Is trabecular bone related to primary stability of miniscrews?

Mariana Marquezan1, Inayá Lima, Ricardo Tadeu Lopes, Eduardo Franzotti Sant'Anna, Margareth Maria Gomes de Souza.   

Abstract

OBJECTIVE: To compare the primary stability of miniscrews inserted into bone blocks of different bone mineral densities (BMDs) with and without cortical bone, and investigate whether some trabecular properties could influence primary stability.
MATERIALS AND METHODS: Fifty-two bone blocks were extracted from fresh bovine pelvic bone. Four groups were created based on bone type (iliac or pubic region) and presence or absence of cortical bone. Specimens were micro-computed tomography imaged to evaluate trabecular thickness, trabecular number, trabecular separation, bone volume density (BV/TV), BMD, and cortical thickness. Miniscrews 1.4 mm in diameter and 6 mm long were inserted into the bone blocks, and primary stability was evaluated by insertion torque (IT), mini-implant mobility (PTV), and pull-out strength (PS).
RESULTS: Intergroup comparison showed lower levels of primary stability when the BMD of trabecular bone was lower and in the absence of cortical bone (P≤.05). The Pearson correlation test showed correlation between trabecular number, trabecular thickness, BV/TV, trabecular BMD, total BMD, and IT, PTV, and PS. There was correlation between cortical thickness and IT and PS (P≤.05).
CONCLUSION: Cancellous bone plays an important role in primary stability of mini-implants in the presence or absence of cortical bone.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24245816      PMCID: PMC8667495          DOI: 10.2319/052513-39.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  34 in total

1.  Primary stability and histomorphometric bone-implant contact of self-drilling and self-tapping orthodontic microimplants.

Authors:  Seçil Çehreli; Ayça Arman-Özçırpıcı
Journal:  Am J Orthod Dentofacial Orthop       Date:  2012-02       Impact factor: 2.650

2.  Bone density at implant sites and its relationship to assessment of bone quality and treatment outcome.

Authors:  Göran Bergkvist; Kwang-Joon Koh; Sten Sahlholm; Eva Klintström; Christina Lindh
Journal:  Int J Oral Maxillofac Implants       Date:  2010 Mar-Apr       Impact factor: 2.804

3.  A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment.

Authors:  Ghada Alsaadi; Marc Quirynen; Katleen Michiels; Reinhilde Jacobs; Daniel van Steenberghe
Journal:  J Clin Periodontol       Date:  2007-04       Impact factor: 8.728

4.  Insertional torque and axial pull-out strength of mini-implants in mandibles of dogs.

Authors:  Karine Kimak Salmória; Orlando Motohiro Tanaka; Odilon Guariza-Filho; Elisa Souza Camargo; Leandro Teixeira de Souza; Hiroshi Maruo
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-06       Impact factor: 2.650

5.  The effect of cortical bone thickness on the stability of orthodontic mini-implants and on the stress distribution in surrounding bone.

Authors:  M Motoyoshi; M Inaba; A Ono; S Ueno; N Shimizu
Journal:  Int J Oral Maxillofac Surg       Date:  2008-10-28       Impact factor: 2.789

6.  Three-dimensional analysis of peri-bone-implant contact of rough-surface miniscrew implants.

Authors:  Hedeki Ikeda; P Emile Rossouw; Phillip M Campbell; Elias Kontogiorgos; Elias Kontogirogos; Peter H Buschang
Journal:  Am J Orthod Dentofacial Orthop       Date:  2011-02       Impact factor: 2.650

Review 7.  Periotest to monitor osseointegration and to check the occlusion in oral implantology.

Authors:  W Schulte; D Lukas
Journal:  J Oral Implantol       Date:  1993       Impact factor: 1.779

8.  Assessment of correlation between computerized tomography values of the bone, and maximum torque and resonance frequency values at dental implant placement.

Authors:  I Turkyilmaz; T F Tözüm; C Tumer; E N Ozbek
Journal:  J Oral Rehabil       Date:  2006-12       Impact factor: 3.837

9.  Orthodontic mini-implant stability and the ratio of pilot hole implant diameter.

Authors:  Miwa Uemura; Mitsuru Motoyoshi; Shinya Yano; Masahito Sakaguchi; Yu Igarashi; Noriyoshi Shimizu
Journal:  Eur J Orthod       Date:  2011-03-10       Impact factor: 3.075

10.  Application of a wireless resonance frequency transducer to assess primary stability of orthodontic mini-implants: an in vitro study in pig ilia.

Authors:  Yu-Yu Su; Benedict Wilmes; Ralf Hönscheid; Dieter Drescher
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Jul-Aug       Impact factor: 2.804

View more
  7 in total

1.  Interradicular trabecular bone density of the lateral maxilla for temporary anchorage devices--a histomorphometric study.

Authors:  Elena Krieger; Heinrich Wehrbein
Journal:  Head Face Med       Date:  2015-02-07       Impact factor: 2.151

2.  Investigating the Mechanical Characteristics of Bone-Metal Implant Interface Using in situ Synchrotron Tomographic Imaging.

Authors:  Sophie Le Cann; Erika Tudisco; Mikael J Turunen; Alessandra Patera; Rajmund Mokso; Magnus Tägil; Ola Belfrage; Stephen A Hall; Hanna Isaksson
Journal:  Front Bioeng Biotechnol       Date:  2019-01-21

3.  Bone quality affects stability of orthodontic miniscrews.

Authors:  Wan-Ping Yu; Ming-Tzu Tsai; Jian-Hong Yu; Heng-Li Huang; Jui-Ting Hsu
Journal:  Sci Rep       Date:  2022-02-18       Impact factor: 4.379

4.  The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2016-06-28       Impact factor: 2.614

5.  Effects of implant tilting and the loading direction on the displacement and micromotion of immediately loaded implants: an in vitro experiment and finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2017-08-28       Impact factor: 2.614

6.  Histomorphometric evaluation of the bone surrounding orthodontic miniscrews according to their adjacent root proximity.

Authors:  Hyun-Ju Oh; Jung-Yul Cha; Hyung-Seog Yu; Chung-Ju Hwang
Journal:  Korean J Orthod       Date:  2018-08-08       Impact factor: 1.372

7.  Evaluation of Primary Stability of Cylindrical and Tapered Implants in Different Bone Types by Measuring Implant Displacement: An In vitro Study.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Tadaaki Kirita
Journal:  Contemp Clin Dent       Date:  2019 Jul-Sep
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.