Two different implant designs and impact of related drilling protocols on primary stability in different bone densities: an in vitro comparison study.

Lars Sennerby, Luca Pagliani, Anders Petersson, Damiano Verrocchi, Stefano Volpe, Peter Andersson.   

Abstract

PURPOSE: To conduct an in vitro comparison study on the primary stability of two different clinically used dental implant designs.
MATERIALS AND METHODS: Eighteen pairs of osteotomies were prepared in fresh bovine bone specimens. The control implant had a subtle tapering and was placed using straight drills. The test implant had a marked tapering and was placed using tapered drills. The bone density at the experimental sites was determined in Hounsfield units (HUs) by using cone beam computed tomography and imaging software. The implants were inserted during continuous registration of insertion torque. The bone blocks were embedded in plaster for firm fixation in a rig for displacement measurements. Resonance frequency analysis (RFA) measurements were taken. A lateral force of 15 N was applied to the RFA transducer and the displacement measured in micrometers. A flex constant (μm/N) was calculated for each measurement.
RESULTS: The test implants displayed statistically significantly higher primary stability than the control implants for all parameters. There was a marked difference in displacement and flex constant in low-density bone in favor of the test implant, but there was no obvious difference in higher-density bone.
CONCLUSION: In this study, placement of a tapered implant design using tapered drills resulted in higher primary stability than a control implant with subtle tapering using straight drills. The results indicate that the novel implant may work particularly well in soft bone densities such as the posterior maxilla. However, clinical studies are needed to confirm this.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26009906     DOI: 10.11607/jomi.3903

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

1.  Evaluation of miniscrew stability using an automatic embedding auxiliary skeletal anchorage device.

Authors:  Yasuhiko Oga; Hiroshi Tomonari; Sangho Kwon; Takaharu Kuninori; Takakazu Yagi; Shouichi Miyawaki
Journal:  Angle Orthod       Date:  2018-09-01       Impact factor: 2.079

2.  A Simplified Prosthetic Implant Loading Protocol: 1-Year Clinical Follow-Up Study.

Authors:  Lorenzo Andreatta; Malin Bjursten Brailsford; Jakob Zwaan
Journal:  Int J Dent       Date:  2020-08-03

3.  A One-year Follow-up Study of a Tapered Hydrophilic Implant Design Using Various Placement Protocols in the Maxilla.

Authors:  Jakob Zwaan; Leonardo Vanden Bogaerde; Herman Sahlin; Lars Sennerby
Journal:  Open Dent J       Date:  2016-12-09

4.  Implant Stability of Biological Hydroxyapatites Used in Dentistry.

Authors:  Maria Piedad Ramírez Fernández; Sergio A Gehrke; Patricia Mazón; Jose L Calvo-Guirado; Piedad N De Aza
Journal:  Materials (Basel)       Date:  2017-06-12       Impact factor: 3.623

5.  Evaluation of Insertion Energy as Novel Parameter for Dental Implant Stability.

Authors:  Tanja Grobecker-Karl; Anthony Dickinson; Siegfried Heckmann; Matthias Karl; Constanze Steiner
Journal:  J Clin Med       Date:  2020-09-15       Impact factor: 4.241

6.  Non-Destructive Removal of Dental Implant by Using the Cryogenic Method.

Authors:  Burak Ak; Emre Gürkan Eroğlu; Abdullah Seckin Ertugrul; Ayla Batu Öztürk; Şakir Necat Yılmaz
Journal:  Medicina (Kaunas)       Date:  2022-06-25       Impact factor: 2.948

7.  Effects of implant tilting and the loading direction on the displacement and micromotion of immediately loaded implants: an in vitro experiment and finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2017-08-28       Impact factor: 2.614

8.  Evaluation of Primary Stability of Cylindrical and Tapered Implants in Different Bone Types by Measuring Implant Displacement: An In vitro Study.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Tadaaki Kirita
Journal:  Contemp Clin Dent       Date:  2019 Jul-Sep
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.