C Yan1, J Cui1, L Huang2, B Du1, L Chen3, G Xue1, S Li1, W Zhang1, L Zhao1, Y Sun1, H Yao1, N Li1, H Zhao1, Y Feng1, S Liu1, Q Zhang1, D Liu4, J Yuan5.
Abstract
OBJECTIVE: To evaluate a reverse traene">nscriptioene">n loop-mediated isothermal aEntities:
Keywords: COVID-19; Detection; RT-LAMP; SARS-CoV-2; Visual
Mesh:
Substances:
Year: 2020 PMID: 32276116 PMCID: PMC7144850 DOI: 10.1016/j.cmi.2020.04.001
Source DB: PubMed Journal: Clin Microbiol Infect ISSN: 1198-743X Impact factor: 8.067
Fig. 1The most appropriate primers and reaction temperature for the reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay. (A) Most suitable primer set for the RT-LAMP assay to amplify the orf1ab gene. (B) Optimal reaction temperature for the RT-LAMP assay with primer orf1ab-4. (C) Most suitable primer set for the RT-LAMP assay to amplify the S gene. (D) Optimal reaction temperature for the RT-LAMP assay with primer S-123. The reaction volume was 25 μL contained 2 μL RNA template, and the template concentration was 1 ng/μL.
Fig. 2Sensitivity of the reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay and conventional PCR for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection. (A,B) Sensitivity of the RT-LAMP assay using primer set orf1ab-4 to target the orf1ab gene for SARS-CoV-2 detection. (D,E) Sensitivity of the RT-LAMP assay using primer set S-123 to target the S gene for SARS-CoV-2 detection. (C,F) Sensitivity of the conventional PCR assay targeting the orf1ab and S genes for SARS-CoV-2 detection. The detection was monitored by turbidity using a Loopamp real-time turbidimeter, and was judged by the naked eye depending on a colour change from orange to green. (B), Lanes 1–9: 108, 107, 106, 105, 104, 103, 102, 101 and 100 copies/μL; (E) lanes 1–8: 107, 106, 105, 104, 103, 102, 101 and 100 copies/μL; (C) and (F) lanes 1–8: 108, 107, 106, 105, 104, 103, 102 and 101 copies/μL. In the sensitivity test, 60 min can be used as the cut-off for the visual detection.
Fig. 3Specificity of the reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection. (A,B) Specificity of the RT-LAMP assay with primer set orf1ab-4 targeted the orf1ab gene for SARS-CoV-2 detection. (C,D) Specificity of the RT-LAMP assay with primer set S-123 targeted the S gene for SARS-CoV-2 detection. The detection was monitored by turbidity using a Loopamp real-time turbidimeter (A, C), and was judged by the naked eye depending on a colour change from orange to green (B, D). S1: human coronavirus (HCoV) -229E-1; S2: HCoV-229E-2; S3: HCoV-NL63-1; S4: HCoV-NL63-2; S5: HCoV-NL63-3; S6: HCoV-OC43-1; S7: HCoV-OC43-2; S8: HCoV-HKU1-1; S9: HCoV-HKU1-2; S10: H3N2-1; S11: H3N2-2; S12: H3N2-3; S13: H3N2-4; S14: H1N1-1; S15: H1N1-2; S16: influenza B-1; S17: influenza B-2; S18: influenza B-3; S19: parainfluenza (PIV) -1-1; S20: PIV-1-2; S21: PIV-2-1; S22: PIV-2-2; S23: PIV-3-1; S24: PIV-3-2; S25: PIV-4-1; S26: PIV-4-2; S27: adenovirus (ADV) -1-1; S28: ADV-1-2; S29: ADV-2-1; S30: ADV-2-2; S31: ADV-3-1; S32: ADV-3-2; S33: ADV-4; S34: ADV-5-1; S35: ADV-5-2; S36: ADV-6; S37: ADV-7-1; S38: ADV-7-2; S39: respiratory syncytial virus (RSV) A-1; S40: RSV A-2; S41: RSV B-1; S42: RSV B-2; S43: human metapneumovirus HMPV-1; S44: HMPV-2; S45: human bocavirus (BoV) -1; S46: BoV-2; S47: rhinovirus (Rh) A-1; S48: Rh A-2; S49: Rh B-1; S50: Rh B-2; S51: Rh C; S52: Mycoplasma pneumoniae (MP) -FH; S53: MP-M129; S54: Haemophilus influenzae; S55: Staphylococcus aureus; S56: Klebsiella pneumoniae; S57: Streptococcus pneumoniae; S58: Pseudomonas aeruginosa; S59: SARS-CoV; S60: Middle East respiratory syndrome-CoV; PC, positive control (pseudo-virus); NC, negative control (distilled water).
Comparison between the reverse transcription loop-mediated isothermal amplification assay and RT-PCR for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection from clinical samples
| Results | RT-LAMP | Real-time RT-PCR | |
|---|---|---|---|
| Positive samples | 58 | 58 | 58 (Ct < 38) |
| Mean ± SD (min) | 25.36 ± 5.09 | 27.21 ± 5.74 | |
| Range | 16–35 | 18–40 | |
| Negative samples | 72 (no change in colour) | 72 (no change in colour) | 72 (no peak) |
RT-LAMP, reverse transcription loop-mediated isothermal amplification; RT-PCR, reverse transcription-polymerase chain reaction; SD, standard deviation.
Range: the shortest detection time – the longest detection time.