Benjamin Gauter-Fleckenstein1, Erol Tülümen2,3, Boris Rudic2,3, Martin Borggrefe2,3, Martin Polednik4, Jens Fleckenstein4.
Abstract
PURPOSE: In the beam penumbra of stereotactic body radiotherapy volumes, dose rate effects in implantable cardioverter-defibrillators (ICDs) may be the predominant cause for failures in the absence of neutron-generating photon energies. We investigate such dose rate effects in ICDs and provide evidence for safe use of lung tumor stereotactic radioablation with flattening filter free (FFF) and flattened 6 Megavolt (MV) beams in ICD-bearing patients.Entities:
Keywords: Cardiac implantable electronic devices; Flattening filter free radiation; Malfunction; Planning target volume; Sterotactic body radiotherapy
Mesh:
Year: 2022 PMID: 35267050 PMCID: PMC9165256 DOI: 10.1007/s00066-022-01911-8
Source DB: PubMed Journal: Strahlenther Onkol ISSN: 0179-7158 Impact factor: 4.033
Fig. 1a Sketch of the experimental setup. Only one implantable cardioverter–defibrillator (ICD) was present on each side at a time; either at 1.0, 2.5, or at 7.0 cm. b Positioning of the ICDs with respect to the radiation field after removing the 2 cm bolus buildup-material. ICDs located at the 1 cm and 2.5 cm distance from field edge
Beam characteristics and experimental setup of the different groups
| Group | |||||
|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | |
| FLAT | FLAT | FFF | FFF | FFF | |
| 23.5 | 23.5 | 7.2 | 7.2 | 7.2 | |
| 1.0 | 2.5 | 1.0 | 2.5 | 7.0 | |
| 10 | 16 | 10 | 10 | 16 |
FLAT flattened photon radiation, FFF flattening filter free photon radiation, ICD implantable cardioverter–defibrillator
Radiation-induced ICD failures, maximum dose, and local dose rates at the specified measurement positions
| Group | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| 3.9 | 1.5 | 3.4 | 1.1 | 0.4 | |
| 16.8 | 6.6 | 46.8 | 15.6 | 6.0 | |
| – | – | ||||
| – | – | ||||
| – | – | – | – | ||
ICD implantable cardioverter–defibrillator
a Letters in parentheses denote manufacturer: Medtronic (M); Saint Jude Medical/Abbott (SJM)
Fig. 2Local dose rate as a function of the distance from a 5 × 5 cm2 radiation field (constant dose rate in the isocenter for FLAT 430 cGy/min, FFF 1400 cGy/min). Zero distances would be at the 50% dose level at SSD = 98 cm. Measured local dose rates at specified positions and fitted curve which was received from Monte Carlo-based dose calculations. Dashed line denotes local dose rate of 6 cGy/min extrapolating to 2.5 cm distance from FLAT beam field edge and 7 cm distance from FFF beam field edge