To evaluate the accuracy of a commercial optical surface tracking (OST) system and to demonstrate how it can be implemented to monitor patient positioning during non-coplanar single isocenter stereotactic treatments of brain metastases. A 3-camera OST system was used (Catalyst HD™, C-RAD) on a TruebeamSTx with a 6DoF couch. The setup accuracy and agreement between the OST system, and CBCT and kV-MV imaging at couch angles 0° and 270°, respectively, were examined. Film measurements at 3 depths in the Rando-Alderson phantom were performed using a single isocenter non-coplanar VMAT plan containing 4 brain lesions. Setup of the phantom was performed with CBCT at couch 0° and subsequently monitored by OST at other couch angles. Setup data for 7 volunteers were collected to evaluate the accuracy and reproducibility of the OST system at couch angles 0°, 45°, 90°, 315°, and 270°. These results were also correlated to the couch rotation offsets obtained by a Winston-Lutz (WL) test. The Rando-Alderson phantom, as well as volunteers, were fixated using open face masks (Orfit). For repeated tests with the Rando-Alderson phantom, deviations between rotational and translational isocenter corrections for CBCT and OST systems are always within 0.2° (pitch, roll, yaw), and 0.1mm and 0.5mm (longitudinal, lateral, vertical) for couch positions 0° and 270°, respectively. Dose deviations between the film and TPS doses in the center of the 4 lesions were -1.2%, -0.1%, -0.0%, and -1.9%. Local gamma evaluation criteria of 2%/2 mm and 3%/1 mm yielded pass rates of 99.2%, 99.2%, 98.6%, 89.9% and 98.8%, 97.5%, 81.7%, 78.1% for the 4 lesions. Regarding the volunteers, the mean translational and rotational isocenter shift values were (0.24 ± 0.09) mm and (0.15 ± 0.07) degrees. Largest isocenter shifts were found for couch angles 45˚ and 90˚, confirmed by WL couch rotation offsets. Patient monitoring during non-coplanar VMAT treatments of brain metastases is feasible with submillimeter accuracy.Entities:
Keywords: brain metastases; non-coplanar; open face mask; optical surface tracking; stereotactic radiosurgery
Mesh:
Year: 2020 PMID: 32237274 PMCID: PMC7324699 DOI: 10.1002/acm2.12866
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1(a) Orfit open face mask and T‐shaped vacuum bag, (b) Catalyst HDTM in kV‐MV setup using the ExaFix‐3 baseplate and (c) in setup at couch 0˚(3 cameras are indicated with arrows)
Fig. 3(a) A five arc single isocenter non‐coplanar VMAT plan with 4 PTVs, (b) verified using film dosimetry with EBT‐XD films in the first 3 slices of the Rando‐Alderson phantom head
The dosimetric agreement between film and TPS doses is presented by the deviations between measured and calculated doses in the center of the 4 PTVs and by a local gamma evaluation criterion of 2%/2mm as well as 3%/1mm for the 4 PTVs irradiated with single isocenter non‐coplanar VMAT and monitored using Catalyst HDTM. To put these numbers into perspective, also the results for PTV‐1 irradiated in a co‐planar set‐up (couch 0˚) are given
| PTV |
Dfilm (Gy) |
DTPS (Gy) |
Δ (Dfilm‐DTPS) (%) |
Agreement score (2%/2mm) |
Agreement score (3%/1mm) |
|---|---|---|---|---|---|
| 1 | 9.71 | 9.82 | −1.2% | 99.2% | 98.8% |
| 1 (co‐planar) | 9.15 | 9.30 | −1.6% | 99.8% | 99.9% |
| 2 | 9.86 | 9.87 | −0.1% | 99.2% | 97.5% |
| 3 | 9.82 | 9.82 | −0.0% | 98.6% | 81.7% |
| 4 | 9.11 | 9.29 | −1.9% | 89.9% | 78.1% |
Repeated WL tests over a 10 months period: the deviation for the gantry position at 0˚ and couch positions 0°, 45°, 90°, 315°, and 270° show that the accuracy of the couch exceeds 0.5 mm for couch angles 45˚ and 90˚
| WL test | Deviation (mm) at couch 0˚ | Deviation (mm) at couch 315˚ | Deviation (mm) at couch 270˚ | Deviation (mm) at couch 90˚ | Deviation (mm) at couch 45˚ |
|---|---|---|---|---|---|
| 1 | 0.35 | 0.17 | 0.25 | 0.63 | 0.74 |
| 2 | 0.39 | 0.17 | 0.25 | 0.63 | 0.55 |
| 3 | 0.39 | 0.17 | 0.39 | 0.63 | 0.63 |
| 4 | 0.17 | 0.25 | 0.17 | 0.39 | 0.39 |
| 5 | 0.55 | 0.17 | 0.17 | 0.63 | 0.63 |
| 6 | 0.25 | 0.17 | 0.25 | 0.63 | 0.63 |
| 7 | 0.39 | 0.39 | 0.55 | 0.39 | 0.52 |
| 8 | 0.17 | 0.35 | 0.17 | 0.55 | 0.39 |
| 9 | 0.35 | 0.17 | 0.25 | 0.63 | 0.74 |
| 10 | 0.63 | 0.35 | 0.52 | 0.74 | 0.74 |
| Mean | 0.36 | 0.24 | 0.30 | 0.59 | 0.60 |
| SD | 0.15 | 0.09 | 0.14 | 0.11 | 0.13 |