Vivian U Y Chow1, Michael L M Cheung1, Monica W K Kan1,2, Anthony T C Chan1,2.
Abstract
Accurate detection of patient shift is essential during radiation therapy such that optimal dose is delivered to the tumor while minimizing radiation to surrounding normal tissues. The shift detectability of a newly developed optical surface and thermal tracking system, which was known as ExacTrac Dynamic (EXTD), was evaluated by comparing its performance with the image guidance under cone-beam computed tomography (CBCT). Anthropomorphic cranial and pelvis phantoms with internal bone-like structures and external heat pad were utilized to study the shift detection discrepancy between EXTD system and CBCT. Random displacements within the range of ± 2 cm for translations and ± 2 degrees for rotations were intentionally applied to the phantom. Positional shifts detected by optical surface and thermal tracking (EXTD_Thml), stereoscopic X-ray (EXTD_Xray), and CBCT were compared in 6 degrees of freedom. The translational difference between EXTD_Thml and CBCT was 0.57 ± 0.41 mm and 0.66 ± 0.40 mm for cranial and pelvis phantom, respectively, while it was 0.60 ± 0.43 mm and 0.76 ± 0.49 mm between EXTD_Xray and CBCT, respectively. For rotational movement, the difference between EXTD_Thml and CBCT was 0.19 ± 0.16° and 0.19 ± 0.22° for cranial and pelvis phantom, respectively, while it was 0.13 ± 0.18° and 0.65 ± 0.46° between EXTD_Xray and CBCT, respectively. This study demonstrated that the EXTD system with thermal mapping ability could offer comparable accuracy for shift detection with CBCT on both cranial and pelvis phantoms.Entities:
Keywords: cone-beam computed tomography; optical surface tracking; patient positioning; stereotactic radiosurgery; thermal mapping
Mesh:
Year: 2022 PMID: 35188333 PMCID: PMC9121052 DOI: 10.1002/acm2.13567
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIGURE 1Components of ExacTrac Dynamic (EXTD) system It consists of two x‐ray tubes on the floor with two flat panel detectors on each side of a set of optical & thermal cameras that is centrally ceiling‐mounted above the couch.
FIGURE 2Anthropomorphic cranial (left) and pelvis (right) verification phantoms. The cranial phantom was fitted with Brainlab cranial 4Pi open face mask. Heat pad was applied on the phantom to generate temperature difference for thermal tracking.
Differences of shift detected by EXTD_Thml and EXTD_Xray against cone‐beam computed tomography (CBCT) for 20 measurements in all six directions for cranial verification phantom
| EXTD_Thml | EXTD_Xray | |||||
|---|---|---|---|---|---|---|
| Direction | Average ± 1 SD | Range |
| Average ± 1 SD | Range |
|
| Vertical | 0.23 ± 0.19 | −0.6, 0.7 | 0.447 | 0.26 ± 0.19 | −0.7, 0.6 | 0.567 |
| Longitudinal | 0.40 ± 0.26 | −0.8, 0.3 | 0.001 | 0.39 ± 0.28 | −0.8, 0.4 | 0.001 |
| Lateral | 0.34 ± 0.26 | −0.9, 0.2 | 0.002 | 0.37 ± 0.26 | −0.9, 0.2 | 0.003 |
| Vectortranslation (mm) | 0.57 ± 0.41 | 0.60 ± 0.43 | ||||
| Roll | 0.11 ± 0.09 | −0.2, 0.3 | 0.640 | 0.06 ± 0.08 | −0.2, 0.2 | 0.623 |
| Pitch | 0.11 ± 0.08 | −0.3, 0.1 | 0.003 | 0.09 ± 0.14 | −0.6, 0.1 | 0.223 |
| Yaw | 0.11 ± 0.10 | −0.2, 0.4 | 0.032 | 0.07 ± 0.09 | –0.3, 0.2 | 0.874 |
| Vectorrotation (°) | 0.19 ± 0.16 | 0.13 ± 0.18 | ||||
The p‐value was calculated using Wilcoxon signed‐ranks test.
The p‐value has a statistical significance of difference (p < 0.05).
Differences of shift detected by EXTD_Thml and EXTD_Xray against cone‐beam computed tomography (CBCT) for 20 measurements in all six directions for pelvis verification phantom
| EXTD_Thml | EXTD_Xray | ||||||
|---|---|---|---|---|---|---|---|
| Direction | Average ± 1 SD | Range |
| Average ± 1 SD | Range |
| |
| Vertical | 0.24 ± 0.17 | −0.7, 0.4 | 0.965 | 0.34 ± 0.17 | −0.8, 0.5 | 0.027 | |
| Longitudinal | 0.47 ± 0.32 | −1.0, 0.2 | 0.001 | 0.59 ± 0.41 | −1.3, 0.1 | <0.001 | |
| Lateral | 0.39 ± 0.18 | −0.8, 0.4 | 0.001 | 0.34 ± 0.21 | −0.8, 0.3 | 0.001 | |
| Vectortranslation (mm) | 0.66 ± 0.40 | 0.76 ± 0.49 | |||||
| Roll | 0.13 ± 0.19 | −0.2, 0.9 | 0.064 | 0.23 ± 0.25 | −0.5, 1.1 | 0.158 | |
| Pitch | 0.11 ± 0.07 | −0.2, 0.2 | 0.829 | 0.51 ± 0.31 | −0.3, 1.3 | 0.001 | |
| Yaw | 0.09 ± 0.08 | −0.1, 0.2 | 0.012 | 0.31 ± 0.24 | −0.6, 1.1 | 0.012 | |
| Vectorrotation (°) | 0.19 ± 0.22 | 0.65 ± 0.46 | |||||
The p‐value was calculated using Wilcoxon signed ranks test.
The p‐value has a statistical significance of difference (p < 0.05).
Differences of shift detected by EXTD_Thml and EXTD_Xray for 20 measurements in all six directions for cranial and pelvis verification phantoms
| Cranial phantom | Pelvis phantom | |||||
|---|---|---|---|---|---|---|
| Direction | Average ± 1 SD | Range |
| Average ± 1 SD | Range |
|
| Vertical | 0.10 ± 0.09 | −0.2, 0.4 | 0.506 | 0.16 ± 0.14 | −0.2, 0.4 | 0.003 |
| Longitudinal | 0.11 ± 0.11 | ‐0.3, 0.3 | 0.872 | 0.19 ± 0.16 | −0.3, 0.6 | 0.012 |
| Lateral | 0.05 ± 0.06 | −0.1, 0.2 | 0.564 | 0.16 ± 0.14 | −0.5, 0.4 | 0.618 |
| Vectortranslation (mm) | 0.16 ± 0.15 | 0.30 ± 0.25 | ||||
| Roll | 0.10 ± 0.11 | −0.2, 0.4 | 0.260 | 0.15 ± 0.11 | −0.4, 0.4 | 0.662 |
| Pitch | 0.12 ± 0.14 | −0.3, 0.6 | 0.044 | 0.50 ± 0.28 | −1.1, 0.5 | 0.001 |
| Yaw | 0.09 ± 0.07 | −0.1, 0.2 | 0.006 | 0.26 ± 0.22 | −0.9, 0.6 | 0.045 |
| Vectorrotation (°) | 0.18 ± 0.19 | 0.58 ± 0.37 | ||||
The p‐value was calculated using Wilcoxon signed ranks test.
The p‐value has a statistical significance of difference (p < 0.05).