Measurement of cone beam CT coincidence with megavoltage isocentre and image sharpness using the QUASAR Penta-Guide phantom.

J R Sykes1, R Lindsay, C J Dean, D S Brettle, D R Magee, D I Thwaites.   

Abstract

For image-guided radiotherapy (IGRT) systems based on cone beam CT (CBCT) integrated into a linear accelerator, the reproducible alignment of imager to x-ray source is critical to the registration of both the x-ray-volumetric image with the megavoltage (MV) beam isocentre and image sharpness. An enhanced method of determining the CBCT to MV isocentre alignment using the QUASAR Penta-Guide phantom was developed which improved both precision and accuracy. This was benchmarked against our existing method which used software and a ball-bearing (BB) phantom provided by Elekta. Additionally, a method of measuring an image sharpness metric (MTF(50)) from the edge response function of a spherical air cavity within the Penta-Guide phantom was developed and its sensitivity was tested by simulating misalignments of the kV imager. Reproducibility testing of the enhanced Penta-Guide method demonstrated a systematic error of <0.2 mm when compared to the BB method with near equivalent random error (s=0.15 mm). The mean MTF(50) for five measurements was 0.278+/-0.004 lp mm(-1) with no applied misalignment. Simulated misalignments exhibited a clear peak in the MTF(50) enabling misalignments greater than 0.4 mm to be detected. The Penta-Guide phantom can be used to precisely measure CBCT-MV coincidence and image sharpness on CBCT-IGRT systems.

Entities:  

Mesh:

Year:  2008        PMID: 18758000     DOI: 10.1088/0031-9155/53/19/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

1.  Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study.

Authors:  Markus Stock; Marlies Pasler; Wolfgang Birkfellner; Peter Homolka; Richard Poetter; Dietmar Georg
Journal:  Radiother Oncol       Date:  2009-08-18       Impact factor: 6.280

2.  On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

Authors:  Weiliang Du; Jennifer L Johnson; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

Review 3.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

4.  Alignment of multiradiation isocenters for megavoltage photon beam.

Authors:  Yin Zhang; Kai Ding; Garth Cowan; Erik Tryggestad; Elwood Armour; Ken Kang-Hsin Wang
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

5.  Evaluation of IsoCal geometric calibration system for Varian linacs equipped with on-board imager and electronic portal imaging device imaging systems.

Authors:  Song Gao; Weiliang Du; Peter Balter; Peter Munro; Andrew Jeung
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

6.  The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases.

Authors:  Ans C C Swinnen; Michel C Öllers; Chin Loon Ong; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2020-04-01       Impact factor: 2.102

7.  Evaluation of the RUBY modular QA phantom for planar and non-coplanar VMAT and stereotactic radiations.

Authors:  Daniela Poppinga; Jana Kretschmer; Leonie Brodbek; Jutta Meyners; Bjoern Poppe; Hui Khee Looe
Journal:  J Appl Clin Med Phys       Date:  2020-08-14       Impact factor: 2.102