Comparison of the expression of the volumetric alteration artifact in cylindrical and triangular fields of view in two cone-beam computed tomography devices.

Deivi Cascante-Sequeira1, Fernanda Coelho-Silva2, Lucas P Lopes Rosado2, Deborah Queiroz Freitas2, Sergio Lins de-Azevedo-Vaz3, Francisco Haiter-Neto2.   

Abstract

OBJECTIVE: To compare the expression of the volumetric alteration (VA) artifact between cylindrical and convex triangular fields of view (FOVs) using high-density materials in different positions in two cone-beam computed tomography (CBCT) devices.
MATERIALS AND METHODS: Cylinders of five high-density materials (amalgam, chromium-cobalt, gutta-percha, titanium, and zirconium) with known physical volume were individually submitted to CBCT acquisition in four positions inside a polymethylmethacrylate phantom using two different FOV shapes (convex triangular and cylindrical) on the Veraviewepocs® R100 (R100) and Veraview® X800 (X800) devices. Two oral radiologists obtained the tomographic volumes by segmenting each cylinder. The difference between the tomographic and physical volumes corresponded to the VA. These values were analyzed by intraclass correlation coefficient and analysis of variance for repeated measures with Tukey post hoc test (α = 5%).
RESULTS: The FOV influenced the VA only in the X800 device (p = 0.014): the VA in the triangular FOV was greater than in the cylindrical FOV. The VA in the triangular FOV of the X800 device was greater than the R100 device (p < 0.0001). The material influenced the expression of the VA only in the R100 device (p < 0.0001); gutta-percha presented the highest VA, being underestimated, and differing from the other materials (p < 0.01).
CONCLUSION: The triangular FOV increased the VA of high-density materials in the X800 device. CLINICAL RELEVANCE: It is important to know if there is an influence on the volumetric alteration artifact of dental materials due to the different image formation geometry in the convex triangular FOV.
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Artifact; Cone-beam computed tomography; Dental materials; Volumetric alteration

Mesh:

Substances:

Year:  2021        PMID: 34324084     DOI: 10.1007/s00784-021-04086-1

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.606


  16 in total

1.  Beam hardening artefacts occur in dental implant scans with the NewTom cone beam CT but not with the dental 4-row multidetector CT.

Authors:  F G Draenert; E Coppenrath; P Herzog; S Müller; U G Mueller-Lisse
Journal:  Dentomaxillofac Radiol       Date:  2007-05       Impact factor: 2.419

2.  On cone-beam computed tomography artifacts induced by titanium implants.

Authors:  Ralf Kurt Willy Schulze; Dorothea Berndt; Bernd d'Hoedt
Journal:  Clin Oral Implants Res       Date:  2009-10-21       Impact factor: 5.977

Review 3.  Artefacts in CBCT: a review.

Authors:  R Schulze; U Heil; D Gross; D D Bruellmann; E Dranischnikow; U Schwanecke; E Schoemer
Journal:  Dentomaxillofac Radiol       Date:  2011-07       Impact factor: 2.419

4.  Inaccuracy of buccal bone thickness estimation on cone-beam CT due to implant blooming: An ex-vivo study.

Authors:  Tony Vanderstuyft; Mihai Tarce; Bahoz Sanaan; Reinhilde Jacobs; Karla de Faria Vasconcelos; Marc Quirynen
Journal:  J Clin Periodontol       Date:  2019-09-09       Impact factor: 8.728

5.  Quantitative evaluation of metal artifacts using different CBCT devices, high-density materials and field of views.

Authors:  Marina Codari; Karla de Faria Vasconcelos; Laura Ferreira Pinheiro Nicolielo; Francisco Haiter Neto; Reinhilde Jacobs
Journal:  Clin Oral Implants Res       Date:  2017-04-22       Impact factor: 5.977

6.  Comparison of manual and dose reduction modes of a MORITA R100 CBCT.

Authors:  Dennis Rottke; Julia Dreger; Kunihiko Sawada; Kazuya Honda; Dirk Schulze
Journal:  Dentomaxillofac Radiol       Date:  2018-10-03       Impact factor: 2.419

Review 7.  Technical aspects of dental CBCT: state of the art.

Authors:  R Pauwels; K Araki; J H Siewerdsen; S S Thongvigitmanee
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

8.  Artefact expression associated with several cone-beam computed tomographic machines when imaging root filled teeth.

Authors:  K F Vasconcelos; L F P Nicolielo; M C Nascimento; F Haiter-Neto; F N Bóscolo; J Van Dessel; M EzEldeen; I Lambrichts; R Jacobs
Journal:  Int Endod J       Date:  2014-11-15       Impact factor: 5.264

9.  Assessment of Volumetric Distortion Artifact in Filled Root Canals Using Different Cone-beam Computed Tomographic Devices.

Authors:  Berkan Celikten; Reinhilde Jacobs; Karla deFaria Vasconcelos; Yan Huang; Laura Ferreira Pinheiro Nicolielo; Kaan Orhan
Journal:  J Endod       Date:  2017-06-30       Impact factor: 4.171

10.  Cone beam computed tomography in dentomaxillofacial radiology: a two-decade overview.

Authors:  Hugo Gaêta-Araujo; Tamara Alzoubi; Karla de Faria Vasconcelos; Kaan Orhan; Ruben Pauwels; Jan W Casselman; Reinhilde Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2020-06-15       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.