Literature DB >> 23768878

Prospects and challenges of rendering tissue density in Hounsfield units for cone beam computed tomography.

Roberto Molteni1.   

Abstract

OBJECTIVES: Limitations in rendering of tissue density in Hounsfield units (HUs) for cone beam computed tomography (CBCT) are described and illustrated using a phantom and two CBCT systems to demonstrate grayscale measurement variability.
MATERIALS AND METHODS: The basis of the HU scale, its correlation with measured computed tomography (CT) numbers, and the limitations in the accuracy of such correlation due to artifacts are discussed. Rendering of tissue densities based on HU values of two CBCT systems [NewTom VGi and Hyperion X9, respectively large and small field of view (FOV)] are measured using a phantom.
RESULTS: Data produced from small FOV CBCT acquisition are generally less affected by artifacts compared with large FOV CBCT.
CONCLUSIONS: Artifacts challenge the accurate conversion of density values into HUs. Care should be taken when interpreting quantitative density measurements obtained with CBCT. With more advanced software and methods, it may be possible to improve the consistency and accuracy of density measurements.
Copyright © 2013 Elsevier Inc. All rights reserved.

Mesh:

Year:  2013        PMID: 23768878     DOI: 10.1016/j.oooo.2013.04.013

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  38 in total

1.  Motion artefacts in cone beam CT: an in vitro study about the effects on the images.

Authors:  Cosimo Nardi; Roberto Molteni; Chiara Lorini; Gian G Taliani; Benedetta Matteuzzi; Elisa Mazzoni; Stefano Colagrande
Journal:  Br J Radiol       Date:  2015-11-18       Impact factor: 3.039

Review 2.  CBCT-based bone quality assessment: are Hounsfield units applicable?

Authors:  R Pauwels; R Jacobs; S R Singer; M Mupparapu
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

3.  Conditions and calibration to obtain comparable grey values between different clinical cone beam computed tomography scanners.

Authors:  Gregory M England; Eun-Sang Moon; Jordan Roth; Toru Deguchi; Allen R Firestone; F Michael Beck; Do-Gyoon Kim
Journal:  Dentomaxillofac Radiol       Date:  2016-11-15       Impact factor: 2.419

4.  Image quality optimization using a narrow vertical detector dental cone-beam CT.

Authors:  Danieli Moura Brasil; Ruben Pauwels; Wim Coucke; Francisco Haiter-Neto; Reinhilde Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2019-01-31       Impact factor: 2.419

5.  Variation in voxel value distribution and effect of time between exposures in six CBCT units.

Authors:  R Spin-Neto; E Gotfredsen; A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2014-02-03       Impact factor: 2.419

6.  Influence of exposure factors on the variability of CBCT voxel values: a phantom study.

Authors:  M L Oliveira; D Q Freitas; G M B Ambrosano; F Haiter-Neto
Journal:  Dentomaxillofac Radiol       Date:  2014-05-28       Impact factor: 2.419

7.  Effect of ultra-low doses, ASIR and MBIR on density and noise levels of MDCT images of dental implant sites.

Authors:  Gerlig Widmann; Reema Al-Shawaf; Peter Schullian; Ra'ed Al-Sadhan; Romed Hörmann; Asma'a A Al-Ekrish
Journal:  Eur Radiol       Date:  2016-09-21       Impact factor: 5.315

8.  Effect of exposure parameters and voxel size on bone structure analysis in CBCT.

Authors:  R Pauwels; T Faruangsaeng; T Charoenkarn; N Ngonphloy; S Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2015-06-09       Impact factor: 2.419

9.  Standardized method to quantify the variation in voxel value distribution in patient-simulated CBCT data sets.

Authors:  R Spin-Neto; E Gotfredsen; A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2014-10-29       Impact factor: 2.419

10.  Assessment of CT numbers in limited and medium field-of-view scans taken using Accuitomo 170 and Veraviewepocs 3De cone-beam computed tomography scanners.

Authors:  Matheus L Oliveira; Guilherme M Tosoni; David H Lindsey; Kristopher Mendoza; Sotirios Tetradis; Sanjay M Mallya
Journal:  Imaging Sci Dent       Date:  2014-11-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.