Ghost-image analysis in phakic intraocular lenses with central hole as a potential cause of dysphotopsia.

Timo Eppig1, Corinna Spira2, Themistoklis Tsintarakis2, Moatasem El-Husseiny2, Alan Cayless2, Marc Müller2, Berthold Seitz2, Achim Langenbucher2.   

Abstract

PURPOSE: To analyse the optical effect of an artificial hole in the optic centre of posterior chamber phakic intraocular lenses.
SETTING: Institute of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
DESIGN: Experimental simulation study.
METHODS: Four eye models with an ametropia of -4 D, -8 D, +4 D, and +8 D were created in the ray tracing software ASAP. Refractive correction of these models was implemented with a model of an Implantable Collamer Lens (ICL). Each eye was set up twice with 1 eye receiving a conventional ICL without a central hole and the second an ICL with a central hole. Ray bundles were traced for lateral visual field angles from 0 to 60 degrees in steps of 1 degree. Ray propagation and retinal illumination were then compared between the 2 ICL models.
RESULTS: All eye models showed ghost images originating from the anterior surface of the ICL. Eye models with the ICL with central hole showed additional light spots in the peripheral areas of the retina originating from reflections at the cylindrical wall of the central hole in the ICL. The average intensity of ghost images in the temporal retinal hemisphere was between 30 and 40 dB less than the maximum intensity of the primary image.
CONCLUSION: A central hole within a posterior chamber phakic intraocular lens may cause stray light and ghost images (positive dysphotopsia) although the on-axis visual quality of the eye with the ICL is mostly unaffected. FINANCIAL DISCLOSURE: The authors have no financial interest in any of the material presented in this paper.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26703506     DOI: 10.1016/j.jcrs.2015.05.034

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  [Intraocular pressure after implantation of an ICL with aquaport : Development of intraocular pressure after implantation of an ICL (model V4c) with aquaport without iridotomy].

Authors:  B Repplinger; T Kohnen
Journal:  Ophthalmologe       Date:  2018-01       Impact factor: 1.059

2.  Effects of tilt and decentration of Visian Implantable Collamer Lens (ICL V4c) on visual quality: an observational study.

Authors:  Lingling Niu; Zhe Zhang; Huamao Miao; Jing Zhao; Meiyan Li; Ji C He; Peijun Yao; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2022-07-05       Impact factor: 2.086

3.  Optical quality after posterior chamber Phakic implantation of an intraocular Lens with a central hole (V4c implantable Collamer Lens) under different lighting conditions.

Authors:  Tianrui He; Yi Zhu; Jibo Zhou
Journal:  BMC Ophthalmol       Date:  2020-03-04       Impact factor: 2.209

4.  One-year natural course of corneal densitometry in high myopic patients after implantation of an implantable collamer lens (model V4c).

Authors:  Xun Chen; Yang Shen; Haipeng Xu; Xiaoying Wang; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2020-02-12       Impact factor: 2.209

5.  Two-Year Visual Outcomes of Evolution Implantable Collamer Lens and Small Incision Lenticule Extraction for the Correction of Low Myopia.

Authors:  Mengjun Fu; Meiyan Li; Yiyong Xian; Zhiqiang Yu; Haorun Zhang; Joanne Choi; Lingling Niu; Xiaoying Wang; Xingtao Zhou
Journal:  Front Med (Lausanne)       Date:  2022-04-13
  5 in total