Eun Joo Choi1, Gwangcheol Go1, Woong Ki Han1, Pyung-Bok Lee1,2.
Abstract
BACKGROUND: The aim of this study was to evaluate radiation exposure to the eye and thyroid in pain physicians during the fluoroscopy-guided cervical epidural block (CEB).Entities:
Keywords: Anesthesia; Cervical Vertebrae; Chronic Pain; Epidural; Eye; Fluoroscopy; Physicians; Protection; Radiation; Thyroid Gland
Year: 2020 PMID: 31888321 PMCID: PMC6944368 DOI: 10.3344/kjp.2020.33.1.73
Source DB: PubMed Journal: Korean J Pain ISSN: 2005-9159
Fig. 1The cervical epidural block (CEB) technique. (A) Lateral view of a pain physician performing a CEB. (B) Anterior view of a pain physician performing a CEB.
Fig. 2Thermoluminescent dosimeters were placed on the forehead (A), outside the thyroid protector (B), inside the thyroid protector (C), inside the apron (D), and outside the apron (E).
Fig. 3Measurement of the distance (A) between the center of the X-ray field (B) and the physician. The distance between the X-ray beam and the physician was measured during imaging after needle manipulation.
Patient Demographics and Radiation-related Procedural Data
| Variable | Fellow (n = 50) | Professor (n = 50) | |
|---|---|---|---|
| Age (yr) | 56.6 ± 12.4 | 54.2 ± 11.3 | 0.96 |
| Sex (male/female) | 26/24 | 27/23 | |
| Height (cm) | 163.1 ± 7.9 | 168.1 ± 1.3 | 0.21 |
| Weight (kg) | 63.1 ± 12.6 | 66.2 ± 6.1 | 0.27 |
| Total time of procedure (min) | 13.3 ± 1.7 | 10.6 ± 1.3 | 0.35 |
| Time of radiation exposure/procedure (sec) | 22.9 ± 8.8 | 17.7 ± 2.4 | 0.18 |
| Radiation dose/procedure (cGy/cm2) | −95.4 ± 66.1 | 89.9 ± 13.3 | 0.16 |
| Physician distance from radiation beam (cm) | 37.5 ± 2.1 | −41.2 ± 3.6 | 0.03 |
Values are presented as the mean ± standard deviation or number only.
Radiation Dose Measured with Thermoluminescent Dosimeters (TLDs)
| Physician | TLD position | Radiation dose (mSv) | Radiation reduction with protection (%) |
|---|---|---|---|
| Fellow | Forehead | 0.74 | |
| Outside of thyroid protector | 0.55 | ||
| Inside of thyroid protector | 0.43 | 22 | |
| Outside of apron | 0.71 | ||
| Inside of apron | 0.47 | 34 | |
| Professor | Forehead | 0.62 | |
| Outside of thyroid protector | 0.47 | ||
| Inside of thyroid protector | 0.32 | 32 | |
| Outside of apron | 0.64 | ||
| Inside of apron | 0.42 | 35 | |
| Control | Inside operating room | 0.40 | |
| Outside operating room | 0.37 |
Fig. 4Annual equivalent dose (approximately 600 cases) was calculated from 2-month (100 cases) exposure data, as measured using thermoluminescent dosimeters. Annual maximum permissible radiation doses set by the International Commission on Radiological Protection are 500 and 20 mSv for the thyroid and crystalline lens, respectively. In: inside, Out: outside.