How protective are the lead aprons we use against ionizing radiation?

Orhan Oyar1, Arzu Kışlalıoğlu.   

Abstract

PURPOSE: To evaluate, in terms of their protective features, the lead aprons used in areas working with ionizing radiation at a hospital by analyzing qualitative and quantitative aspects using a variety of methods.
MATERIALS AND METHODS: Eighty-five protective lead aprons used in our hospital's clinics to work with ionizing radiation were analyzed in the radiology unit. Each apron was identified by registering the unit from which it had been obtained and by how long it had been used, its storage condition, and its lead thickness. X-ray films of the aprons, controlled according to their appearances, durability and cleanliness, were taken to evaluate their internal structure; their permeability was measured with electronic dosimeters in terms of their absorbent features. All of these data were compared with the results acquired from brand-new, Turkish Standards Institution approved aprons having different lead thicknesses.
RESULTS: Regarding internal structure homogeneity, only 13 (15.3%) of 85 aprons were found to be at normal levels and usable. A total of 14 (16.5%) of the remaining 72 aprons' radiation absorptions were at normal levels, but folds were observed in their protective lead layers. The remaining 58 aprons (68.2%) were found to be defective. All of the aprons were considered to be defective in terms of their radiation permeability.
CONCLUSION: All of the aprons were found insufficient for protection and were more radioparent than the defined limits; it was concluded that they must be replaced by new ones.

Mesh:

Substances:

Year:  2011        PMID: 22020951     DOI: 10.4261/1305-3825.DIR.4526-11.1

Source DB:  PubMed          Journal:  Diagn Interv Radiol        ISSN: 1305-3825            Impact factor:   2.630


  5 in total

1.  Evaluations for Determination of Optimum Shields in Nuclear Medicine.

Authors:  Parvaresh R; Jalili M; Haghparast A; Khoshgard K; Eivazi M T; Ghorbani M
Journal:  J Biomed Phys Eng       Date:  2020-10-01

2.  Quality Assurance of Personal Radiation Shield for Kilovoltage Photon: A Multicentre Experience.

Authors:  Omemh Bawazeer
Journal:  Risk Manag Healthc Policy       Date:  2021-03-24

3.  Is It Safe to Use a Lead Screen During Hip Arthroscopy?

Authors:  Alexander Rahill; Leah Biffin; Camdon Fary; Alasdair G Sutherland; Phong Tran
Journal:  Arthrosc Sports Med Rehabil       Date:  2021-02-03

4.  Microbial Contamination Risk and Disinfection of Radiation Protective Garments.

Authors:  Stephen Balter; Michelle A Rodriguez; Janett A Pike; Norman J Kleiman
Journal:  Health Phys       Date:  2021-02-01       Impact factor: 2.922

5.  Radiation exposure to the eyes and thyroid during C-arm fluoroscopy-guided cervical epidural injections is far below the safety limit.

Authors:  Eun Joo Choi; Gwangcheol Go; Woong Ki Han; Pyung-Bok Lee
Journal:  Korean J Pain       Date:  2020-01-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.