Literature DB >> 31880323

An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties.

Wolfgang Lechner1, Dietmar Georg1, Hugo Palmans2,3.   

Abstract

PURPOSE: To present an analytical formalism for the in depth assessment of uncertainties of field output factors in small fields related to detector positioning based on dose profile measurements. Additionally, a procedure for the propagation of these uncertainties was developed.
METHODS: Based on the assumption that one dimensional and two dimensional second-order polynomial functions can be fitted to dose profiles of small photon beams, equations for the calculation of the expectation value, the variance, and the standard deviation were developed. The following fitting procedures of the dose profiles were considered: A one-dimensional case (1D), a quasi two-dimensional case (2Dq) based on independently measured line profiles and a full 2D case (2Df) which also considers cross-correlations in a two-dimensional dose distribution. A rectangular and a Gaussian probability density function (PDF) characterizing the probability of possible positions of the detector relative to the maximum dose were used. Uncertainty components such as the finite resolution of the scanning water phantom, the reproducibility of the determination of the position of the maximum dose, and the reproducibility of the collimator system were investigated. This formalism was tested in a 0.5 x 0.5 cm2 photon field where dose profiles were measured using a radiochromic film, a synthetic diamond detector, and an unshielded diode detector. Additionally, the dose distribution measured with the radiochromic film was convoluted with a convolution kernel mimicking the active volume of the unshielded diode.
RESULTS: Analytic expressions for the calculation of uncertainties on field output factors were found for the 1D, the 2Dq, and the 2Df case. The uncertainty of the field output factor related to the relative position of the detector to the maximum dose increased quadratically with increasing limits of possible detector positions. Analysis of the radiochromic film showed that the 2Dq case gave a more conservative assessment of the uncertainty compared to the 2Df case with a difference of < 0.1%. The 2Dq case applied to the film measurements agreed well with the same approach as was applied to the unshielded diode. The investigated uncertainty components propagated to an uncertainty of the field output factors of 0.5% and 0.4% for the synthetic diamond and the unshielded diode, respectively. Additionally, the expectation value was lower than the maximum dose. The difference was 0.4% and 0.3% for the synthetic diamond and the unshielded diode, respectively.
CONCLUSIONS: The assessment of uncertainties of field output factors related to detector positioning is feasible using the proposed formalism. The 2Dq case is applicable when using online detectors. Accurate positioning in small fields is essential for accurate dosimetry as its related uncertainty increases quadratically. The observed drop of the expectation value needs to be considered in small field dosimetry.
© 2019 The Authors. Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  positioning uncertainty; radiation oncology; small field dosimetry

Year:  2020        PMID: 31880323      PMCID: PMC7078844          DOI: 10.1002/mp.13991

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


INTRODUCTION

Modern radiotherapy treatment techniques such as intensity‐modulated radiation therapy (IMRT), volumetric modulated arc therapy (VMAT), and stereotactic radiotherapy utilize small fields for creating a dose distribution optimal for the treatment of the patient. Depending on the size and complexity of the target volume on the one hand and the employed treatment technique, as well as the sequencing algorithms on the other hand, the frequency of small fields compared to the total fluence can vary.1, 2, 3, 4, 5 Using small fields in radiation oncology requires a state of the art treatment planning system suitable for this task. For beam model creation and commissioning of such a system, special procedures on the determination of field output factors in the small fields need to be considered. In a joint effort, the IAEA and AAPM published a code of practice on small field dosimetry, recommending experimental procedures as well as detector and field specific correction factors. This code of practice also provides an assessment of the uncertainties of the correction factors themselves.6, 7 These correction factors and the associated uncertainties were based on several publications using experimental8, 9, 10, 11, 12, 13, 14, 15, 16 and Monte Carlo data.17, 18, 19, 20, 21, 22 Besides the uncertainty of the field output correction factors in small fields, additional uncertainties gain importance. One of these additional contributions to the uncertainty budget is the positioning accuracy of the detector relative to the maximum dose. In small fields, proper alignment is critical, but there are several influencing factors which are beyond control. These include the accuracy of the scanning system, the accuracy of determining the position of the maximum dose and the position of the collimator when repeating measurements. Bouchard et al performed a Monte Carlo study characterizing the impact of misalignment of the detector relative to the maximum dose. They investigated a maximum displacement of the detector relative to the maximum dose of ±1 mm assuming a rectangular probability density function (PDF) of the position of the detector.23 This study shed some light on the magnitude of this type of uncertainty of field output factors. The positioning uncertainty is influenced by the shape of the dose profile which is, amongst other parameters, dependent on the collimator design, on the design of the detector and the assumed maximum displacement of the detector. Therefore, the results presented by Bouchard et al.23 cannot be generalized to other detector types, beams, or displacement limits of the detector other than used in that study. In this work, an analytical formalism on determination of uncertainties due to misalignment of a detector relative to the maximum dose based on profile measurements on small fields is presented. Besides the development of the formalism, the main questions addressed in this work were whether the whole two‐dimensional beam profile is required or whether two independently measured line profiles are sufficient for a reasonable assessment of positioning uncertainties in small fields. Furthermore, the propagation of this type of uncertainty was explored.

MATERIALS AND METHODS

General formalism

This formalism was developed to assess the contribution of uncertainties related to the position of the detector with respect to the maximum dose, calculated as the relative standard deviation , to the combined standard uncertainty of field output factor measurements. It is considered a type B uncertainty of a single measurement of the maximum dose. The approach follows the recommendations and concepts of the JCGM.24 Based on the assumption that the dose profile of a small field near the maximum dose can be described using a second‐order polynomial function and a given PDF of the detector position relative to the maximum dose , the expectation value , the variance , and the relative standard deviation of the dose measured by the detector can be expressed as follows: where and are the positions of the maximum of the polynomial used for fitting the dose profiles. Note that for one‐dimensional problems only one‐dimensional functions and integrals were investigated. For better readability only the formulas for calculating the expectation value and the variance are shown. With these, the relative standard deviation was calculated using Eq. (3) shown in Table 5, Figs. 2 and 3.
Table 5

Relative standard deviation of the dose for selected limits of detector displacement. For these examples a symmetric limit of detector displacement () was used.

σrel/%Function a/mm
Detector0.050.250.51
Film Dx,y <0.10.150.62.6
Film Dx, Dy <0.10.160.72.6
DiodeE Dx, Dy <0.10.160.72.7
microDiamond Dx, Dy <0.10.180.72.9
Figure 2

Relative standard deviation of the dose as a function of . A two‐dimensional rectangular probability density function with symmetric limits () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com]

Figure 3

Relative standard deviation of the dose as a function of . A two‐dimensional Gaussian probability density function with symmetric standard deviations () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com]

Solutions of these formulas were used for calculating the contribution to the uncertainty of the dose due to detector positioning. Three different cases were investigated further. A simple one dimensional dose distribution, described as A quasi two‐dimensional dose distribution, where two independent polynomial functions were fitted to two independent dose profiles in and direction, expressed as follows: The two‐dimensional (2D) second‐order polynomial functionwas fitted to a two‐dimensional dose distribution and is henceforth referred to as full 2D dose distribution. The resulting expressions from the application of Eqs. (1), (2) and (3) to the different fits describing the respective dose distributions are summarized in Table 1 for different PDFs. A graphical representation of the fitted dose profile and a rectangular as well as a Gaussian PDF are shown in Fig. 1. Mathematical details on the approach and an example on the calculation of this type of uncertainty can be found in the Supporting information.
Table 1

A summary of expressions for the calculation of the expectation value and the variance for the assessment uncertainties of point dose measurements due to the relative displacement of the detector with respect to the maximum dose. Two different probability density functions (PDFs), a rectangular and a Gaussian, were applied to the one dimensional, quasi two dimensional and full two dimensional dose distribution. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively.

Probability density function
Dose distributionRectangularGaussian
1D
gx 12a (8) 12πsx2e-x22sx2 (17)
EDx p00-p1024p20+a2p203 (9) p00-p1024p20+sx2p20 (18)
VarDx 4a4p20245 (10) 2sx4p202 (19)
Quasi 2D
gx;gy 12a;12b (11) 12πsx2e-x22sx2;12πsy2e-y22sy2 (20)
EDx;EDy p00-p1024p20+a2p203;p00-p0124p02+b2p023 (12) p00-p1024p20+sx2p20;p00-p0124p02+sy2p02 (21)
VarDx;VarDy 4a4p20245;4b4p02245 (13) 2sx4p202;2sy4p022 (22)
Full 2D
gx,y 14ab (14) 12πsxsye-12x2sx2+y2sy2 (23)
EDx,y a2p20+b2p023+C (15) sx2p20+sy2p02+C (24)
VarDx,y 4a4p202+5a2b2p112+4b4p02245 (16) 2sx4p202+sx2sy2p112+2sy4p022 (25)
C p02p102-p01p10p11+p00p112+p012p20-4p00p02p20p112-4p02p20 (26)
Figure 1

Example of a one dimensional profile. On the left, the probability of the detector position with respect to the maximum dose is assumed to be equal within the limits − and . On the right, the probability of the detector position with respect to the maximum dose is assumed to follow a Gaussian distribution with the standard deviation .

A summary of expressions for the calculation of the expectation value and the variance for the assessment uncertainties of point dose measurements due to the relative displacement of the detector with respect to the maximum dose. Two different probability density functions (PDFs), a rectangular and a Gaussian, were applied to the one dimensional, quasi two dimensional and full two dimensional dose distribution. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively. Example of a one dimensional profile. On the left, the probability of the detector position with respect to the maximum dose is assumed to be equal within the limits − and . On the right, the probability of the detector position with respect to the maximum dose is assumed to follow a Gaussian distribution with the standard deviation .

Uncertainty propagation

As the dose distribution around the maximum dose is not linear, conventional Gaussian uncertainty propagation is not applicable. To investigate the propagation of uncertainties, various convolutions of rectangular and Gaussian PDFs were performed and the resulting PDFs were applied to the formalism described in Eqs. (1) and (2). The resulting equations for the one dimensional case are summarized in Tables 2 and 3.
Table 2

Formulas for the expectation value using various combinations of probability density functions (PDFs): Eq. (27) two rectangular PDFs, Eq. (28) a Gaussian and a rectangular PDF, Eq. (29) a rectangular and a Gaussian PDF and Eq. (30) two Gaussian PDFs. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively.

EDx Rectangular a1 Gaussian s1
Rectangular a2 p00-p1024p20+p20a123+a223 (27) p00-p1024p20+p20s12+a223 (28)
Gaussian s2 p00-p1024p20+p20a123+s22 (29) p00-p1024p20+p20s12+s22 (30)
Table 3

Formulas for the variance using various combinations of probability density functions (PDFs): Eq. (31) two rectangular PDFs, Eq. (32) a Gaussian and a rectangular PDF, Eq. (33) a rectangular and a Gaussian PDF and Eq. (34) two Gaussian PDFs. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively.

VarDx Rectangular a1 Gaussian s1
Rectangular a2 p2024a14+20a12a22+4a2445 (31) p2022s14+43s12a22+4a2445 (32)
Gaussian s2 p2024a1445+43s22a12+2s24 (33) 2p202s12+s222 (34)
Formulas for the expectation value using various combinations of probability density functions (PDFs): Eq. (27) two rectangular PDFs, Eq. (28) a Gaussian and a rectangular PDF, Eq. (29) a rectangular and a Gaussian PDF and Eq. (30) two Gaussian PDFs. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively. Formulas for the variance using various combinations of probability density functions (PDFs): Eq. (31) two rectangular PDFs, Eq. (32) a Gaussian and a rectangular PDF, Eq. (33) a rectangular and a Gaussian PDF and Eq. (34) two Gaussian PDFs. The parameters and are the half width of the rectangular PDF and the parameters and are the standard deviation of the Gaussian PDF in and direction, respectively.

Experimental data

All measurements were performed in a 6 MV photon beam produced by a Versa HD (Elekta AB, Stockholm, Sweden) equipped with an Agility multi‐leaf collimator (MLC) with 5 mm isocentric leaf width. The investigated nominal field size was 0.5 cm x 0.5 cm. All measurements were performed using an SSD of 90 cm and a depth of 10 cm water. The 2D dose distribution was determined using Gafchromic EBT3 films (Ashland Advanced Materials, Bridgewater, NJ, USA) in a solid water phantom (Gammex/Sun Nuclear Corporation, Melbourne, Australia). The film was scanned using an Epson V700 scanner (Epson, Nagano, Japan) 24 h after irradiation.25, 26 The resulting dose distribution was convolved with a disc‐shaped convolution kernel with 1.2 mm diameter to mimic the active volume of a PTW 60017 DiodeE detector (PTW, Freiburg, Germany). For testing the full 2D approach, the parameters of Eq. (7) were fitted to a 2.4 mm x 2.4 mm area around the maximum dose. Line profiles in x and y direction were also extracted and Eqs. (5) and (6) were fitted to the respective line profiles. To assess the differences between the full 2D approach and the quasi 2D approach using two independent one‐dimensional (1D) profiles in clinically realistic scenarios, both methods were executed and compared. One‐dimensional profiles in the x an y direction were acquired using a PTW 60019 microDiamond and a PTW 60017 DiodeE in a PTW MP3 water phantom. The microDiamond is a synthetic diamond detector. The active volume has a diameter 2.2 mm and a thickness of 1 µm. The DiodeE is a p‐type silicon diode with a diameter of 1.2 mm and a thickness of 30 µm. A step size of 0.1 mm was used for profile scanning using the online detectors. The resulting assessments of uncertainty of the DiodeE detector were compared to the film measurements mimicking measurements using the DiodeE. This allowed a comparison of uncertainties derived using a full two dimensional dose distribution with uncertainties derived from two independent line profiles. Additionally, other sources of uncertainties related to the relative position of the DiodeE to the maximum dose were investigated. The reproducibility of finding the position of the maximum dose was assessed by repeated acquisition of five dose profiles without switching the beam off between measurements, thereby keeping the MLCs and jaws fixed. The reproducibility of the position of maximum dose itself when changing the field geometry was assessed by repeated measurements of dose profiles. For these measurements the field size was set to 10 x 10 cm2 and reset to the 0.5 x 0.5 cm2 field between measurements. This process was repeated five times in one measurement session.

RESULTS

The results of the fitting procedures for the different dose distributions are summarized in Table 4. As depicted in Figs. 2 and 3, with increasing limits of the detector displacement of the rectangular PDF or the standard deviation of the Gaussian PDF also increased following a quadratic function. The examples presented in Table 5 represent prototypes of uncertainty contributions encountered in small fields due to positioning of the detector with a rectangular PDF. A limit of detector displacement of 0.05, 0.25, and 1 mm correspond to the accuracy of a state of the art scanning water phantom, a previous generation scanning water phantom and the uncertainty of the room lasers, respectively.
Table 4

Fitting parameters of the film measurements considering the full two‐dimensional dose distribution, the film measurements of using two independent one‐dimnesional (1D) dose distributions, two independent 1D dose distributions acquired with the DiodeE as well as the microDiamond.

DetectorFunction p00 p10 p20 p01 p02 p11
Film Dx,y 225.6−1.494−14.32−0.2671−11.940.2027
Film Dx, Dy 225.0−1.106−15.090.0071−12.760
DiodeE Dx, Dy 1.002−0.0183−0.0611−0.0129−0.06310
microDiamond Dx, Dy 0.998−0.0020−0.0582−0.0012−0.07470
Fitting parameters of the film measurements considering the full two‐dimensional dose distribution, the film measurements of using two independent one‐dimnesional (1D) dose distributions, two independent 1D dose distributions acquired with the DiodeE as well as the microDiamond. Relative standard deviation of the dose for selected limits of detector displacement. For these examples a symmetric limit of detector displacement () was used. Relative standard deviation of the dose as a function of . A two‐dimensional rectangular probability density function with symmetric limits () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com] Relative standard deviation of the dose as a function of . A two‐dimensional Gaussian probability density function with symmetric standard deviations () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com] The aspect of a lower expectation value than the maximum dose is shown in Figs. 4 and 5. The expectation value with respect to the maximum of the dose distribution is shown as a function of the investigated limit of detector displacement for the rectangular PDF and standard deviation of the Gaussian PDF. This ratio decreased following a quadratic function with increasing limits of detector displacement or standard deviation. This drop of the expectation value was larger than 1% for > 0.5 mm for the rectangular PDF and for > 0.3 mm for the Gaussian PDF.
Figure 4

Expectation value of the dose with respect to the maximum dose as a function of . A two‐dimensional rectangular probability density function with symmetric limits () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com]

Figure 5

Expectation value of the dose with respect to the maximum dose as a function of . A two‐dimensional Gaussian probability density function with symmetric standard deviations () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com]

Expectation value of the dose with respect to the maximum dose as a function of . A two‐dimensional rectangular probability density function with symmetric limits () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com] Expectation value of the dose with respect to the maximum dose as a function of . A two‐dimensional Gaussian probability density function with symmetric standard deviations () was used for this calculation. [Color figure can be viewed at http://wileyonlinelibrary.com] The standard uncertainty and the drop of the expectation value derived using the two line profiles, which were fitted independently to the 2D dose measured using film, agreed well with those derived using the DiodeE. A direct comparison for the two investigated detectors showed a slightly larger uncertainty for the microDiamond having a larger active volume compared to the DiodeE. Acquiring five profiles without changing the field geometry between the measurements showed a standard deviation of the position of the maximum dose relative to the detector position of 0.05 and 0.01 mm in and direction, respectively. Acquiring five profiles with changing the field size between the measurements revealed a standard deviation of the position of the maximum dose relative to the detector position of 0.04 and 0.22 mm in and direction, respectively. The square sum of these two contributions to the positioning uncertainty were 0.06 and 0.22 mm for and , respectively. The scanner resolution of the water phantom was isotropic 0.1 mm, as stated by the manufacturer. Therefore, is equal to with 0.05 mm. In this example, the total uncertainty was based on an uncertainty contribution with an underlying rectangular and a Gaussian PDF. These uncertainties propagated according to Eqs. (28) and Eq. (32) or Eqs. (29) and (33) in Tables 2 and 3 to a combined relative standard uncertainty related to the detector position of 0.5% for the microDiamond and 0.4% for the DiodeE. The ratio of the expectation value to the maximum dose was 0.996 and 0.997 for the microDiamond and the DiodeE, respectively.

DISCUSSION

In this work an analytical formalism for the assessment of uncertainties of field output factors due to positioning uncertainties in small fields was introduced. The uncertainty of 2.7% (shown in Table 5) derived for the DiodeE using a rectangular PDF with a half‐width of 1 mm was comparable to the value of 3.3% published by Bouchard et al. for the same type of detector.23 It needs to be mentioned that the 0.5 x 0.5 cm2 field used in this work had an equivalent square field size, as defined in TRS‐483,6 of 0.56 cm. With clinically realistic spatial uncertainties the relative standard deviation was 0.5% and 0.4% for the microDiamond and the DiodeE, respectively. For these examples, the uncertainty contribution related to positioning is approximately half the uncertainty of the field output correction factor of 0.8% as stated in TRS‐483 for these types of detectors and that particular field size.27 Additionally, the uncertainty of determining the field size itself is a major contributor to the total uncertainty of field output factor measurements. The reason for that is the high sensitivity of the field output factor with respect to small field sizes. This sensitivity can be in the order of 20%/mm for an 0.5 cm × 0.5 cm field. Assuming an uncertainty of only 0.1 mm of determining the field size would already cause an uncertainty related to the determination of the field size of 2%. For field sizes of 1 cm x 1 cm and larger, the positioning uncertainty becomes negligible for the investigated detectors (results not shown). A re‐assessment of the field size dependence of the positioning uncertainty needs to be conducted for other detectors, especially for detector with larger active volumes. A comparison of Eq. (13) with Eq. (16) and Eq. (22) with Eq. (25) in Table 1 shows that the only difference between the variance of two independent profiles and the full 2D expression is the term containing the parameter characterizing the cross‐correlation . Assuming all other parameters of the full 2D dose distribution are the same as of the quasi 2D dose distribution would mean that the variance of the full 2D distribution is always larger than the variance of the quasi 2D distribution using two independent 1D dose profiles. However, with parameters derived from the same EBT3 film shown in Table 1, the variance of the quasi 2D dose distributions was actually higher compared to the variance of the full 2D dose distributions. This effect was the same for the rectangular and the Gaussian PDF. Therefore, the quasi 2D approach overestimates the uncertainties compared to the full 2D approach. This might be partially caused by the fitting procedure itself. Fitting the two 1D 2nd polynomials gave a slightly larger curvature than fitting the 2D second‐order polynomial to the dose distribution. Field output measurements using online detectors require the measurement of dose profiles before the actual field output factor measurement can be conducted. These dose profiles can be used as input for the proposed quasi 2D approach and give a reasonable assessment of the uncertainties due to the relative position of the detector to the maximum dose. The slightly larger uncertainty of the microDiamond detector compared to the DiodeE might be related to the differences in the size of the active volume. This difference could also be caused by a slight change in collimator positions between measurements with the microDiamond and the DiodeE. Further investigations with a fixed collimator such as stereotactic cones are necessary. For detectors with larger volumes compared to solid state detectors, such as small ionization chambers, an increase of the uncertainty related to positioning of the detector is expected. Using this formalism, the detector and field specific uncertainties related to positioning can be calculated for any detector. The correct choice of the PDF for the particular uncertainty contribution is essential. For this work, the value of interest was the maximum dose. If the maximum dose is believed to be within a certain interval with no further information on the probability distribution of possible positions, the rectangular PDF will give the most conservative assessment. This is the case for a scanning water phantom which can drive to discrete positions only. For assessing the uncertainty related to the collimation system, repeated measurements were performed. For that type of uncertainty, a Gaussian PDF is the appropriate choice. Equally as important as the correct choice of the PDF for describing a particular uncertainty component, is the selection of the correct set of equations when propagating uncertainties. This work revealed that conventional propagation of uncertainties cannot be applied to uncertainties related to positioning uncertainties of a detector in small fields. When propagating an uncertainty component with a rectangular PDF and an uncertainty component with a Gaussian PDF, Eqs. (28) and Eq. (32) or (29) and (33) in Tables 2 and 3 need to be used, as additional terms would be missed if conventional propagation of uncertainties were used. Propagating uncertainty components described by Gaussian PDFs only can be done by adding the variances of the PDFs in the position domain and then calculating the dosimetric uncertainty using the proposed formulas as shown in Eqs. (30) and (34) in Tables 2 and 3. Other types or combinations of PDFs were not investigated in this work. The development of the uncertainty due to positioning as a function of possible detector displacements highlights the importance of accurate detector and collimator positioning. A reduction of the spatial uncertainty by half, for example, from 0.5 to 0.25 mm, will reduce the uncertainty of the dose approximately by a factor of four (see Table 5). Another observation of this work was the decrease of the expectation value of the distribution with respect to the maximum dose which is the value of interest when measuring field output factors. The implication of this is that no matter how often these measurements are repeated, the measured field output factors will in the vast majority of cases be lower than the maximum dose. For limits of detector displacement of 0.5 mm for a rectangular PDF or standard deviation of 0.3 mm for a Gaussian PDF, this drop is already 1% and increases to 4% for maximum displacements of 1 mm and a standard deviation of 0.5 mm. The example of the microDiamond detector with clinically realistic spatial uncertainties , , and shows that the expectation value would be 0.4% lower than the maximum dose. There are three approaches to encounter this issue. First, following the spirit of the JCGM,24 the inverse of the ratio of the expectation value to the maximum dose could be used as correction factor applied to the field output measurement. Second, the difference could be included as additional contribution to the uncertainty budget. Third, this difference could be neglected provided that it is sufficiently small. This problem is less pronounced for online detectors. For offline detectors, where accurate positioning is much more challenging, the uncertainty due to detector positioning will be higher. With the increasing use of small fields in radiotherapy the need of auditing these fields also increases. One way to conduct such an audit is to distribute offline detectors to participating centers and evaluate the results centrally in the auditing institution. Auditing institutions need to consider the uncertainties investigated here when providing audits for field sizes below 1 x 1 cm2. For the full 2D approach, the PDFs in and direction were assumed to be uncorrelated as strong correlations were not expected. Considering correlations of these two PDFs would result in very complicated expressions which are not useful in clinical practice, but can be used if necessary. No general analytical expressions were found for error propagation for the full 2D approach involving rectangular and Gaussian PDFs. These calculations can be performed numerically.

CONCLUSIONS

An analytical formalism for the assessment of uncertainties due to detector position relative to the maximum dose in small fields was presented. The quasi two dimensional case can be used for the assessment of positioning uncertainties in small fields using online detectors. Provided that positioning in small fields is performed according to international guidelines, the uncertainty component of field output factors related to positioning of the detector for fields with an equivalent square field size in the order of 0.6 cm can be kept below 0.5%. Special care in selection of the correct equations must be taken when propagating uncertainty components with different underlying PDFs.

CONFLICT OF INTEREST

The authors have no conflict to disclose. Data S1. Mathematical details on the approach and an example on the calculation of the uncertainty of the maximum dose due to positioning. Click here for additional data file.
  24 in total

1.  Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations.

Authors:  Gavin Cranmer-Sargison; Steve Weston; Narinder P Sidhu; David I Thwaites
Journal:  Radiother Oncol       Date:  2011-09-24       Impact factor: 6.280

2.  Correction factor measurements for multiple detectors used in small field dosimetry on the Varian Edge radiosurgery system.

Authors:  Sean Tanny; Nicholas Sperling; E Ishmael Parsai
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

3.  Application of the Exradin W1 scintillator to determine Ediode 60017 and microDiamond 60019 correction factors for relative dosimetry within small MV and FFF fields.

Authors:  T S A Underwood; B C Rowland; R Ferrand; L Vieillevigne
Journal:  Phys Med Biol       Date:  2015-08-13       Impact factor: 3.609

4.  Measurement of output factors for small photon beams.

Authors:  Otto A Sauer; Jürgen Wilbert
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

5.  Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery.

Authors:  Dávid Papp; Jan Unkelbach
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

6.  Detector density and small field dosimetry: integral versus point dose measurement schemes.

Authors:  T S A Underwood; H C Winter; M A Hill; J D Fenwick
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

7.  Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.

Authors:  Ralf Dreindl; Dietmar Georg; Markus Stock
Journal:  Z Med Phys       Date:  2013-09-20       Impact factor: 4.820

Review 8.  Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA-AAPM international Code of Practice for reference and relative dose determination.

Authors:  Hugo Palmans; Pedro Andreo; M Saiful Huq; Jan Seuntjens; Karen E Christaki; Ahmed Meghzifene
Journal:  Med Phys       Date:  2018-10-17       Impact factor: 4.071

9.  Plan averaging for multicriteria navigation of sliding window IMRT and VMAT.

Authors:  David Craft; Dávid Papp; Jan Unkelbach
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

10.  Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: a PENELOPE Monte Carlo study.

Authors:  Hamza Benmakhlouf; Josep Sempau; Pedro Andreo
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

View more
  3 in total

1.  Technical Note: Design and commissioning of a water phantom for proton dosimetry in magnetic fields.

Authors:  Hermann Fuchs; Fatima Padilla-Cabal; Andreas Hummel; Dietmar Georg
Journal:  Med Phys       Date:  2020-12-08       Impact factor: 4.071

2.  Implementation and Challenges of International Atomic Energy Agency/American Association of Physicists in Medicine TRS 483 Formalism for Field Output Factors and Involved Uncertainties Determination in Small Fields for TomoTherapy.

Authors:  Rajesh Kinhikar; Suryakant Kaushik; Chandrashekhar Tambe; Sudarshan Kadam; Shrikant Kale; Rituraj Upreti
Journal:  J Med Phys       Date:  2021-09-08

3.  A multi-institutional evaluation of small field output factor determination following the recommendations of IAEA/AAPM TRS-483.

Authors:  Wolfgang Lechner; Rodolfo Alfonso; Mehenna Arib; M Saiful Huq; Anas Ismail; Rajesh Kinhikar; José M Lárraga-Gutiérrez; Karthick Raj Mani; Nkosingiphile Maphumulo; Otto A Sauer; Shaima Shoeir; Sivalee Suriyapee; Karen Christaki
Journal:  Med Phys       Date:  2022-07-08       Impact factor: 4.506

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.