Literature DB >> 17654901

Measurement of output factors for small photon beams.

Otto A Sauer1, Jürgen Wilbert.   

Abstract

A variety of detectors and procedures for the measurement of small field output factors are discussed in the current literature. Different detectors with or without corrections are recommended. Correction factors are often derived by Monte Carlo methods, where the bias due to approximations in the model is difficult to judge. Over that, results appear to be contradictory in some cases. In this work, output factors were measured for field sizes from 4 mm up to 180 mm side length with different detectors. A simple linear correction for the energy response of solid state detectors is proposed. This led to identical values down to 8 mm field size, as long as the size of the detector is small against the field size. The correction was of the order of a few percent. For a shielded silicon diode it was well below 1%. A physically meaningful function is proposed in order to calculate output factors for arbitrary field sizes with high accuracy.

Entities:  

Mesh:

Year:  2007        PMID: 17654901     DOI: 10.1118/1.2734383

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

2.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

3.  Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter resolution.

Authors:  Muhammad Ramish Ashraf; Petr Bruza; Brian W Pogue; Nathan Nelson; Benjamin B Williams; Lesley A Jarvis; David J Gladstone
Journal:  Med Phys       Date:  2019-10-04       Impact factor: 4.071

4.  Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam.

Authors:  Kyohei Fukata; Satoru Sugimoto; Chie Kurokawa; Akito Saito; Tatsuya Inoue; Keisuke Sasai
Journal:  Radiol Phys Technol       Date:  2018-04-04

5.  Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system.

Authors:  Corey Clift; Andrew Thomas; John Adamovics; Zheng Chang; Indra Das; Mark Oldham
Journal:  Phys Med Biol       Date:  2010-02-04       Impact factor: 3.609

6.  The influence of small field output factors simulated uncertainties on the calculated dose in VMAT plans for brain metastases: a multicentre study.

Authors:  Stefania Clemente; Maria Daniela Falco; Elisabetta Cagni; Cinzia Talamonti; Mafalda Boccia; Eva Gino; Elena Lorenzini; Federica Rosica; Serenella Russo; Alessandro Alparone; Daniele Zefiro; Christian Fiandra
Journal:  Br J Radiol       Date:  2021-01-22       Impact factor: 3.039

7.  Repeatability of Small Field Output Factor Measurements with Various Detectors.

Authors:  Zakithi Lungile Mpumelelo Msimang; Debbie van der Merwe; Nkosingiphile Maphumulo
Journal:  J Med Phys       Date:  2021-05-05

8.  Small fields measurements with radiochromic films.

Authors:  Antonio Gonzalez-Lopez; Juan-Antonio Vera-Sanchez; Jose-Domingo Lago-Martin
Journal:  J Med Phys       Date:  2015 Apr-Jun

9.  Dosimetric characterization of Elekta stereotactic cones.

Authors:  Egor Borzov; Alexander Nevelsky; Raquel Bar-Deroma; Itzhak Orion
Journal:  J Appl Clin Med Phys       Date:  2017-12-20       Impact factor: 2.102

10.  Dosimetry for small fields in stereotactic radiosurgery using gafchromic MD-V2-55 film, TLD-100 and alanine dosimeters.

Authors:  Guerda Massillon-J L; Diego Cueva-Prócel; Porfirio Díaz-Aguirre; Miguel Rodríguez-Ponce; Flor Herrera-Martínez
Journal:  PLoS One       Date:  2013-05-09       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.