Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.

Ralf Dreindl1, Dietmar Georg2, Markus Stock3.   

Abstract

Gafchromic® EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6 MV, 10 MV and 18 MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB® was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations.
Copyright © 2013. Published by Elsevier GmbH.

Entities:  

Keywords:  EBT2; EBT3; Eigenschaften; Film dosimetry; Filmdosimetrie; Langzeit-Effekte; characterization; long-term effects

Mesh:

Year:  2013        PMID: 24055395     DOI: 10.1016/j.zemedi.2013.08.002

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  27 in total

1.  Correction of lateral response artifacts from flatbed scanners for dual-channel radiochromic film dosimetry.

Authors:  Yuichi Akino; Hiroya Shiomi; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Takero Hirata; Hirokazu Mizuno; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2021-03-10       Impact factor: 2.724

2.  Measurement of percentage dose at the surface for a 6 MV photon beam.

Authors:  O O Galván De la Cruz; M A Rodríguez-Ávila; T Rivera-Montalvo; O A García Garduño
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-18

3.  In vivo surface dose measurement using GafChromic film dosimetry in breast cancer radiotherapy: comparison of 7-field IMRT, tangential IMRT and tangential 3D-CRT.

Authors:  Volker Rudat; Alaa Nour; Abdul Aziz Alaradi; Adel Mohamed; Saleh Altuwaijri
Journal:  Radiat Oncol       Date:  2014-07-15       Impact factor: 3.481

4.  Sum signal dosimetry: A new approach for high dose quality assurance with Gafchromic EBT3.

Authors:  Davide Cusumano; Maria Luisa Fumagalli; Francesco Ghielmetti; Linda Rossi; Giuliano Grossi; Raffaella Lanzarotti; Laura Fariselli; Elena De Martin
Journal:  J Appl Clin Med Phys       Date:  2017-02-02       Impact factor: 2.102

5.  The combination of the error correction methods of GAFCHROMIC EBT3 film.

Authors:  Yinghui Li; Lixin Chen; Jinhan Zhu; Xiaowei Liu
Journal:  PLoS One       Date:  2017-07-27       Impact factor: 3.240

6.  Robust optimization of VMAT for lung cancer: Dosimetric implications of motion compensation techniques.

Authors:  Ben R Archibald-Heeren; Mikel V Byrne; Yunfei Hu; Meng Cai; Yang Wang
Journal:  J Appl Clin Med Phys       Date:  2017-08-08       Impact factor: 2.102

7.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

8.  Methodology for radiochromic film analysis using FilmQA Pro and ImageJ.

Authors:  Michelle E Howard; Michael G Herman; Michael P Grams
Journal:  PLoS One       Date:  2020-05-21       Impact factor: 3.240

9.  Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.

Authors:  Elsa Y León Marroquin; José A Herrera González; Miguel A Camacho López; José E Villarreal Barajas; Olivia A García-Garduño
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams.

Authors:  Nashrulhaq Tagiling; Raizulnasuha Ab Rashid; Siti Nur Amirah Azhan; Norhayati Dollah; Moshi Geso; Wan Nordiana Rahman
Journal:  Heliyon       Date:  2018-10-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.