Application of the Exradin W1 scintillator to determine Ediode 60017 and microDiamond 60019 correction factors for relative dosimetry within small MV and FFF fields.

T S A Underwood1, B C Rowland, R Ferrand, L Vieillevigne.   

Abstract

In this work we use EBT3 film measurements at 10 MV to demonstrate the suitability of the Exradin W1 (plastic scintillator) for relative dosimetry within small photon fields. We then use the Exradin W1 to measure the small field correction factors required by two other detectors: the PTW unshielded Ediode 60017 and the PTW microDiamond 60019. We consider on-axis correction-factors for small fields collimated using MLCs for four different TrueBeam energies: 6 FFF, 6 MV, 10 FFF and 10 MV. We also investigate percentage depth dose and lateral profile perturbations. In addition to high-density effects from its silicon sensitive region, the Ediode exhibited a dose-rate dependence and its known over-response to low energy scatter was found to be greater for 6 FFF than 6 MV. For clinical centres without access to a W1 scintillator, we recommend the microDiamond over the Ediode and suggest that 'limits of usability', field sizes below which a detector introduces unacceptable errors, can form a practical alternative to small-field correction factors. For a dosimetric tolerance of 2% on-axis, the microDiamond might be utilised down to 10 mm and 15 mm field sizes for 6 MV and 10 MV, respectively.

Entities:  

Mesh:

Year:  2015        PMID: 26271097     DOI: 10.1088/0031-9155/60/17/6669

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  15 in total

1.  Dosimetric characterization and behaviour in small X-ray fields of a microchamber and a plastic scintillator detector.

Authors:  Massimo Pasquino; Claudia Cutaia; Lorenzo Radici; Serena Valzano; Eva Gino; Carlo Cavedon; Michele Stasi
Journal:  Br J Radiol       Date:  2016-11-09       Impact factor: 3.039

2.  Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

Authors:  Yujiao Qin; Hualiang Zhong; Ning Wen; Karen Snyder; Yimei Huang; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

3.  Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector.

Authors:  Suresh H Chaudhari; Rishabh Dobhal; Rajesh A Kinhikar; Sudarshan S Kadam; Deepak D Deshpande
Journal:  J Med Phys       Date:  2017 Jan-Mar

4.  Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

Authors:  Madelaine K Tyler; Paul Z Y Liu; Christopher Lee; David R McKenzie; Natalka Suchowerska
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

5.  A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2018-12-24       Impact factor: 4.071

6.  Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS.

Authors:  Jesse D Snyder; Rodney J Sullivan; Xingen Wu; Elizabeth L Covington; Richard A Popple
Journal:  J Appl Clin Med Phys       Date:  2019-09       Impact factor: 2.102

7.  The role of radiation-induced charge imbalance on the dose-response of a commercial synthetic diamond detector in small field dosimetry.

Authors:  Hui Khee Looe; Daniela Poppinga; Rafael Kranzer; Isabel Büsing; Tuba Tekin; Ann-Britt Ulrichs; Björn Delfs; Dennis Vogt; Jan Würfel; Björn Poppe
Journal:  Med Phys       Date:  2019-05-02       Impact factor: 4.071

8.  Evaluation of the PTW microDiamond in edge-on orientation for dosimetry in small fields.

Authors:  Owen J Brace; Sultan F Alhujaili; Jason R Paino; Duncan J Butler; Dean Wilkinson; Brad M Oborn; Anatoly B Rosenfeld; Michael L F Lerch; Marco Petasecca; Jeremy A Davis
Journal:  J Appl Clin Med Phys       Date:  2020-05-22       Impact factor: 2.102

9.  High spatial resolution inorganic scintillator detector for high-energy X-ray beam at small field irradiation.

Authors:  Sree Bash Chandra Debnath; Carole Fauquet; Agnes Tallet; Anthony Goncalves; Sébastien Lavandier; Franck Jandard; Didier Tonneau; Julien Darreon
Journal:  Med Phys       Date:  2020-01-23       Impact factor: 4.071

10.  An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties.

Authors:  Wolfgang Lechner; Dietmar Georg; Hugo Palmans
Journal:  Med Phys       Date:  2020-01-26       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.