Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: a PENELOPE Monte Carlo study.

Hamza Benmakhlouf1, Josep Sempau2, Pedro Andreo3.   

Abstract

PURPOSE: To determine detector-specific output correction factors,[Formula: see text], in 6 MV small photon beams for air and liquid ionization chambers, silicon diodes, and diamond detectors from two manufacturers.
METHODS: Field output factors, defined according to the international formalism published byAlfonso et al. [Med. Phys. 35, 5179-5186 (2008)], relate the dosimetry of small photon beams to that of the machine-specific reference field; they include a correction to measured ratios of detector readings, conventionally used as output factors in broad beams. Output correction factors were calculated with the PENELOPE Monte Carlo (MC) system with a statistical uncertainty (type-A) of 0.15% or lower. The geometries of the detectors were coded using blueprints provided by the manufacturers, and phase-space files for field sizes between 0.5 × 0.5 cm(2) and 10 × 10 cm(2) from a Varian Clinac iX 6 MV linac used as sources. The output correction factors were determined scoring the absorbed dose within a detector and to a small water volume in the absence of the detector, both at a depth of 10 cm, for each small field and for the reference beam of 10 × 10 cm(2).
RESULTS: The Monte Carlo calculated output correction factors for the liquid ionization chamber and the diamond detector were within about ± 1% of unity even for the smallest field sizes. Corrections were found to be significant for small air ionization chambers due to their cavity dimensions, as expected. The correction factors for silicon diodes varied with the detector type (shielded or unshielded), confirming the findings by other authors; different corrections for the detectors from the two manufacturers were obtained. The differences in the calculated factors for the various detectors were analyzed thoroughly and whenever possible the results were compared to published data, often calculated for different accelerators and using the EGSnrc MC system. The differences were used to estimate a type-B uncertainty for the correction factors. Together with the type-A uncertainty from the Monte Carlo calculations, an estimation of the combined standard uncertainty was made, assigned to the mean correction factors from various estimates.
CONCLUSIONS: The present work provides a consistent and specific set of data for the output correction factors of a broad set of detectors in a Varian Clinac iX 6 MV accelerator and contributes to improving the understanding of the physics of small photon beams. The correction factors cannot in general be neglected for any detector and, as expected, their magnitude increases with decreasing field size. Due to the reduced number of clinical accelerator types currently available, it is suggested that detector output correction factors be given specifically for linac models and field sizes, rather than for a beam quality specifier that necessarily varies with the accelerator type and field size due to the different electron spot dimensions and photon collimation systems used by each accelerator model.
© 2014 American Association of Physicists in Medicine.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24694131     DOI: 10.1118/1.4868695

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Radiological tissue equivalence of deformable silicone-based chemical radiation dosimeters (FlexyDos3D).

Authors:  Yi Du; Ruoxi Wang; Meijiao Wang; Haizhen Yue; Yibao Zhang; Hao Wu; Weihu Wang
Journal:  J Appl Clin Med Phys       Date:  2019-06-11       Impact factor: 2.102

2.  Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam.

Authors:  Kyohei Fukata; Satoru Sugimoto; Chie Kurokawa; Akito Saito; Tatsuya Inoue; Keisuke Sasai
Journal:  Radiol Phys Technol       Date:  2018-04-04

3.  Variation of kQclin,Qmsr (fclin,fmsr) for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio.

Authors:  Paolo Francescon; Sam Beddar; Ninfa Satariano; Indra J Das
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

4.  Characterization of a microSilicon diode detector for small-field photon beam dosimetry.

Authors:  Yuichi Akino; Masateru Fujiwara; Keita Okamura; Hiroya Shiomi; Hirokazu Mizuno; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

5.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

Review 6.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

7.  Beam commissioning of the first clinical biology-guided radiotherapy system.

Authors:  Bin Han; Dante Capaldi; Nataliya Kovalchuk; Eric Simiele; John White; Daniel Zaks; Lei Xing; Murat Surucu
Journal:  J Appl Clin Med Phys       Date:  2022-04-28       Impact factor: 2.243

8.  Dosimetric characterization of Elekta stereotactic cones.

Authors:  Egor Borzov; Alexander Nevelsky; Raquel Bar-Deroma; Itzhak Orion
Journal:  J Appl Clin Med Phys       Date:  2017-12-20       Impact factor: 2.102

9.  Survey of 5 mm small-field output factor measurements in Australia.

Authors:  Christopher P Oliver; Duncan J Butler; Viliami Takau; Ivan Williams
Journal:  J Appl Clin Med Phys       Date:  2018-01-25       Impact factor: 2.102

10.  Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Authors:  Yuichi Akino; Archana Gautam; Len Coutinho; Jan Würfel; Indra J Das
Journal:  J Radiat Res       Date:  2015-08-12       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.