Wei Mao1, Haofei Ni1, Linli Li1, Yiqun He1, Xujun Chen1, Han Tang1, Youhai Dong1.
Abstract
OBJECTIVES: Different criteria for assessing the reduction quality of trochanteric fractures have been reported. The Baumgaertner reduction quality criteria (BRQC) are relatively common and the Chang reduction quality criteria (CRQC) are relatively new. The objectives of the current study were to compare the reliability of the BRQC and CRQC in predicting mechanical complications and to investigate the clinical implications of the CRQC.Entities:
Keywords: Baumgaertner reduction quality criteria; Chang reduction quality criteria; Mechanical complications; Trochanteric fractures
Year: 2019 PMID: 31728190 PMCID: PMC6825041 DOI: 10.1302/2046-3758.810.BJR-2019-0032.R1
Source DB: PubMed Journal: Bone Joint Res ISSN: 2046-3758 Impact factor: 5.853
Baumgaertner reduction quality criteria
| Criteria |
|---|
| a. Anteroposterior view: normal or slight valgus neck-shaft angle[ |
| b. Lateral view: less than 20° of angulation |
| a. Anteroposterior view: less than 4 mm of displacement of any fragments |
| b. Lateral view: less than 4 mm of displacement of any fragments |
| Good: both criteria met |
| Acceptable: only one criterion met |
| Poor: neither criterion met |
Slight valgus means a valgus of no more than 10°[4,15]
Fig. 1Radiograph showing positive medial cortex support, which is defined as a situation where the lower edge of the medial cortex of the femoral head–neck fragment is located medially to the upper edge of the medial cortex of the femoral shaft, with the displacement less than one cortex thickness.
Fig. 2Radiograph showing negative medial cortical support, which is defined as a situation where the lower edge of the medial cortex of the femoral head–neck fragment is located laterally to the upper edge of the medial cortex of the femoral shaft, regardless of the distance of displacement.
Chang reduction quality criteria
| Item | Score |
|---|---|
| I. Alignment | |
| a. Anteroposterior view: normal or slight valgus neck-shaft angle[ | 1 |
| b. Lateral view: less than 20° of angulation | 1 |
| II. Displacement | |
| a. Anteroposterior view: neutral or positive medial cortical support[ | 1 |
| b. Lateral view: smooth anterior cortical contact[ | 1 |
| Reduction quality | |
| Excellent | 4 |
| Acceptable | 2 or 3 |
| Poor | 0 or 1 |
Slight valgus means a valgus of no more than 10°[4,15]
Neutral medial cortical support is shown in Supplementary Figure a
The displacement is less than half of the cortex thickness
Types and numbers of mechanical complications; 29 mechanical complications occurred in 26 patients
| Mechanical complication | n |
|---|---|
| Implant failure[ | 2 |
| Varus displacement | 19 |
| Excessive lateral migration of the blade | 8 |
Including one cut-out and one loss of fixation, both with nonunion
Reliability between the two observers for different variables
| Variable | ICC or κ | 95% CI | Reliability |
|---|---|---|---|
| Tip-apex distance, mean measures ICC | 0.888 | 0.841 to 0.921 | Almost perfect |
| AO/OTA classification, κ | 0.635 | 0.491 to 0.749 | Excellent |
| Blade position, κ | 0.420 | 0.295 to 0.544 | Moderate |
| Baumgaertner reduction quality criteria, κ | 0.589 | 0.458 to 0.705 | Moderate |
| Chang reduction quality criteria, κ | 0.731 | 0.615 to 0.833 | Excellent |
| Singh Osteoporosis Index, κ | 0.339 | 0.213 to 0.457 | Poor |
ICC, intraclass correlation coefficient; κ, kappa coefficient; CI, confidence interval; AO/OTA, AO Foundation and Orthopaedic Trauma Association
Univariate analysis of patient demographics and clinical characteristics
| Characteristic | Without mechanical complications (n = 101) | With mechanical complications (n = 26) | Crude OR (95% CI) | p-value |
|---|---|---|---|---|
| Mean age, yrs (range) | 69.7 (27 to 94) | 74.0 (54 to 91) | 1.03 (0.99 to 1.06) | 0.068[ |
| Male, n (%) | 43 ( | 11 ( | 0.99 (0.41 to 2.37) | 1.000[ |
| Left side, n (%) | 55 ( | 13 ( | 0.84 (0.35 to 1.98) | 0.826[ |
| Mean TAD, mm (range) | 22.9 (7.2 to 47.9) | 26.7 (16.0 to 48.2) | 1.07 (1.01 to 1.13) | 0.020[ |
| ASA classification, n (%) | 0.309[ | |||
| I | 25 ( | 2 ( | N/A | |
| II | 47 ( | 17 ( | N/A | |
| III | 27 ( | 6 ( | N/A | |
| IV | 2 ( | 1 ( | N/A | |
| AO/OTA classification, n (%) | 0.094[ | |||
| A1 | 51 ( | 11 ( | N/A | |
| A2 | 44 ( | 9 ( | N/A | |
| A3 | 6 ( | 6 ( | N/A | |
| Blade position | § | § | N/A | < 0.001[ |
| BRQC | § | § | N/A | < 0.001[ |
| CRQC | § | § | N/A | < 0.001[ |
| SOI, n (%) | 0.167[ | |||
| I | 2 ( | 0 ( | N/A | |
| II | 2 ( | 1 ( | N/A | |
| III | 14 ( | 8 ( | N/A | |
| IV | 52 ( | 11 ( | N/A | |
| V | 22 ( | 6 ( | N/A | |
| VI | 9 ( | 0 ( | N/A |
Student’s t-test
Chi-squared test
Statistically significant
§The distribution of values is indicated in Figure 3 and Figure 4
OR, odds ratio; CI, confidence interval; TAD, tip-apex distance; ASA, American Society of Anesthesiologists; N/A, not applicable; AO/OTA, AO Foundation and Orthopaedic Trauma Association; BRQC, Baumgaertner reduction quality criteria; CRQC, Chang reduction quality criteria; SOI, Singh Osteoporosis Index
Fig. 3The blade position according to the Cleveland zone. The first number represents the group with mechanical complications and the second number represents the group without mechanical complications. Percentages demonstrate the proportion of mechanical complications associated with each Cleveland zone.
Fig. 4Distribution of outcomes according to Baumgaertner reduction quality criteria (BRQC) and Chang reduction quality criteria (CRQC). The rates of mechanical complications were 8.6% (BRQC-Good), 21.7% (BRQC-Acceptable), 88.9% (BRQC-Poor), 3.8% (CRQC-Excellent), 30.6% (CRQC-Acceptable), and 92.3% (CRQC-Poor).
The results of multivariate logistic regression analysis
| Characteristic | Adjusted OR (95% CI) | p-value |
|---|---|---|
| Tip-apex distance, mm | 1.088 (1.007 to 1.175) | 0.025[ |
| Blade position | N/A | 0.201 |
| Baumgaertner reduction quality criteria | N/A | 0.505 |
| < 0.001[ | ||
| Excellent | 0.003 (0.000 to 0.030) | < 0.001[ |
| Acceptable | 0.026 (0.003 to 0.243) | 0.001[ |
Statistically significant
The poor reduction was the reference category
OR, odds ratio; CI, confidence interval; N/A, not applicable
Fig. 5The receiver operating characteristic (ROC) curves comparing performance of Baumgaertner reduction quality criteria (BRQC) and Chang reduction quality criteria (CRQC) in predicting mechanical complications. The area under the curve (AUC) of the BRQC was 0.74 (95% confidence interval (CI) 0.65 to 0.81). The AUC of the CRQC was 0.87 (95% CI 0.80 to 0.92). The difference between the AUCs was significant (p = 0.012; z-test).