| Literature DB >> 31707918 |
Michele M Kim1, Douglas Bollinger1, Chris Kennedy1, Wei Zou1, Ryan Scheuermann1, Boon-Keng Kevin Teo1, James M Metz1, Lei Dong1, Taoran Li1.
Abstract
The Halcyon is Varian's latest linear accelerator that offers a single 6X flattening-filter-free beam with a jawless design that features a new dual layer multileaf collimator system with faster speed and reduced transmission. Dosimetric characteristics of the dual layer multileaf collimator system including transmission, dosimetric leaf gap, and tongue and groove effects were measured. Ionization chambers, diode arrays, and an electronic portal imaging device were used to measure various multileaf collimator characteristics. Transmission through both multileaf collimator banks was found to be 0.008%, while the distal and proximal banks alone had transmission values of 0.4%. The penumbra was slightly sharper for fields using only the distal multileaf collimator bank but found to be largely independent of leaf position with values between 2.7 to 3.0 mm at dmax for the combined multileaf collimator banks. The dosimetric leaf gap was measured for the proximal and distal multileaf collimator banks both individually and together and found to have values of -0.216 mm, -0.225 mm, and 0.964 mm, respectively. Measurements of dosimetric leaf gap at the leaf edge and midline were also performed. Tongue and groove effects were investigated with both the electronic portal imaging device and a 2-dimensional array of diodes.Entities:
Keywords: Halcyon; MLC; dosimetric leaf gap; dynamic beam flattening; multileaf collimator; tongue and groove
Mesh:
Year: 2019 PMID: 31707918 PMCID: PMC6843729 DOI: 10.1177/1533033819883641
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Transmission through both MLC banks, distal bank, and proximal bank measured with the portal dosimeter and a CC13 ion chamber. The value modeled in the TPS (0.47%) is shown with square symbols. MLC indicates multileaf collimator.
Dosimetric Leaf Gap for the Proximal, Distal, and Combined MLCs.
| MLC Bank | Measurement Location | Dosimetric Leaf Gap (mm) | |
|---|---|---|---|
| MapCheck2 | CC13 | ||
| Proximal | Leaf midline (Y = 0) | 0.053 | −0.216 |
| Leaf edge (Y = 5 mm) | −0.251 | ||
| Distal | Leaf edge (Y = 0) | −0.294 | −0.225 |
| Leaf midline (Y = 5 mm) | 0.190 | ||
| Combined | Y = 0 | 1.054 | 0.964 |
| Y = 5 mm | 1.108 | ||
Abbreviation: MLC, multileaf collimator.
Penumbra (mm) Measured for a 2 × 10 cm2 Field at Various Leaf Positions at Depths of 10 cm and dmax Using the Proximal, Distal, and Combined MLC Banks.
| Leaf Position (cm) | 13 | 7 | 1 | −5 | −11 |
|---|---|---|---|---|---|
| Proximal d10 | 5.0 | 4.5 | 4.6 | 4.7 | 5.0 |
| Distal d10 | 4.1 | 3.7 | 3.6 | 3.7 | 4.1 |
| Combined d10 | 3.7 | 3.6 | 3.7 | 3.9 | 4.0 |
| Proximal dmax | 3.4 | 3.2 | 3.3 | 3.5 | 3.7 |
| Distal dmax | 2.6 | 2.5 | 2.6 | 2.7 | 2.8 |
| Combined dmax | 2.7 | 2.7 | 2.7 | 2.8 | 3.0 |
Abbreviation: MLC, multileaf collimator.
Figure 2.Flattened and nonflattened field profiles in the Y direction measured with MapCheck2 for various field sizes.
Figure 3.The tongue and groove effect for both proximal and distal banks of MLCs measured with (A) MapCheck2 compared to TPS and (B) portal dosimetry compared to TPS. MLCs indicates multileaf collimators; TPS, treatment planning system.
Figure 4.Rounded leaf end schematic.
Figure 5.Example of tongue and groove effect with collimator angles set to 0. The dose color wash for pelvis treatment is shown for the (A) coronal and (B) sagittal views along with the (C) dose profile to highlight the tongue and groove effect.