| Literature DB >> 30284378 |
Vida Karimnia1, Matthew D Belley1,2,3, Robert Rodgers1,2,3, Michael Price1,2,3.
Abstract
Linear accelerator (linac) commissioning and quality assurance measurements are time-consuming tasks that often require a water tank scanning system to acquire profile scans for full characterization of dosimetric beam properties. To increase efficiency, a method is demonstrated to acquire variable resolution, photon beam profile data using a commercially available ion chamber array (0.5 cm detector spacing). Field sizes of 2 × 2, 5 × 5, 10 × 10, and 15 × 15 cm2 were acquired at depths in solid water of dmax , 5 cm, and 10 cm; additionally, beam profiles for field sizes of 25 × 25 and 40 × 40 cm2 were acquired at 5 cm depth in solid water at x-ray energies of 6 and 23 MV. 1D composite profiles were generated by combining discrete point measurements made at multiple couch positions. The 1D composite profile dataset was evaluated against a commissioning dataset acquired with a 3D water tank scan system utilizing (a) 0.125 cc ion chamber for 5 × 5, 10 × 10, 15 × 15, 25 × 25, and 40 × 40 field sizes and (b) a solid state detector for 2 × 2 cm2 field size. The two datasets were compared to the gamma criteria at 1%/1 mm and 2%/2 mm tolerance. Almost all pass rates exceeded 95% at 2%/2 mm except for the 6 MV 2 × 2 cm2 field size at dmax . Pass rates at 1%/1 mm ranged from 51% to 99%, with an average pass rate of 82%. A fourfold reduction in MU was achieved for scans larger than 15 × 15 cm2 using this method compared to the water tank scans. Further, dynamic wedge measurements acquired with the ion chamber array showed reasonable agreement with the treatment planning system. This method opens up new possibilities for rapid acquisition of variable resolution 2D-3D dosimetric data mitigating the need for acquiring all scan data with in-water measurements.Entities:
Keywords: gamma analysis; ion chamber array; photon beam profiles; quality assurance
Mesh:
Year: 2018 PMID: 30284378 PMCID: PMC6236837 DOI: 10.1002/acm2.12466
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experiment setup for 5 cm depth crossline acquired with couch shifts of (a) 0 cm, (b) 0.1 cm, and (c) 0.2 cm.
Field sizes and SSD values used for data acquisition
| Square field sizes (cm2) | Depths (cm) | SSD (cm) | MU | Energy (MV) | Solid water, square size (cm2) | Couch shift increment (cm) | Ic‐Profiler™ array axis |
|---|---|---|---|---|---|---|---|
| 5 |
| 100 | 50 | 6, 23 | 30 | 0.1 |
|
| 10 |
| 100 | 50 | 6, 23 | 30 | 0.1 |
|
| 10 | 5 (EDW) | 100 | 50 | 6 | 30 | 0.1 |
|
| 15 |
| 100 | 50 | 6, 23 | 30 | 0.1 |
|
| 2 |
| 100 | 50 | 6 | 30 | 0.05 |
|
| 25 | 5 | 100 | 50 | 6, 23 | 30 | 0.1 |
|
| 40 | 5 | 70 | 50 | 6 | 40 | 0.1 |
|
Sample calculation of the coordinates based on the measurements when the couch is oriented normally
| Measurement number |
|
| Δ |
|
|
| Δ |
| Couch/support angle (deg) |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 0 | 0 | 0 | 0 | 147.5 | 147.5 | 0 | 0 | 0 |
| 2 | 0 | 0.1 | 0.1 | −0.1 | 147.5 | 147.5 | 0 | 0 | 0 |
| 3 | 0 | 0.2 | 0.2 | −0.2 | 147.5 | 147.5 | 0 | 0 | 0 |
| 4 | 0 | 0.3 | 0.3 | −0.3 | 147.5 | 147.5 | 0 | 0 | 0 |
| 5 | 0 | 0.4 | 0.4 | −0.4 | 147.5 | 147.5 | 0 | 0 | 0 |
| 6 | 0 | 0 | 0 | 0 | 147.5 | 147.5 | 0 | 0 | 0 |
| 7 | 0 | 0 | 0 | 0 | 147.5 | 147.6 | 0.1 | −0.1 | 0 |
| 8 | 0 | 0 | 0 | 0 | 147.5 | 147.7 | 0.2 | −0.2 | 0 |
| 9 | 0 | 0 | 0 | 0 | 147.5 | 147.8 | 0.3 | −0.3 | 0 |
| 10 | 0 | 0 | 0 | 0 | 147.5 | 147.9 | 0.4 | −0.4 | 0 |
Figure 2(a) Couch oriented normally (0°) and (b) Couch rotated 90° (agrees with IEC 61217 coordinate system).
Figure 3Graph of photon beam profile data at 2%/2 mm criteria. (a) 6 MV, 10 × 10 cm2 FS, 10 cm depth, gamma pass rate = 99.40% (b) 23 MV, 10 × 10 cm2 FS, 10 cm depth, gamma pass rate = 100% (c) 6 MV, 25 × 25 cm2 FS, 5 cm depth, gamma pass rate = 88.30% (d) 23 MV, 25 × 25 cm2 FS, 5 cm depth, gamma pass rate = 89.37% (e) 6 MV, 2 × 2 cm2 FS, 1.5 cm depth, gamma pass rate = 99.20% (f) 6 MV, 40 × 40 cm2 FS, 5 cm depth, gamma pass rate = 78.60%.
Percentage of points satisfying gamma with 1%/1 mm and 2%/2 mm criteria scans. “–” indicates data were not collected. The data for 2 × 2 cm2 are after convolution
| Square field size (cm2) | Depth (cm) | Shift (cm) | (2%/2 mm) gamma pass rate | (1%/1 mm) gamma pass rate | ||
|---|---|---|---|---|---|---|
| 6 MV | 23 MV | 6 MV | 23 MV | |||
| 2 | 1.5 | 0.05 | 94.2 | – | 51.5 | – |
| 5 | 0.05 | 100 | – | 65.7 | – | |
| 10 | 0.05 | 100 | – | 63.4 | – | |
| 5 | 1.5 | 0.1 | 100 | – | 96.0 | – |
| 3.6 | 0.1 | – | 100 | – | 79.9 | |
| 5 | 0.1 | 100 | 100 | 90.6 | 78.9 | |
| 10 | 0.1 | 99.4 | 100 | 97.8 | 79.3 | |
| 10 | 1.5 | 0.1 | 100 | – | 99.38 | – |
| 3.6 | 0.1 | – | 100 | – | 95.7 | |
| 5 | 0.1 | 100 | 100 | 87.5 | 95.1 | |
| 10 | 0.1 | 99.4 | 100 | 69.9 | 95.8 | |
| 15 | 1.5 | 0.1 | 100 | – | 97.77 | – |
| 3.6 | 0.1 | – | 100 | – | 96.08 | |
| 5 | 0.1 | 99.30 | 100 | 75.48 | 90.78 | |
| 10 | 0.1 | 95.99 | 100 | 66.5 | 96.42 | |
| 25 | 5 | 0.1 | 88.30 | 89.37 | 72.85 | 51.79 |
| 40 (SSD = 70 cm) | 5 | 0.1 | 76.13 | – | 63.44 | – |
Dosimetric agreement for small field 3D Scanner tank data (Edge detector) convolved to match measured IC‐Profiler™ data. E = 6 MV, inline scans, Composite Point Spacing = 0.05 cm
| Square field size (cm2) | Depth (cm) | (1%/1 mm) gamma pass rate | (2%/2 mm) gamma PASS rate | ||
|---|---|---|---|---|---|
| No‐convolution (%) | Convolution (%) | No‐convolution (%) | Convolution (%) | ||
| 2 | 1.5 | 51.1 | 51.5 | 86.5 | 94.2 |
| 2 | 5 | 65.0 | 65.7 | 99.4 | 100 |
| 2 | 10 | 59.7 | 63.4 | 99.5 | 100 |
Figure 4Dose difference between the convolution and non‐convolution method for 6 MV, 2 × 2 cm2 field size, 10 cm depth.
Dosimetric agreement of 2 × 2 cm2, 3D Scanner tank data (Edge detector) convolved to match IC‐Profiler™ data at various composite point spacing. E = 6 MV, inline scans, 5 cm depth
| Composite point spacing (cm) | Sampled points increase factor | Gamma pass rate (%) | |
|---|---|---|---|
| 1%/1 mm | 2%/2 mm | ||
| 0.50 (no shifts) | 1.0× | 22.6 | 34.3 |
| 0.20 | 2.5× | 40.3 | 70.2 |
| 0.15 | 3.3× | 58.6 | 75.1 |
| 0.10 | 5.0× | 62.4 | 91.2 |
| 0.05 | 10.0× | 65.7 | 100 |
Figure 5Comparison between 60° EDW for 6 MV, 100 cm SSD, 5 cm depth, 10 × 10 cm2 field size using IC‐Profiler™ composite (composite point spacing = 0.1 cm) and TPS Exported Data with Y2 jaw, and voxel size 0.12 cm. Normalization was only performed on the composite profile.