Literature DB >> 31531906

Enhance the effectiveness of clinical laboratory critical values initiative notification by implementing a closed-loop system: A five-year retrospective observational study.

Runqing Li1, Tengjiao Wang1, Lijun Gong1, Jingxiao Dong1, Nan Xiao1, Xiaohuan Yang1, Dong Zhu1, Zhipeng Zhao1.   

Abstract

BACKGROUND: Accurate and timely clinical laboratory critical values notification are crucial steps in supporting effective clinical decision making, thereby improving patient safety.
METHODS: A closed-loop laboratory critical value notification system was developed by a multidisciplinary team of clinicians, laboratorians, administrators, and information technology experts. All the laboratory critical values that occurred at Beijing Tsinghua Changgung Hospital (BTCH, Beijing, China) from 2015 to 2019 were analyzed and studied retrospectively.
RESULTS: The total number (ratio) of institutional laboratory critical values to all reported items at BTCH from 2015 to 2019 was 38 020/7 706 962 (0.49%). Percentage distribution points of critical value boundaries based on patients' test reports are 0.007% ~ 6.04% for low boundaries and 71.70% ~ 99.99% for high boundaries. After the intervention, the timely notification ratio, notification receipt ratio, and timely notification receipt ratio of critical values of ED, IPD, and total patients had increased, with a significant difference (P < .001). Five quality indicators, such as notification ratio, timely notification ratio, notification receipt ratio, timely notification receipt ratio, and clinician response ratio over a 5-year period, were 100%, 94%, 97%, 92%, and 99%, respectively.
CONCLUSIONS: We enhanced the effectiveness of clinical laboratory critical values initiative notification by implementing a closed-loop system and intervening. Clinical critical values and quality indicators should be analyzed and monitored to avoid adversely affecting patient care.
© 2019 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals, Inc.

Entities:  

Keywords:  hospital information system; laboratory critical values; patient safety; quality indicators

Mesh:

Year:  2019        PMID: 31531906      PMCID: PMC7031628          DOI: 10.1002/jcla.23038

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


INTRODUCTION

There has been increased concern about issues involved with enhancing the effectiveness of clinical laboratory critical values notification since the publication of a report entitled “When to panic over abnormal values” by George Lundberg in the 1970s.1 Laboratory critical values present a pathophysiological state at such variance with normal as to be life‐threatening if an action is not taken quickly and for which an effective action is possible.2 Critical values are needed to be proactively identified and reported timely and accurately so as to support effective clinical decision‐making based on the test results.3, 4 The effectiveness of clinical laboratory critical values notification will directly be related with the safety of patients and affect the satisfaction of customers to laboratory service.2 Meanwhile, accreditation institutions, such as ISO 15189, College of American Pathologists (CAP), and Joint Commission International (JCI), established the mandatory requirement for laboratory critical values management, including the identification, notification, handling, documentation, auditing, and quality indicators monitoring of laboratory critical values.5, 6 A growing number of publications have addressed the reporting of critical values.3, 7, 8, 9, 10, 11, 12 A CAP‐sponsored study of 121 institutions determined that it takes a total of 7 minutes for technician to notify clinicians about a critical result once testing was complete.13 It took up a lot of time reporting thousands of critical values by laboratories each year. On the other hand, a CAP Q‐Probes study in 623 institutions showed that about 5% of critical value telephone calls were abandoned, with the largest percentage abandoned for outpatients.14 There were some problems with the effectiveness of critical value notifications. The typical processes of laboratory critical value notification are as follows. A laboratory critical value is (a) first perceived by a technician in the laboratory, (b) then reported by the technician to clinicians or nurses in time, (c) then the notification transferred and received by the clinician, (d) then clinician response is made for the patient, and (e) documentation of the response is recorded in the patient's electronic medical record (EMR). A closed‐loop laboratory critical value notification system was developed based on the above five steps, and quality indicators were designed to monitor the notification process of laboratory critical values. A 5‐year retrospective observational study about laboratory critical values was introduced.

MATERIALS AND METHODS

Setting

All clinical laboratory critical values that occurred in the emergency department (ED), inpatient department (IPD), and outpatient department (OPD) of a 1000‐bed tertiary hospital at Beijing Tsinghua Changgung Hospital (BTCH, Beijing, China) were documented and analyzed retrospectively from January 2015 through June 2019. These included all critical values for hematology, coagulation, clinical chemistry, and microbiology testing. A closed‐loop laboratory critical value notification system combined with mobile phone short message and phone call was developed by a multidisciplinary team of clinicians, laboratorians, administrators, and information technology experts. As we previously reported, the system was applied to the clinic since 2015 throughout the entire hospital.15, 16

Establishing a critical value list

Laboratory items to be notified with critical values were selected by laboratory director in discussion with the clinicians who use laboratory services, referring to relevant literature.13, 14, 17, 18, 19, 20, 21 Considering the needs of special patients, such as cardiac surgery patients, critical test (high‐sensitivity troponin T), and its thresholds were also added into the critical value list.22 Critical value thresholds were set by consideration of relevant patient characteristics, clinical conditions, and the needs of clinicians to meet the special requirements of different patients for critical value boundaries.3, 18 And critical value boundaries were evaluated by calculating the percentage distribution points of the critical value boundaries based on the patients' data distribution. All the critical items and thresholds were implemented in hospital since January 2015 and modified through the annual discussion meeting with clinicians (as shown in Table 1).
Table 1

Critical values by tests for all patients from 2015 to 2019

Critical value items and thresholdsNumber of critical valuesConstituent ratio (%)a Total number of reportsIncidence ratio (%)b Percentage distributionc
Clinical chemistry
High‐sensitivity troponin T, ≥0.053 ng/mL841022.1286 0029.7871.70%
Urea nitrogen, ≥25 mmol/L (70 mg/dL)33148.72366 9880.9098.60%
Potassium, ≤2.5 or ≥6.2 mmol/L26496.97388 6330.680.28%, 99.54%
Creatinine, ≥600 umol/L (6.787 mg/dL)24726.50382 7110.6598.00%
Glucose, ≤2.7 or ≥27.78 mmol/L (≤48.65 or ≥ 500.54 mg/dL)21475.65380 9170.560.04%, 99.97%
Sodium, ≤120 or ≥160 mmol/L6491.71387 7840.170.03%, 99.97%
Arterial partial pressure of carbon dioxide (blood gas), ≤20 or ≥70 mm Hg6021.5820 5842.921.51%, 99.99%
Calcium (serum), ≤1.5 or ≥3.5 mmol/L5831.53364 3680.160.03%, 99.99%
Arterial partial pressure of oxygen (blood gas), ≤50 mm Hg5801.5320 5842.8212.97%
Cholinesterase, ≤2130 U/L5541.46260 1920.216.04%
Bicarbonate (blood gas), ≤10 or ≥40 mmol/L3470.9120 5841.691.20%, 98.22%
pH value (blood gas), ≤7.2 or ≥7.63180.8420 5841.542.43%, 99.86%
Hematology
WBC count, ≤2* or ≥30*109/L35399.31595 9010.590.61%, 99.71%
Hemoglobin, ≤60 g/L (6 g/dL)21775.73595 9010.370.60%
Platelets count, ≤20* or ≥1000*109/L15804.16595 9010.270.43%, 99.98%
Neutrophils count, ≤0.5*109/L11613.05595 9010.190.28%
Percentage of primitive cells (peripheral blood), ≥1%1640.431 884 0840.01NA
Coagulation
Fibrinogen, ≤1.0 g/L9392.47163 2270.580.89%
Thrombin time, ≥150 s7932.09162 7550.4999.99%
Activated partial thromboplastin time, ≤15 or ≥100 s5621.48165 1400.340.01%, 99.70%
Prothrombin time, ≤9 or ≥70 s1320.35173 4100.080.007%, 99.93%
Microbiology
Blood culture, positive22795.9923 5879.66NA
Gram stain (sterile body fluid), positive20695.4451 2244.04NA
Total38 0201007 706 9620.49NA

Constituent ratio, the ratio between the number of critical values of a certain test and the total number of critical values of all twenty‐three test items.

Incidence ratio, the ratio between the number of critical values of a certain test and the total number of the corresponding item reported.

Percentage distribution is shown as the percentage distribution points of the low and high boundaries for the critical value of a test versus the frequency distribution of patients' reports of the test.

Critical values by tests for all patients from 2015 to 2019 Constituent ratio, the ratio between the number of critical values of a certain test and the total number of critical values of all twenty‐three test items. Incidence ratio, the ratio between the number of critical values of a certain test and the total number of the corresponding item reported. Percentage distribution is shown as the percentage distribution points of the low and high boundaries for the critical value of a test versus the frequency distribution of patients' reports of the test.

Intervention introduced (September 2015): Established quality control circle to improve the effectiveness of critical notification

Quality control circle (QCC) was established by a multidisciplinary team of laboratorians, nurses, and information technology experts to enhance the effectiveness of critical values notification. Three quality improvement strategies derived from the QCC implemented in hospital since September 2015, including (a) establish critical value notification policy and conduct employee education and assessment, (b) optimize the laboratory critical value notification system to display a pop‐up window to alert the technician when the critical values are generated, and (c) set up five quality indicators to monitor the whole process of critical values notification.

Design of laboratory critical value notification system and implementation of closed‐loop management

The flowchart of laboratory critical values notification is shown in Figure 1.
Figure 1

The flowchart of critical value notification process. Abbreviations: EMR, electronic medical record; HIS, hospital information system; LIS, laboratory information system

The flowchart of critical value notification process. Abbreviations: EMR, electronic medical record; HIS, hospital information system; LIS, laboratory information system The initial step involves critical values are perceived, verified, and then reported to clinical caregivers by technician within a certain time frame.19, 20 When a measured value triggers its critical value boundaries, the report will change color and a pop‐up window will show up in the laboratory information system (LIS) to remind the technician of the generation of critical value. The critical value will be verified before reporting to clinicians, including rechecking the specimen, repeating test,23 or contacting with clinicians for confirmation. The time frame criterion of notifying clinical caregivers of the critical values by a technician is 30 minutes for ED patients and 60 minutes for OPD and IPD patients. Two quality indicators, notification ratio (number of critical values notified by technician/total number of critical values required to notify × 100%) and timely notification ratio (number of critical values notified by technician within a certain time frame/total number of critical values required to notify × 100%), are used to monitor whether the critical value is reported and whether it was reported within the required time limits, respectively. Critical values are received by clinical caregivers and documented within a certain time frame. A locked screen will show on the caregivers’ computers when the message of critical values is received. The caregivers (usually primary nurses or clinicians) are required to document the acknowledgment of receipt of receiving notifications and input their employee card number and password to unlock the screen in time. Once done, the receipt message will transfer back to LIS. Meanwhile, Short Message Service (SMS) is employed to send a mobile phone short message, including “patient identification, critical value and results, time, and technician in charge”, to the patient's clinician. The documentation of critical value receipt is required within 15 minutes for ED patients, 45 minutes for IPD patients, and 480 minutes for OPD patients. The time frame criteria of documentation are set by meeting attended by relevant clinician, laboratory director, and hospital administrators. Another two quality indicators, notification receipt ratio (number of critical value receipts of caregivers acknowledgment/total number of critical values required to notify × 100%) and timely notification receipt ratio (number of critical value receipts of caregivers acknowledgment within a certain time frame/total number of critical values required to notify × 100%), are used to monitor whether the receipt of critical value is acknowledged by the caregivers and whether it was acknowledged within the required time limits, respectively. Additionally, if the caregivers do not confirm receipt in the notification information system within the above time frame criteria, then critical values are reported by technician over the telephone, and the call information is then documented in the system. An appropriate response is made by the doctors who were informed and the response is documented in the EMR, meanwhile, the records are transmitted from hospital information system (HIS) to LIS. Another indicator, clinician response ratio (number of critical values responsed by clinician/total number of critical values required to notify × 100%) is used to monitor whether the diagnosis or treatment for the critical value is made by doctors. In a word, five quality indicators, notification ratio, timely notification ratio, notification receipt ratio, timely notification receipt ratio, and clinician response ratio, are applied for monitoring the whole process of laboratory critical value management.

Statistical analysis

The TAT data of pre‐analytical, analytical, post‐analytical, and total analytical phase of laboratory critical values showed a skewed distribution by Kolmogorov‐Smirnov normality test (P < .01), the median and inter‐quartile range of the TAT were used for statistical analysis. Five critical value indicators were expressed as percentages. IBM SPSS Statistics for Windows, version 24 (IBM Corp.) and Microsoft Excel 2006 (Microsoft) were used for statistical analysis.

RESULTS

Critical value items, thresholds, and their percentage distribution

There were 7 706 962 test reports of 23 test items at BTCH from January 2015 through June 2019, of which 38 020 (0.49%) reports were notified as critical values, an average of about 32 critical values a day. Of the total critical values, most (24 050, 63%) were from inpatient department (IPD) patients, followed by emergency department (ED) patients (9211, 24%) and outpatient department (OPD) patients (4759, 13%). The top five items (thresholds, constituent ratio) in order of constituent ratios of critical values were high‐sensitivity troponin T (≥0.053 ng/mL, 22.12%), WBC count (≤2* or ≥30*109/L, 9.31%), urea nitrogen [≥25 mmol/L (70 mg/dL), 8.72%], potassium (≤2.5 or ≥6.2 mmol/L, 6.97%), and creatinine [≥600 umol/L (6.787 mg/dL), 6.50%]. According to the incidence ratios of critical values, the top five items (thresholds, incidence ratio) were high‐ sensitivity troponin T (≥0.053 ng/mL, 9.78%), blood culture (positive, 9.66%), Gram stain (sterile body fluid; positive, 4.04%), arterial partial pressure of carbon dioxide (blood gas; ≤20 or ≥70 mm Hg, 2.92%), and arterial partial pressure of oxygen (blood gas; ≤50 mm Hg, 2.82%). The percentage distribution points of critical value boundaries based on patients’ test reports are 0.007% ~ 6.04% for low boundaries and 71.70% ~ 99.99% for high boundaries, as shown in Table 1.

Turnaround time of critical value notification

The median (inter‐quartile range) of the turnaround time (TAT) of pre‐analytical (from sample collection to registration), analytical (from sample registration to critical values reporting), post‐analytical (from critical values reporting to notification of caregivers), post‐analytical (from critical values reporting to clinician response recorded in the EMR), and total analytical (from sample collection to clinician response recorded in the EMR) phase of all laboratory critical values at BTCH over 5 years were 27 (11, 80), 41 (27, 89), 1 (0, 2), 323 (52, 3255), and 648 (192, 3836) minutes, respectively. The median (inter‐quartile range) of the turnaround time of critical value notification by tests from 2015 to 2019 are listed in Table 2.
Table 2

The median (inter‐quartile range) minutes of the turnaround time (TAT) of critical value notification by tests from 2015 to 2019

Critical value items and thresholdsDetection system and method

Pre‐analytical TAT

(from sample collection to registration)

Analytical TAT

(from sample registration to critical values reporting)

Post‐analytical TAT

(from critical values reporting to notification of caregivers)

Post‐analytical TAT

(from critical values reporting to clinician's response recorded in the EMR)

Total TAT

(from sample collection to clinician's response recorded in the EMR)

Clinical chemistry
High‐sensitivity troponin T, ≥0.053 ng/mLRoche Cobas c8000, electrochemiluminescence29 (13, 73)32 (26, 42)1 (0, 2)188 (41, 2029)327 (139, 2120)
Urea nitrogen, ≥25 mmol/L (70 mg/dL)Siemens Advia 2400/Roche Cobas c8000, urease colorimetric36 (12, 93)50 (35, 99)1 (0, 3)544 (73, 5708)770 (221, 5939)
Potassium, ≤2.5 or ≥6.2 mmol/LSiemens Advia 2400/Roche Cobas c8000/Roche b 221, ion‐selective electrode

18 (6, 39; routine)a

30 (12, 94.5; urgent)b

11 (5, 21; blood gas)c

198 (127, 316; routine)

37 (29, 51; urgent)

9 (5, 15; blood gas)

20 (1, 90; routine)

1(0, 2; urgent)

1 (0, 2; blood gas)

2581 (199, 37 926; routine)

253 (47, 2657; urgent)

224 (45, 1445; blood gas)

2919 (610, 38 090; routine)

453 (169, 2803; urgent)

290 (87, 1535; blood gas)

Creatinine, ≥600 umol/L (6.787 mg/dL)Siemens Advia 2400/Roche Cobas c8000, enzymatic

12 (2, 60; routine)

31 (11, 83; urgent)

153 (100, 252; routine)

36 (28, 48; urgent)

10 (1, 67; routine)

1(0, 2; urgent)

3968 (328, 31 691; routine)

802 (81, 10 235; urgent)

4369 (1066, 31 942; routine)

931(194, 10 363; urgent)

Glucose, ≤2.7 or ≥27.78 mmol/L (≤48.65 or ≥500.54 mg/dL)Siemens Advia 2400/Roche Cobas c8000, enzymatic

112 (20, 204; routine)

28 (11, 88; urgent)

13 (6, 27; blood gas)

199 (134, 295; routine)

41 (32, 54; urgent)

9 (6, 16; blood gas)

1 (0, 17; routine)

1 (0, 2; urgent)

1 (0, 2; blood gas)

1053 (117, 3148; routine)

246 (49, 2978; urgent)

305 (64, 1462; blood gas)

1443 (538, 4256; routine)

425 (166, 3112; urgent)

343 (105, 1611; blood gas)

Sodium, ≤120 or ≥160 mmol/LSiemens Advia 2400/Roche Cobas c8000, ion‐selective electrode

59 (19, 179; routine)

25 (11, 62; urgent)

219 (138, 338; routine)

35 (27, 47; urgent)

1 (0, 2; routine)

1 (0, 2; urgent)

181 (31, 1200; routine)

258 (44, 3162; urgent)

671 (408, 1637; routine)

376 (150, 3223; urgent)

Arterial partial pressure of carbon dioxide (blood gas), ≤20 or ≥70 mm HgRoche Cobas b211, ion‐selective electrode11 (5, 20)7 (5, 13)1 (0, 3)271 (46, 2310)328 (74, 2395)
Calcium (serum), ≤1.5 or ≥3.5 mmol/LSiemens Advia 2400/Roche Cobas c8000, colorimetric

60 (15, 144; routine)

37 (14, 88; urgent)

255 (165, 364; routine)

41 (32, 52; urgent)

1 (0, 17; routine)

1 (0, 2; urgent)

309 (66, 2720; routine)

203 (47, 1589; urgent)

1225 (464, 3493; routine)

392 (168, 1984; urgent)

Arterial partial pressure of oxygen (blood gas), ≤50 mm HgRoche Cobas b211, ion‐selective electrode12 (6, 21)8 (5, 14)1 (0, 2)205 (37, 2735)269 (68, 2862)
Cholinesterase, ≤2130 U/LRoche Cobas c8000, butyrylthiocholine (Trinder)20 (10, 51)32 (26, 41)1 (0, 2)678 (68, 8065)741(151, 8112)
Bicarbonate (blood gas), ≤10 or ≥40 mmol/LRoche Cobas b211, calculated12 (5, 20)8(5, 12)1(0, 3)318 (63, 2730)348 (88, 2853)
pH value (blood gas), ≤7.2 or ≥7.6Roche Cobas b211, ion‐selective electrode8 (4, 16)8 (5, 13)1 (0, 3)190 (35, 2549)257 (63, 2740)
Hematology
WBC count, ≤2* or ≥30*109/LSysmex XN‐9000, electrical impedance53 (15, 142)43 (20, 80)1(0, 2)229(39, 2745)446(206, 2861)
Hemoglobin, ≤60 g/L (6 g/dL)Sysmex XN‐9000, sodium dodecyl sulfate hemoglobin30 (12, 107)27 (14, 54)1 (0, 2)259 (41, 3002)445 (160, 3135)
Platelets count, ≤20* or ≥1000*109/LSysmex XN‐9000, electrical impedance52 (17, 138)44 (21, 80)1 (0, 2)259 (49, 2856)465 (222, 3034)
Neutrophils count, ≤0.5*109/LSysmex XN‐9000, light scattering48 (13, 142)50 (21, 89)1 (0, 3)307(50, 2824)519 (239, 3016)
Percentage of primitive cells (peripheral blood), ≥1%Sysmex XN‐9000/Microscope, counting103 (29, 161)115 (88, 164)1 (0, 2)253 (48, 1639)893 (307, 1964)
Coagulation
Fibrinogen, ≤1.0 g/LSysmex CS5100, coagulation40 (16, 82)90 (53, 150)1 (0, 2)270 (39, 1915)510 (213, 2160)
Thrombin time, ≥150 sSysmex CS1500, coagulation25(12, 57)92 (55, 144)1 (0, 2)462 (49, 2564)637 (232, 2740)
Activated partial thromboplastin time, ≤15 or ≥100 sSysmex CS1500, coagulation21 (11, 48)79 (49, 141)1 (0, 2)329 (35, 2364)497 (188, 2629)
Prothrombin time, ≤9 or ≥70 sSysmex CS1500, coagulation24 (12, 48)115 (66, 191)1 (0, 2)285 (29, 3670)549 (235, 3816)
Microbiology
Blood culture, positiveBD blood culture25 (14, 39)1250 (967, 3031)1 (0, 2)230 (16, 1045)3176 (1328, 5850)
Gram stain (sterile body fluid), positiveMicroscope, manual22 (12, 39)1139 (870, 1523)1 (0, 2)974 (115, 3951)2633 (1485, 5617)
Total 27 (11, 80)41 (27, 89)1 (0, 2)323 (52, 3255)648 (192, 3836)

Abbreviation: EMR, electronic medical record.

Routine: the test was analyzed in the way of routine examination by a routine instrument (Siemens Advia 2400). The same below.

Urgent: the test was analyzed in the way of urgent examination by an urgent instrument (Roche Cobas c8000). The same below.

Blood gas: the test was analyzed by the blood gas instrument (Roche b 221). The same below.

The median (inter‐quartile range) minutes of the turnaround time (TAT) of critical value notification by tests from 2015 to 2019 Pre‐analytical TAT (from sample collection to registration) Analytical TAT (from sample registration to critical values reporting) Post‐analytical TAT (from critical values reporting to notification of caregivers) Post‐analytical TAT (from critical values reporting to clinician's response recorded in the EMR) Total TAT (from sample collection to clinician's response recorded in the EMR) 18 (6, 39; routine)a 30 (12, 94.5; urgent)b 11 (5, 21; blood gas)c 198 (127, 316; routine) 37 (29, 51; urgent) 9 (5, 15; blood gas) 20 (1, 90; routine) 1(0, 2; urgent) 1 (0, 2; blood gas) 2581 (199, 37 926; routine) 253 (47, 2657; urgent) 224 (45, 1445; blood gas) 2919 (610, 38 090; routine) 453 (169, 2803; urgent) 290 (87, 1535; blood gas) 12 (2, 60; routine) 31 (11, 83; urgent) 153 (100, 252; routine) 36 (28, 48; urgent) 10 (1, 67; routine) 1(0, 2; urgent) 3968 (328, 31 691; routine) 802 (81, 10 235; urgent) 4369 (1066, 31 942; routine) 931(194, 10 363; urgent) 112 (20, 204; routine) 28 (11, 88; urgent) 13 (6, 27; blood gas) 199 (134, 295; routine) 41 (32, 54; urgent) 9 (6, 16; blood gas) 1 (0, 17; routine) 1 (0, 2; urgent) 1 (0, 2; blood gas) 1053 (117, 3148; routine) 246 (49, 2978; urgent) 305 (64, 1462; blood gas) 1443 (538, 4256; routine) 425 (166, 3112; urgent) 343 (105, 1611; blood gas) 59 (19, 179; routine) 25 (11, 62; urgent) 219 (138, 338; routine) 35 (27, 47; urgent) 1 (0, 2; routine) 1 (0, 2; urgent) 181 (31, 1200; routine) 258 (44, 3162; urgent) 671 (408, 1637; routine) 376 (150, 3223; urgent) 60 (15, 144; routine) 37 (14, 88; urgent) 255 (165, 364; routine) 41 (32, 52; urgent) 1 (0, 17; routine) 1 (0, 2; urgent) 309 (66, 2720; routine) 203 (47, 1589; urgent) 1225 (464, 3493; routine) 392 (168, 1984; urgent) Abbreviation: EMR, electronic medical record. Routine: the test was analyzed in the way of routine examination by a routine instrument (Siemens Advia 2400). The same below. Urgent: the test was analyzed in the way of urgent examination by an urgent instrument (Roche Cobas c8000). The same below. Blood gas: the test was analyzed by the blood gas instrument (Roche b 221). The same below.

Enhance the effectiveness of critical values notification by the intervention

Three quality improvement strategies derived from the QCC were implemented throughout the hospital in September 2015, as shown in Figure 2. To analyze the effects of interventions, baseline data were collected for a 9‐month period (January 2015 through September 2015 as the pre‐intervention period and October 2015 through June 2016 as the initial post‐intervention comparison period). After the intervention, timely notification ratio, notification receipt ratio, and timely notification receipt ratio of critical values of ED, IPD, and total patients were all increased, with a significant difference for the two periods (P < .001, Table 3).
Figure 2

Five quarterly quality indicators at Beijing Tsinghua Changgung Hospital over a 5‐y period

Table 3

Comparing critical value indicators between 2015 and 2016

SitesJanuary‐September, 2015 (before intervention)October, 2015‐June, 2016 (after intervention)
Notification ratio (No./Total No.)Timely notification ratio (No./Total No.)Notification receipt ratio (No./Total No.)Timely notification receipt ratio (No./Total No.)Clinician response ratio (No./Total No.)Notification ratio (No./Total No.)Timely notification ratio (No./Total No.)Notification receipt ratio (No./Total No.)Timely notification receipt ratio (No./Total No.)Clinician response ratio (No./Total No.)
ED100% (286/286)88% (253/286)99% (285/286)94% (270/286)98% (281/286)100% (1132/1132)95% (1073/1132)a 100% (1132/1132)a 96% (1092/1132)98% (1107/1132)
IPD100% (961/961)90% (869/961)96% (923/961)95% (912/961)99% (947/961)100% (2177/2177)95% (2057/2177)a 99% (2165/2177)a 99% (2157/2177)a 99% (2160/2177)
OPD100% (80/80)90% (72/80)68% (54/80)43% (34/80)99% (79/80)100% (327/327)91% (299/327)67% (220/327)49% (160/327)99% (325/327)
Total100% (1327/1327)90% (1194/1327)95% (1262/1327)92% (1216/1327)98% (1307/1327)100% (3636/3636)94% (3429/3636)a 97% (3517/3636)a 94% (3409/3636)a 99% (3592/3636)

Abbreviations: ED, emergency department; IPD, inpatient department; OPD, outpatient department.

Chi‐square test, P < .001 vs before intervention.

Five quarterly quality indicators at Beijing Tsinghua Changgung Hospital over a 5‐y period Comparing critical value indicators between 2015 and 2016 Abbreviations: ED, emergency department; IPD, inpatient department; OPD, outpatient department. Chi‐square test, P < .001 vs before intervention.

Quality indicators of critical values

Five quality indicators, such as notification ratio, timely notification ratio, notification receipt ratio, timely notification receipt ratio, and clinician response ratio, from total patients over a 5‐year period at BTCH, are 100%, 94%, 97%, 92%, and 99%, respectively. However, critical values from OPD patients show relatively poor indicators, timely notification ratio, notification receipt ratio, and timely notification receipt ratio are 92%, 72%, and 48%, respectively. Five quarterly quality indicators of critical values were shown in Figure 2 and Table 4.
Table 4

Five quality indicators of critical value for all patients from 2015 to 2019

SitesTotal number of critical values required to notifyNumber of critical values notifiedNotification ratio (%)Number of critical values notified by technician within a certain time framea Timely notification ratio (%)Number of notifications receipt of caregivers acknowledgmentNotification receipt ratio (%)Number of notifications receipt of caregivers acknowledgment within a certain time frameb Timely notification receipt ratio (%)Number of critical values responsed by clinicianClinician response ratio (%)
ED817281721007811968171100789897806499
IPD25 85425 85410024 3179425 6519925 1819725 52099
OPD39943994100368492287872191948396099
Total38 02038 02010035 8129436 7009734 9989237 54499

Abbreviations: ED, emergency department; IPD, inpatient department; OPD, outpatient department.

The time frame criterion of notifying clinical caregivers of the critical values by technician is 30 min for ED patients and 60 min for OPD and IPD patients.

The time frame criterion of the documentation of critical value receipt is 15 min for ED patients, 45 min for IPD patients, and 480 min for OPD patients.

Five quality indicators of critical value for all patients from 2015 to 2019 Abbreviations: ED, emergency department; IPD, inpatient department; OPD, outpatient department. The time frame criterion of notifying clinical caregivers of the critical values by technician is 30 min for ED patients and 60 min for OPD and IPD patients. The time frame criterion of the documentation of critical value receipt is 15 min for ED patients, 45 min for IPD patients, and 480 min for OPD patients.

DISCUSSION

A complete critical value notification and response process should be established in hospitals to provide safe and high‐quality medical services.24 This study described here was a 5‐year retrospective observational report of laboratory critical values notification after implementing the electronic closed‐loop notification system. The main strengths of the study were as follows: data coverage for 5 years, a large number of objects (38 020 critical values of over 7 million item reports), and multiple service practice sites, including the ED, IPD, and OPD. Previously, laboratory critical values notification was often made by telephone and read‐back. It was more time‐consuming and easy to have missing reports or even false reports.5, 14, 25 The ratio of errors made by telephone contacts for critical values was 3.5% reported by Joan Barenfanger et al25 and 5.0% reported by Peter J et al14 Our study had clearly documented that implementing a electronic closed‐loop laboratory critical value notification system combining with HIS, mobile phone short message, and phone call was an effective intervention to improve the critical values initiative notification.5, 24 The total incidence ratio of critical values over a 5‐year period was 0.49%, which was higher than that of 0.25% in Massachusetts General Hospital26 (Medical Center Teaching Hospital, USA), and lower than that of 0.96% in Zhejiang University First Affiliated Hospital27 (Tertiary Teaching Hospital, China) and 0.57% in Sun Yat‐sen University Ophthalmic Center7 (Special Hospital, China). The remarkable inter‐laboratory differences in the critical values notification existed between different hospitals. Excessively reporting critical values may make clinicians less sensitive to true critical values. Consensus on the items and their thresholds of critical values should be established by clinical laboratorians and clinicians together, based on the characteristic of the institution itself and percentage distribution of critical value thresholds.2, 18 We previously reported that the percentage distribution points of the critical value boundaries can be evaluated on the basis of the patients' data distribution.16 The data could provide references for the review meeting with clinicians. We further studied the timeliness of notification, the median time from a technician notification of the critical value until the time the critical value was reported successfully to caregivers was 1 minute (Table 2), which was much shorter than the reported 6 minutes suggested by Carmen Ricos et al28 and 7 minutes in a CAP Q‐Probes study of 121 Institutions.13 On the other hand, the median time of post‐analytical TAT (from critical values reporting to clinician response and recorded in the EMR) for total critical values was 323 minutes, which was much longer than that of pre‐analytical TAT (from sample collection to registration, 27 minutes) and analytical TAT (from sample registration to critical values reporting, 41 minutes). This prompted the group of critical value management of the hospital should optimize the procedures to ensure the clinicians get the information as soon as possible and treat the patients in time.29 The introduced quality improvement strategies from the QCC contributed to greatly improve the effectiveness of critical values notification. The study showed that poor indicators were from outpatients and that more attention should be payed to OPD critical value management.14 The quality indicators were used to monitor the whole process of critical value notification, point‐to‐point communication improvements were carried out in the department with deficiencies in the indicators of critical value notification. The continuous monitoring of quality indicator data allowed identification all possible improvements, promoted the reduction of errors, and improved quality of the critical value notification, thus guaranteeing patient safety.20, 21 This study may provide some ideas for other hospitals, including how to establish the flowchart of notification, how to set items and thresholds, and how to define related quality indicators to monitor the whole process. Further study on the personalized application of critical values for different types of patients in different departments is needed. Managers of hospitals and laboratories should attach more importance to the construction of the critical value notification system, and the closed‐loop management, thus ensuring patient safety.

CONFLICT OF INTEREST

The authors declare that they have no conflict of interest.
  27 in total

1.  Improving patient safety by repeating (read-back) telephone reports of critical information.

Authors:  Joan Barenfanger; Robert L Sautter; Diane L Lang; Susan M Collins; Donna M Hacek; Lance R Peterson
Journal:  Am J Clin Pathol       Date:  2004-06       Impact factor: 2.493

2.  Application of a rules-based natural language parser to critical value reporting in anatomic pathology.

Authors:  Scott R Owens; Ulysses G J Balis; David R Lucas; Jeffrey L Myers
Journal:  Am J Surg Pathol       Date:  2012-03       Impact factor: 6.394

3.  Critical values, panic values, or alert values?

Authors:  G L Hortin; G Csako
Journal:  Am J Clin Pathol       Date:  1998-04       Impact factor: 2.493

4.  Critical values: ASCP practice parameter. American Society of Clinical Pathologists.

Authors:  K Emancipator
Journal:  Am J Clin Pathol       Date:  1997-09       Impact factor: 2.493

5.  Defining a roadmap for harmonizing quality indicators in Laboratory Medicine: a consensus statement on behalf of the IFCC Working Group "Laboratory Error and Patient Safety" and EFLM Task and Finish Group "Performance specifications for the extra-analytical phases".

Authors:  Laura Sciacovelli; Mauro Panteghini; Giuseppe Lippi; Zorica Sumarac; Janne Cadamuro; César Alex De Olivera Galoro; Isabel Garcia Del Pino Castro; Wilson Shcolnik; Mario Plebani
Journal:  Clin Chem Lab Med       Date:  2017-08-28       Impact factor: 3.694

Review 6.  Extra-analytical quality indicators and laboratory performances.

Authors:  Laura Sciacovelli; Ada Aita; Mario Plebani
Journal:  Clin Biochem       Date:  2017-03-25       Impact factor: 3.281

7.  Critical values comparison: a College of American Pathologists Q-Probes survey of 163 clinical laboratories.

Authors:  Elizabeth A Wagar; Richard C Friedberg; Rhona Souers; Ana K Stankovic
Journal:  Arch Pathol Lab Med       Date:  2007-12       Impact factor: 5.534

8.  Enhance the effectiveness of clinical laboratory critical values initiative notification by implementing a closed-loop system: A five-year retrospective observational study.

Authors:  Runqing Li; Tengjiao Wang; Lijun Gong; Jingxiao Dong; Nan Xiao; Xiaohuan Yang; Dong Zhu; Zhipeng Zhao
Journal:  J Clin Lab Anal       Date:  2019-09-18       Impact factor: 2.352

9.  Utility and necessity of repeat testing of critical values in the clinical chemistry laboratory.

Authors:  Aijun Niu; Xianxia Yan; Lin Wang; Yan Min; Chengjin Hu
Journal:  PLoS One       Date:  2013-11-19       Impact factor: 3.240

10.  Using Plan-Do-Check-Act Circulation to Improve the Management of Panic Value in the Hospital.

Authors:  Suo-Wei Wu; Tong Chen; Yong Xuan; Xi-Wu Xu; Qi Pan; Liang-Yu Wei; Chao Li; Qin Wang
Journal:  Chin Med J (Engl)       Date:  2015-09-20       Impact factor: 2.628

View more
  2 in total

1.  Enhance the effectiveness of clinical laboratory critical values initiative notification by implementing a closed-loop system: A five-year retrospective observational study.

Authors:  Runqing Li; Tengjiao Wang; Lijun Gong; Jingxiao Dong; Nan Xiao; Xiaohuan Yang; Dong Zhu; Zhipeng Zhao
Journal:  J Clin Lab Anal       Date:  2019-09-18       Impact factor: 2.352

2.  Baseline assessment of staff perception of critical value practices in government hospitals in Kuwait.

Authors:  Talal ALFadhalah; Buthaina Al Mudaf; Haya Al Tawalah; Wadha A Al Fouzan; Gheed Al Salem; Hanaa A Alghanim; Samaa Zenhom Ibrahim; Hossam Elamir; Hamad Al Kharji
Journal:  BMC Health Serv Res       Date:  2022-08-03       Impact factor: 2.908

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.