Shading correction for volumetric CT using deep convolutional neural network and adaptive filter.

Xiaokun Liang1,2, Na Li1,2, Zhicheng Zhang1,2, Shaode Yu1,2, Wenjian Qin1,2, Yafen Li1,2, Shupeng Chen3, Huailing Zhang4, Yaoqin Xie1.   

Abstract

BACKGROUND: Shading artifact may lead to CT number inaccuracy, image contrast loss and spatial non-uniformity (SNU), which is considered as one of the fundamental limitations for volumetric CT (VCT) application. To correct the shading artifact, a novel approach is proposed using deep learning and an adaptive filter (AF).
METHODS: Firstly, we apply the deep convolutional neural network (DCNN) to train a human tissue segmentation model. The trained model is implemented to segment the tissue. According to the general knowledge that CT number of the same human tissue is approximately the same, a template image without shading artifact can be generated using segmentation and then each tissue is filled with the corresponding CT number of a specific tissue. By subtracting the template image from the uncorrected image, the residual image with image detail and shading artifact are generated. The shading artifact is mainly low-frequency signals while the image details are mainly high-frequency signals. Therefore, we proposed an adaptive filter to separate the shading artifact and image details accurately. Finally, the estimated shading artifacts are deleted from the raw image to generate the corrected image.
RESULTS: On the Catphan©504 study, the error of CT number in the corrected image's region of interest (ROI) is reduced from 109 to 11 HU, and the image contrast is increased by a factor of 1.46 on average. On the patient pelvis study, the error of CT number in selected ROI is reduced from 198 to 10 HU. The SNU calculated from the ROIs decreases from 24% to 9% after correction.
CONCLUSIONS: The proposed shading correction method using DCNN and AF may find a useful application in future clinical practice.

Entities:  

Keywords:  Shading artifact; adaptive filter (AF); deep convolution neural network; volumetric CT (VCT)

Year:  2019        PMID: 31448210      PMCID: PMC6685805          DOI: 10.21037/qims.2019.05.19

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  25 in total

1.  Direct measurement and analytical modeling of scatter in portal imaging.

Authors:  L Spies; P M Evans; M Partridge; V N Hansen; T Bortfeld
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Shading correction for on-board cone-beam CT in radiation therapy using planning MDCT images.

Authors:  Tianye Niu; Mingshan Sun; Josh Star-Lack; Hewei Gao; Qiyong Fan; Lei Zhu
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

3.  Scatter correction for cone-beam computed tomography using moving blocker strips: a preliminary study.

Authors:  Jing Wang; Weihua Mao; Timothy Solberg
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

4.  The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors.

Authors:  J H Siewerdsen; D J Moseley; B Bakhtiar; S Richard; D A Jaffray
Journal:  Med Phys       Date:  2004-12       Impact factor: 4.071

5.  Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT.

Authors:  Wojciech Zbijewski; Freek J Beekman
Journal:  IEEE Trans Med Imaging       Date:  2006-07       Impact factor: 10.048

6.  Scatter correction method for X-ray CT using primary modulation: theory and preliminary results.

Authors:  Lei Zhu; N Robert Bennett; Rebecca Fahrig
Journal:  IEEE Trans Med Imaging       Date:  2006-12       Impact factor: 10.048

7.  Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging.

Authors:  Heng Li; Radhe Mohan; X Ronald Zhu
Journal:  Phys Med Biol       Date:  2008-11-07       Impact factor: 3.609

8.  An Accurate Scatter Measurement and Correction Technique for Cone Beam Breast CT Imaging Using Scanning Sampled Measurement (SSM) Technique.

Authors:  Xinming Liu; Chris C Shaw; Tianpeng Wang; Lingyun Chen; Mustafa C Altunbas; S Cheenu Kappadath
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-02-28

9.  Scatter correction for cone-beam CT in radiation therapy.

Authors:  Lei Zhu; Yaoqin Xie; Jing Wang; Lei Xing
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

10.  Projection correlation based view interpolation for cone beam CT: primary fluence restoration in scatter measurement with a moving beam stop array.

Authors:  Hao Yan; Xuanqin Mou; Shaojie Tang; Qiong Xu; Maria Zankl
Journal:  Phys Med Biol       Date:  2010-10-12       Impact factor: 3.609

View more
  1 in total

1.  A Deep Unsupervised Learning Model for Artifact Correction of Pelvis Cone-Beam CT.

Authors:  Guoya Dong; Chenglong Zhang; Xiaokun Liang; Lei Deng; Yulin Zhu; Xuanyu Zhu; Xuanru Zhou; Liming Song; Xiang Zhao; Yaoqin Xie
Journal:  Front Oncol       Date:  2021-07-16       Impact factor: 6.244

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.