L Spies1, P M Evans, M Partridge, V N Hansen, T Bortfeld.
Abstract
In this study a direct measurement of scatter in portal imaging for various air gaps and scatterer thicknesses at a beam energy of 6 MV is presented. The experimental data are compared with results from a Monte Carlo (MC) scatter model. In the regime where the air gap is larger than 9.3 cm the MC and the experiment agree. Based on this MC model an analytical model is developed, which takes all important interaction processes into account. It comprises a rigorous treatment of first order scattering and an estimation of photons scattered more than once within the phantom. This estimation is based on the assumption that higher order scattering can be considered as isotropically distributed around a certain scatter origin located in the midplane of the phantom. It is found that relative deviations between the MC model and the analytical model are of 2% to 3% in regions where scattering is very large.Mesh:
Year: 2000 PMID: 10757598 DOI: 10.1118/1.598914
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071