Shading correction for on-board cone-beam CT in radiation therapy using planning MDCT images.

Tianye Niu1, Mingshan Sun, Josh Star-Lack, Hewei Gao, Qiyong Fan, Lei Zhu.   

Abstract

PURPOSE: Applications of cone-beam CT (CBCT) to image-guided radiationtherapy (IGRT) are hampered by shading artifacts in the reconstructed images. These artifacts are mainly due to scatter contamination in the projections but also can result from uncorrected beam hardening effects as well as nonlinearities in responses of the amorphous silicon flat panel detectors. While currently, CBCT is mainly used to provide patient geometry information for treatment setup, more demanding applications requiring high-quality CBCT images are under investigation. To tackle these challenges, many CBCT correction algorithms have been proposed; yet, a standard approach still remains unclear. In this work, we propose a shading correction method for CBCT that addresses artifacts from low-frequency projection errors. The method is consistent with the current workflow of radiation therapy.
METHODS: With much smaller inherent scatter signals and more accurate detectors, diagnostic multidetector CT (MDCT) provides high quality CT images that are routinely used for radiation treatment planning. Using the MDCT image as "free" prior information, we first estimate the primary projections in the CBCT scan via forward projection of the spatially registered MDCT data. Since most of the CBCT shading artifacts stem from low-frequency errors in the projections such as scatter, these errors can be accurately estimated by low-pass filtering the difference between the estimated and raw CBCT projections. The error estimates are then subtracted from the raw CBCT projections. Our method is distinct from other published correction methods that use the MDCT image as a prior because it is projection-based and uses limited patient anatomical information from the MDCT image. The merit of CBCT-based treatment monitoring is therefore retained.
RESULTS: The proposed method is evaluated using two phantom studies on tabletop systems. On the Catphan 600 phantom, our approach reduces the reconstruction error from 348 Hounsfield unit (HU) without correction to 4 HU around the object center after correction, and from 375 HU to 17 HU in the high-contrast regions. In the selected regions of interest (ROIs), the average image contrast is increased by a factor of 3.3. When noise suppression is implemented, the proposed correction substantially improves the contrast-to-noise ratio (CNR) and therefore the visibility of low-contrast objects, as seen in a more challenging pelvis phantom study. Besides a significant improvement in image uniformity, a low-contrast object of approximately 25 HU, which is otherwise buried in the shading artifacts, can be clearly identified after the proposed correction due to a CNR increase of 3.1. Compared to a kernel-based scatter correction method coupled with an analytical beam hardening correction, our approach also shows an overall improved performance with some residual artifacts.
CONCLUSIONS: By providing effective shading correction, our approach has the potential to improve the accuracy of more advanced CBCT-based clinical applications for IGRT, such as tumor delineation and dose calculation.

Entities:  

Mesh:

Year:  2010        PMID: 21089775     DOI: 10.1118/1.3483260

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  43 in total

1.  Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

2.  Evaluation of a cone beam CT artefact reduction algorithm.

Authors:  B Bechara; C A McMahan; H Geha; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-02-23       Impact factor: 2.419

3.  Proton dose calculation on scatter-corrected CBCT image: Feasibility study for adaptive proton therapy.

Authors:  Yang-Kyun Park; Gregory C Sharp; Justin Phillips; Brian A Winey
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

4.  Joint CT/CBCT deformable registration and CBCT enhancement for cancer radiotherapy.

Authors:  Yifei Lou; Tianye Niu; Xun Jia; Patricio A Vela; Lei Zhu; Allen R Tannenbaum
Journal:  Med Image Anal       Date:  2013-02-04       Impact factor: 8.545

5.  Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Authors:  Meng Wu; Andreas Keil; Dragos Constantin; Josh Star-Lack; Lei Zhu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

6.  Local filtration based scatter correction for cone-beam CT using primary modulation.

Authors:  Lei Zhu
Journal:  Med Phys       Date:  2016-11       Impact factor: 4.071

7.  Relationship between x-ray illumination field size and flat field intensity and its impacts on x-ray imaging.

Authors:  Xue Dong; Tianye Niu; Xun Jia; Lei Zhu
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

8.  Accelerated barrier optimization compressed sensing (ABOCS) reconstruction for cone-beam CT: phantom studies.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

9.  Dosimetric study on learning-based cone-beam CT correction in adaptive radiation therapy.

Authors:  Tonghe Wang; Yang Lei; Nivedh Manohar; Sibo Tian; Ashesh B Jani; Hui-Kuo Shu; Kristin Higgins; Anees Dhabaan; Pretesh Patel; Xiangyang Tang; Tian Liu; Walter J Curran; Xiaofeng Yang
Journal:  Med Dosim       Date:  2019-04-01       Impact factor: 1.482

10.  Statistical image-domain multimaterial decomposition for dual-energy CT.

Authors:  Yi Xue; Ruoshui Ruan; Xiuhua Hu; Yu Kuang; Jing Wang; Yong Long; Tianye Niu
Journal:  Med Phys       Date:  2017-02-21       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.