| Literature DB >> 31289734 |
S Badiou1,2, O Vuillot3, A S Bargnoux1,2, N Kuster1,2, S Lefebvre3, M Sebbane3, J P Cristol1,2, A M Dupuy1.
Abstract
BACKGROUND: Centrifugation is a consuming time step which participates to increase the turnaround time (TAT) in laboratories, likewise hemolysis sample that needs a re-sampling could delay management of patients. Recently, it has been postulated that BD Barricor™ tube could allow to decrease the centrifugation time and prevent hemolysis, two key feature to ensure high-quality results.Aim of the study was to evaluate the impact of replacing 4 mL BD vacutainer heparin lithium tube by low vacuum 3.5 mL BD vacutainer Barricor™ tube in an emergency department (ED) on hemolysis rate and TAT.Entities:
Keywords: Barricor; Emergency; Hemolysis; Low vacuum; Turnaround time TAT
Year: 2019 PMID: 31289734 PMCID: PMC6593183 DOI: 10.1016/j.plabm.2019.e00128
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Fig. 1Study design.
Fig. 2Percentage of samples according to HI values and type of tubes.
Comparison of re-sampling need according to Roche® values for hemolysis interference for each analyte during the two periods.
| Roche® values | Period 1 | Period 2 | p value | |
|---|---|---|---|---|
| HI over the limit | HI over the limit | |||
| mg/dl | (%) | (%) | ||
| ALAT | 90 | 8.5 | 2.2 | <0.001 |
| ASAT | 40 | 16.7 | 5.6 | <0.001 |
| Bicarbonate | 600 | 0.2 | 0 | 0.48 |
| Creatinine | 800 | 0.2 | 0 | 0.52 |
| CRP | 1000 | 0.1 | 0 | 0.99 |
| Potassium | 90 | 8.6 | 2.4 | <0.001 |
| Sodium | 1000 | 0.1 | 0 | 0.99 |
| Hs-cTnT | 100 | 7.0 | 2.3 | <0.001 |
ALAT, Alanine Aminotransferase; ASAT, Aspartate Aminotransferase; CRP, C-Reactive Protein (CRP); Hs-cTnT, High sensitivity Troponin T.
Lab turnaround time (TAT) 95% and percent of results available at 30 min according to the period.
| n | Period 1 | TAT 95% (min) | n | Period 2 | TAT 95% (min) | |
|---|---|---|---|---|---|---|
| % of results at 30 min | % of results at 30 min | |||||
| ALAT | 448 | 8 | 77 | 502 | 40* | 70* |
| ASAT | 448 | 7 | 79 | 502 | 39* | 75* |
| Bicarbonate | 1053 | 9 | 71 | 1057 | 51* | 62* |
| Creatinine | 1081 | 3 | 76 | 1101 | 39* | 65* |
| CRP | 1038 | 3 | 75 | 1053 | 40* | 64* |
| Potassium | 1094 | 15 | 75 | 1103 | 58* | 61* |
| Sodium | 1084 | 16 | 70 | 1099 | 59* | 59* |
| Hs-cTnT | 673 | 2 | 88 | 703 | 14* | 77* |
ALAT, Alanine Aminotransferase; ASAT, Aspartate Aminotransferase; CRP, C-Reactive Protein (CRP); hs-cTnT, High sensitivity Troponin T.
*, p < 0.001: comparison of period 2 versus period 1.
Fig. 3Graphical representation of lab TAT between the two periods (Black: use of BarricorTM tube, Blue: Use of regular Heparin lithium tube).