Low vacuum and discard tubes reduce hemolysis in samples drawn from intravenous catheters.

Connie Heiligers-Duckers1, Nathalie A L R Peters2, Jose J P van Dijck2, Jan M J Hoeijmakers2, Marcel J W Janssen3.   

Abstract

OBJECTIVES: In-vitro hemolysis is a great challenge to emergency departments where blood is drawn from intravenous catheters (IVCs). Although high quality samples can be obtained by straight needle venipuncture, IVCs are preferred for various reasons. The aim of this study was to identify blood collection practices that reduce hemolysis while using IVC. DESIGN AND METHODS: The study was conducted at an emergency department where blood is drawn in ≥ 90% of patients from IVC. Hemolysis, measured spectrophotometrically, was compared between syringe and vacuum tubes. The following practices were tested in combination with vacuum collection; a Luer-slip adapter, a Luer-lock adapter, discard tubes and low vacuum tubes. Each intervention lasted 1 week and retrieved 154 to 297 samples. As reference, hemolysis was also measured in vacuum tubes retrieved from departments where only straight needle venipuncture is performed.
RESULTS: Vacuum collection led to more hemolytic samples compared with syringe tubes (24% versus 16% respectively, p=0.008). No difference in hemolysis was observed between the Luer-slip and the Luer-lock adapter. The use of discard (17% hemolytic, p=0.045) and low vacuum tubes (12% hemolytic, p<0.001) substantially decreased hemolysis. None of the interventions reduced the hemolysis rate to the level observed when drawing blood by straight needle venipuncture (3%, p<0.02).
CONCLUSIONS: In summary, both discard and low vacuum tubes reduce hemolysis while drawing blood from IVC. Of these practices the use of a low vacuum tube is preferred considering the less volume of blood and the amount of tubes drawn.
Copyright © 2013 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Blood collection; Discard tube; Hemolysis; IVC; Intravenous catheter; Low vacuum tube; intravenous catheter

Mesh:

Year:  2013        PMID: 23603143     DOI: 10.1016/j.clinbiochem.2013.04.005

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  5 in total

1.  Blood Samples of Peripheral Venous Catheter or The Usual Way: Do Infusion Fluid Alters the Biochemical Test Results?

Authors:  Mahboobeh Taghizadeganzadeh; Mohammadreza Yazdankhahfard; Mohammadreza Farzaneh; Kamran Mirzaei
Journal:  Glob J Health Sci       Date:  2015-11-03

2.  European survey on preanalytical sample handling - Part 1: How do European laboratories monitor the preanalytical phase? On behalf of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase (WG-PRE).

Authors:  Janne Cadamuro; Giuseppe Lippi; Alexander von Meyer; Mercedes Ibarz; Edmee van Dongen; Michael Cornes; Mads Nybo; Pieter Vermeersch; Kjell Grankvist; Joao Tiago Guimaraes; Gunn B B Kristensen; Barbara de la Salle; Ana-Maria Simundic
Journal:  Biochem Med (Zagreb)       Date:  2019-06-15       Impact factor: 2.313

3.  Improved quality of samples and laboratory turnaround time using 3.5 mL low vacuum BD Vacutainer® Barricor tubes in the emergency department.

Authors:  S Badiou; O Vuillot; A S Bargnoux; N Kuster; S Lefebvre; M Sebbane; J P Cristol; A M Dupuy
Journal:  Pract Lab Med       Date:  2019-06-12

4.  European survey on preanalytical sample handling - Part 2: Practices of European laboratories on monitoring and processing haemolytic, icteric and lipemic samples. On behalf of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase (WG-PRE).

Authors:  Janne Cadamuro; Giuseppe Lippi; Alexander von Meyer; Mercedes Ibarz; Edmee van Dongen; Michael Cornes; Mads Nybo; Pieter Vermeersch; Kjell Grankvist; Joao Tiago Guimaraes; Gunn B B Kristensen; Barbara de la Salle; Ana-Maria Simundic
Journal:  Biochem Med (Zagreb)       Date:  2019-06-15       Impact factor: 2.313

5.  A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator.

Authors:  Ghassaan Gawria; Linda Tillmar; Eva Landberg
Journal:  J Clin Lab Anal       Date:  2019-10-11       Impact factor: 2.352

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.