The BD Barricor blood collection tube is an acceptable and robust alternative to the PST for use with the Beckman AccuTnI+3 assay.

Anna K Füzéry1, Joshua E Raizman2, Bobbi Lynn Goudreau3, Karin Moses3, Keri Niemann3, Janice Park3, George S Cembrowski2.   

Abstract

OBJECTIVES: BD Canada recently released a blood collection tube with a novel mechanical separator called the Barricor. We evaluated this tube as an alternate sample type for cardiac troponin I (cTnI) testing using the Beckman Coulter AccuTnI+3 assay. DESIGN AND METHODS: 3014 paired patient specimens (Barricor, plasma separator tube or PST) were obtained from the emergency departments and cardiac care units of nine hospitals in and around Edmonton, Alberta. After centrifugation, each plasma sample was analyzed for cTnI using the Beckman Coulter AccuTnI+3 assay. In addition, selected samples were analyzed multiple times within a single run or over 4-5days to generate imprecision data for the assay.
RESULTS: Repeatability and within-laboratory studies revealed an imprecision of <10% at concentrations above 0.025μg/L for the Barricor as well as BD's traditional PST. Paired patient sample comparisons over the full range of the assay yielded linear regression slopes ranging from 0.956 to 1.011 and Pearson correlation coefficients ranging from 0.993 to 0.999. At a lower range of results closer to the manufacturer's 99th percentile cutoffs correlation was slightly worse, but still acceptable, with linear regression slopes ranging from 0.967 to 1.211 and Pearson correlation coefficients ranging from 0.983 to 0.987. Notably, at these lower concentrations the agreement between individual PST and Barricor results worsened with decreasing cTnI concentration. Differences between pairs of results became particularly large (-50 to +400%) at PST cTnI concentrations ≤0.015μg/L. Closer inspection of the data around the 0.02 and 0.04μg/L 99th percentile cutoffs revealed a number of discordances between PST and Barricor results, with at least some of these attributable to false elevations in the PST results.
CONCLUSIONS: Together, our results suggest that the Barricor blood collection tube is good alternative to the traditional PST for cTnI testing using the AccuTnI+3 assay. The Barricor appears to minimize spurious, nonreproducible, and false elevations in cTnI results for a subset of patients but additional studies are needed to determine if it reduces overall false elevations. cTnI results below 0.04μg/L may still be of questionable accuracy even with the use of this new tube.
Copyright © 2017 The Canadian Society of Clinical Chemists. All rights reserved.

Entities:  

Keywords:  AccuTnI+3; Barricor; Evaluation; False elevations; Troponin

Mesh:

Substances:

Year:  2017        PMID: 28583435     DOI: 10.1016/j.clinbiochem.2017.05.015

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  6 in total

1.  Comparison of Barricor™ vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability.

Authors:  Anne Marie Dupuy; Stéphanie Badiou; Delphine Daubin; Anne Sophie Bargnoux; Chloé Magnan; Kadda Klouche; Jean Paul Cristol
Journal:  Biochem Med (Zagreb)       Date:  2018-04-15       Impact factor: 2.313

2.  Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.

Authors:  Sunghwan Shin; Jongwon Oh; Hyung-Doo Park
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

3.  Improved quality of samples and laboratory turnaround time using 3.5 mL low vacuum BD Vacutainer® Barricor tubes in the emergency department.

Authors:  S Badiou; O Vuillot; A S Bargnoux; N Kuster; S Lefebvre; M Sebbane; J P Cristol; A M Dupuy
Journal:  Pract Lab Med       Date:  2019-06-12

4.  Evaluation of the Barricor Tube in 28 Routine Chemical Tests and Its Impact on Turnaround Time in an Outpatient Clinic.

Authors:  Soo Young Moon; Han Sol Lee; Min Soon Park; In-Suk Kim; Sun Min Lee
Journal:  Ann Lab Med       Date:  2021-05-01       Impact factor: 3.464

5.  Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa.

Authors:  Siyabonga P Khoza; Sarah Ford; Ernest P Buthelezi; Donald M Tanyanyiwa
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

6.  Switching from serum to plasma: Implementation of BD Vacutainer® Barricor™ Plasma Blood Collection Tubes improves sample quality and laboratory turnaround time.

Authors:  Christian Ramakers; Brendan Meyer; Wanfei Yang; Elizabeth Plokhoy; Yan Xiong; Stephen Church; Nitin Kaushik
Journal:  Pract Lab Med       Date:  2019-12-10
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.