Quality of plasma samples and BD Vacutainer Barricor tubes: Effects of centrifugation.

Andrea Padoan1, Martina Zaninotto2, Elisa Piva2, Laura Sciacovelli2, Ada Aita3, Adriano Tasinato2, Mario Plebani3.   

Abstract

BACKGROUNDS: The BD Vacutainer® Barricor™ Plasma collection tube (BD Barricor) uses an innovative non-gel separation method. This study compared the plasma residual cell count (PRCC) obtained from BD Barricor and from BD PST II plasma tubes.
METHODS: Four BD Barricors and one BD PST II were collected from 40 donors. BD PST II was centrifuged at 1300g/10 min, while the BD Barricors were centrifuged at 1800g/10 min, 4000g/3 min, 4000g/7 min and 4000g/15 min. PRCC was evaluated measuring white blood cells (WBC), red blood cells (RBC) and Platelets (PLT) counts by Siemens ADVIA 2120. Cell-free hemoglobin was quantified by haemolysis index (HI) by Roche Cobas c501.
RESULTS: BD PST II Median WBC, RBC and PLT counts were 0.38 (109/L), 0.0291 (1012/L) and 113.5 (109/L), respectively. Considering the BD PST II as reference, PRCC differences were expressed as median bias percentage. WBC showed a significant reduction at all the conditions (p < 0.01), being the reductions: 63.9% (1800g/10 min), 69.9% (4000g/3 min), 75.0% (4000g/7 min) and 82.7% (4000g/15 min). RBC reductions 29.7% (1800g/10 min), 33.8% (4000g/3 min), 39.6% (4000g/7 min) and 66.4 (4000g/15 min) were all significant (p < 0.01). PLT reductions were 1.6% at 1800g/10 min (p = ns), 1.2% at 4000g/3 min (p = ns), 27.1% at 4000g/7 min (p = 0.046) and 46.6% at 4000g/15 min (p = 0.005). BD Barricor centrifuged for 7 and 15 min at 4000g showed an increased haemolysis.
CONCLUSIONS: BD Barricors plasma quality improved with increasing the centrifugation times but already at 4000g/3 min, the suggested centrifugation condition, a significant improvement was achieved.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  BD Barricor; Blood analyses; Blood collection tubes; Plasma; Plasma quality; Preanalytical; Residual cell counts; Vacuum tubes

Mesh:

Year:  2018        PMID: 29753681     DOI: 10.1016/j.cca.2018.05.018

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  5 in total

1.  Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.

Authors:  Sunghwan Shin; Jongwon Oh; Hyung-Doo Park
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

2.  Improved quality of samples and laboratory turnaround time using 3.5 mL low vacuum BD Vacutainer® Barricor tubes in the emergency department.

Authors:  S Badiou; O Vuillot; A S Bargnoux; N Kuster; S Lefebvre; M Sebbane; J P Cristol; A M Dupuy
Journal:  Pract Lab Med       Date:  2019-06-12

3.  A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator.

Authors:  Ghassaan Gawria; Linda Tillmar; Eva Landberg
Journal:  J Clin Lab Anal       Date:  2019-10-11       Impact factor: 2.352

4.  Switching from serum to plasma: Implementation of BD Vacutainer® Barricor™ Plasma Blood Collection Tubes improves sample quality and laboratory turnaround time.

Authors:  Christian Ramakers; Brendan Meyer; Wanfei Yang; Elizabeth Plokhoy; Yan Xiong; Stephen Church; Nitin Kaushik
Journal:  Pract Lab Med       Date:  2019-12-10

5.  Biochemical, immunochemical and serology analytes validation of the lithium heparin BD Barricor blood collection tube on a highly automated Roche COBAS8000 instrument.

Authors:  Davide Ferrari; Marta Strollo; Matteo Vidali; Andrea Motta; Marina Pontillo; Massimo Locatelli
Journal:  Acta Biomed       Date:  2020-03-19
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.