Jafri Kuthubutheen1, Amandeep Grewal2,3, Sean Symons2,3, Julian Nedzelski2,3, David Shipp2,3, Vincent Lin2,3, Joseph Chen2,3.
Abstract
OBJECTIVES: To determine if cochlear duct length and cochlear basal diameter, measured using routinely available radiology software, affect hearing outcomes after cochlear implantation with two different length electrodes.Entities:
Mesh:
Year: 2019 PMID: 31275978 PMCID: PMC6582906 DOI: 10.1155/2019/5849871
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1The A value measured in the oblique coronal plane (a). A straight line-measuring tool (b) is used to measure the outer wall cochlear duct length to 360 degrees. A side profile view (c) indicates the individual points used to calculate the length till 720 degrees.
Preoperative demographics. SD = standard deviation. P-value obtained using the Mann-Whitney U test and Chi square test.
| Group | n | Mean age of implantation | Sex | Side of |
|---|---|---|---|---|
| Flex 31 | 34 | 62 (14.5) | 20:14 | 16 Left |
| 18 Right | ||||
|
| ||||
| Flex 28 | 21 | 63 (9.8) | 10:11 | 8 Left |
| 13 Right | ||||
|
| ||||
| Combined | 55 | 62 (12.8) | 30: 25 | 24 Left |
| 31 Right | ||||
|
| ||||
| p-value | 0.81 | 0.58 | 0.58 | |
Hearing and insertion outcome measures for each group. PTA = pure tone average at 250Hz, 500Hz and 1000Hz, HINT = hearing in noise test, and CNC = consonant nucleus consonant. P-value obtained using the Mann-Whitney U test, with significant values indicated in italic. Numbers in brackets indicate standard deviations.
| Flex 31 | Flex 28 | p-value | |
|---|---|---|---|
| Preo op | |||
| PTA (dB) | 69 (13.2) | 72 (8.9) | 0.555 |
| Best aided CNC quiet (%) | 12.3 (15.8) | 30.9 (16.1) |
|
| HINT sentence quiet (%) | 36 (24.9) | 45.2 (22.2) | 0.218 |
| AZBIO sentence quiet (%) | n/a | 33.3 (20.6) | |
| 6 months post op | |||
| Best aided CNC quiet (%) | 52.6 (22.8) | 59.7 (17.9) | 0.399 |
| HINT sentence quiet (%) | 83.7 (21.8) | n/a | |
| AZBIO sentence quiet (%) | n/a | 63.3 (19.99) | |
| CNC score shift (%) | 40.7 (24.1) | 28.9 (17.4) | 0.055 |
| Insertion outcomes | |||
| (1) Degrees of electrode insertion on plain X-ray | 489.4 degrees | 525 degrees | 0.165 |
| (2) Number of channels inserted on plain X-ray (out of 12) | 10.85 (1.08) | 11.33 (0.73) | 0.12 |
| Correlation |
|
| |
| between (1) and (2) | (p<0.01) | (p<0.01) |
Cochlear metrics for all subjects. There is no statistically significant difference in all measurements between subjects receiving either electrode.
| Minimum | Maximum | Mean | Standard Deviation | ||
|---|---|---|---|---|---|
| A-vaue | Flex 31 | 8.1 | 9.8 | 8.94 | 1.63 |
| Flex 28 | 8.1 | 9.5 | 8.87 | 0.38 | |
| Combined | 8.91 | 0.37 | |||
|
| |||||
| Outer wall length to 360 degrees | Flex 31 | 29.8 | 35.9 | 32.52 | 1.41 |
| Flex 28 | 27.8 | 34.7 | 31.91 | 1.63 | |
| Combined | 32.29 | 1.51 | |||
|
| |||||
| Outer wall length to 720 degrees | Flex 31 | 19.6 | 23.5 | 21.4 | 0.94 |
| Flex 28 | 18.3 | 22.6 | 21.1 | 1.09 | |
| Combined | 21.3 | 1 | |||
Figure 2A histogram showing the normal distribution of the outer wall CDL to 720 degrees in all 55 patients.
A-value correlations with postoperative outcomes at 6 months. There are no statistically significant correlations between cochlear metrics listed on the left column and shifts in PTA, CNC, or AZBio scores which are not shown here. ∗∗ p<0.01, ∗ p<0.05.
| Flex 31 | Flex 28 | ||||
|---|---|---|---|---|---|
| CNC (%) | HINT (%) | CNC (%) | AZBIO (%) | Electrodes inserted | |
| Ac | 0.36 | 0.34 |
|
| 0.19 |
| Outer wall length 7200 | 0.19 | 0.19 |
|
| 0.27 |
| Degree of insertion | 0.23 | 0.22 | -0.08 | -0.09 | |
| Electrodes inserted on X-ray | 0.26 | 0.24 | 0.054 | -0.08 | |
Figure 3A plot showing a statistically significant correlation between the Ac value and the AZBio sentence score (a) and CNC word score (b) in the Flex 28 group as well as the correlation between the outer wall cochlear duct length and the AZBio sentence score (c) and CNC word score (d) at 6 months.