| Literature DB >> 31238402 |
William H Cook1, James McKelvie1, Henry B Wallace1, Stuti L Misra1.
Abstract
Purpose: To evaluate the agreement of selected higher order aberration measurements between aberrometers based on three different wavefront technologies.Entities:
Keywords: Higher order aberrations; LASIK; wavefront aberrometry
Mesh:
Year: 2019 PMID: 31238402 PMCID: PMC6611280 DOI: 10.4103/ijo.IJO_1464_18
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Bland-Altman analysis of total ocular aberrations from different aberrometers
| Aberration | Mean Diff. | LoA | |
|---|---|---|---|
| Vertical coma | |||
| iDesign - OPD-Scan III | 0.026 | −0.025 to+0.078 | |
| iDesign - Zywave | 0.034 | 0.152 | −0.184 to+0.252 |
| OPD-Scan III - Zywave | 0.008 | 0.687 | −0.182 to+0.199 |
| Horizontal coma | |||
| iDesign - OPD-Scan III | 0.002 | 0.768 | −0.067 to+0.071 |
| iDesign - Zywave | −0.001 | 0.969 | −0.204 to+0.202 |
| OPD-Scan III - Zywave | −0.003 | 0.879 | −0.188 to+0.182 |
| Vertical trefoil | |||
| iDesign - OPD-Scan III | 0.033 | −0.078 to+0.143 | |
| iDesign - Zywave | −0.039 | 0.118 | −0.266 to+0.188 |
| OPD-Scan III - Zywave | −0.072 | −0.370 to+0.255 | |
| Oblique trefoil | |||
| iDesign - OPD-Scan III | −0.040 | 0.067 | −0.233 to+0.154 |
| iDesign - Zywave | 0.019 | 0.124 | −0.091 to+0.129 |
| OPD-Scan III - Zywave | 0.058 | −0.129 to+0.245 | |
| Primary spherical | |||
| iDesign - OPD-Scan III | 0.022 | −0.074 to+0.118 | |
| iDesign - Zywave | 0.030 | −0.042 to+0.102 | |
| OPD-Scan III - Zywave | 0.008 | 0.551 | −0.109 to+0.124 |
Diff.: mean difference (μm); LoA: limits of agreement (μm)
Figure 1Agreement of selected ocular higher order aberration measurements (in micrometers) obtained by three wavefront aberrometers: Zywave, iDesign, and OPD-Scan III (OPD-Scan). Central line = mean of the difference between the two methods; peripheral lines = 95% limits of agreement
Bland-Altman analysis of corneal aberrations from GALILEI™ G2 and OPD-Scan III
| Aberration | Mean Diff. | LoA | |
|---|---|---|---|
| Vertical coma | 0.018 | 0.136 | −0.091 to+0.127 |
| Horizontal coma | −0.025 | −0.106 to+0.055 | |
| Vertical trefoil | −0.017 | 0.194 | −0.133 to+0.100 |
| Oblique trefoil | −0.022 | 0.302 | −0.219 to+0.175 |
| Primary spherical | −0.043 | −0.094 to+0.008 |
Diff.: mean difference (μm); LoA: limits of agreement (μm)
Figure 2Agreement of selected corneal higher order aberration measurements (in micrometers) obtained by two wavefront aberrometers using Bland–Altman plots: GALILEI™ G2 (GALILEI) and OPD-Scan III (OPD-Scan). Central line = mean of the difference between the two methods; peripheral lines = 95% limits of agreement