| Literature DB >> 31223408 |
Simeon-Pierre Choukem1,2,3, Colette Sih1,2, Daniel Nebongo2, Philomene Tientcheu4, André-Pascal Kengne5.
Abstract
INTRODUCTION: capillary glucose measurement using point-of-care glucometers is an essential part of diabetes care. We determined the technical accuracy, clinical accuracy and precision of commonly available glucometers against standard spectrophotometry in Cameroon.Entities:
Keywords: Cameroon; clinical accuracy; glucometer; precision; technical accuracy
Mesh:
Substances:
Year: 2019 PMID: 31223408 PMCID: PMC6561013 DOI: 10.11604/pamj.2019.32.118.15553
Source DB: PubMed Journal: Pan Afr Med J
absolute difference between each glucometer and reference method results
| Glucometer | Mean difference | 95% CI | Mean bias (%) | P |
|---|---|---|---|---|
| Accu-Chek® Active | +17.4 | 15.4 - 19.5 | 18.7 | <0.0001 |
| CodeFreeTM | +26.5 | 21.3 - 31.8 | 29.1 | <0.0001 |
| MylifeTMPuraTM | +15.5 | 11.7 - 19.3 | 16.1 | <0.0001 |
| OneTouch® Ultra®2 | +12.0 | 9.2 - 14.9 | 13.8 | <0.0001 |
CI: confidence interval
correlations between the reference value and glucometer-based values,and regression equations
| Glucometer | Overall (95% CI) | Women (95% CI) | Men (95% CI) | p | Regression equation |
|---|---|---|---|---|---|
| Accu-chek® Active | 0.96 (0.94-0.97) | 0.93 (0.88-0.96) | 0.97 (0.95-0.98) | 0.03 | 0.9117×AccuChek - 6.9173 |
| CodeFreeTM | 0.87 (0.81-0.91) | 0.73 (0.57-0.83) | 0.95 (0.92-0.97) | <0.001 | 0.85690×CodeFree - 4.52115 |
| MylifeTMPuraTM | 0.97 (0.95-0.98) | 0.94 (0.90-0.97) | 0.98 (0.96-0.99) | 0.01 | 1.05663×MyllifePura- 15.98433 |
| OneTouch® Ultra®2 | 0.94 (0.91-0.96) | 0.87 (0.79-0.92) | 0.96 (0.94-0.98) | <0.001 | 0.90853×OneTouch- 2.93158 |
Women vs. men; CI: confidence interval
Figure 1Bland-Altman plots for agreement between glucometers; (A) Accu-Chek® Active; (B) CodeFreeTM; (C) MylifeTMPuraTM; (D) OneTouch® Ultra®2) and the reference method, glucometer values minus reference value for each participant (y-axis) is plotted against the mean of the two measurements (x-axis) and are represented by the black dots; the horizontal doted blue line through zero is the line of perfect agreement between the two measurements; the parallel solid black line is the mean bias; the green colour band around the solid black line is the standard error around the mean bias estimates, while the shaded area represents the 95% confidence intervals; the linear curve of best fit is also shown (broken oblique blue line)
proportion of values respecting ISO 2013 criteria for each glucometer
| GLUCOMETER | ≤ 100 mg/dL | >100 mg/dL |
|---|---|---|
| Accu-Chek® Active | 34/75 (45.3%) | 14/32 (43.8%) |
| CodeFreeTM | 34/76 (44.7%) | 16/32 (50.0%) |
| MylifeTMPuraTM | 54/76 (71.1%) | 27/30 (90.0%) |
| OneTouch® Ultra®2 | 49/75 (65.3%) | 20/30 (66.7%) |
Figure 2Parke’s consensus error grid analysis for glucometers; (A) Accu-Chek® Active; (B) CodeFreeTM; (C): MylifeTMPuraTM; (D): OneTouch® Ultra®2), Zone A: results given by the glucometer are clinically accurate within+/-20% of the reference; zone B: the error of the results given by the glucometer is above 20%, but would lead to minor or no treatment error; zone C: results given by the glucometer would begin to lead to treatment decisions opposite to that called for by the real blood glucose levels; zone D: results given by the glucometer lead to a failure to detect and treat errors; zone E: results given by the glucometer lead to erroneous treatment of hypo or hyperglycemia