Performance of the Roche Accu-Chek active glucose meter to screen for gestational diabetes mellitus using fasting capillary blood.

Gurdeep S Dhatt1, Mukesh M Agarwal, Yusra Othman, Satish C Nair.   

Abstract

BACKGROUND: Point-of-care (POC) blood glucose measurement using glucose meters is used by diabetes patients to mange their disease. POC glucose testing also is also used in tight glycemic control protocols and as a screening tool for diabetes. We report the performance and effectiveness of the Accu-Chek® Active (Roche Diagnostics GmbH, Mannheim, Germany) glucose meter to screen for gestational diabetes mellitus (GDM) using blood fasting capillary glucose (FCG).
METHODS: To screen for GDM, 1,465 pregnant women underwent an oral glucose glucose tolerance test. Correlation between the FCG and fasting plasma glucose (FPG) levels was determined by Passing and Bablok regression analysis. Total error (TE) of the glucometer was ascertained using the Bland-Altman method with the DXC-800 analyzer (Beckman-Coulter Instruments, Brea, CA) as the reference method. The area under the receiver operator characteristic curve was used to analyze the performance of the FCG to predict GDM.
RESULTS: FPG and FCG identified 361 (24.6%) and 338 (23%) women as having GDM, respectively. The Bland-Altman TE at 95% limits of agreement was -11.1% to 10.8%. The area under the receiver operator characteristic curve was 0.953 (95% confidence interval 0.943 to 0.964).
CONCLUSIONS: The Roche Accu-Chek Active glucometer meets analytical and clinical quality requirements. A TE of±15% is acceptable for glucose meters used in ambulatory care, including home self-monitoring of blood glucose, and different TE targets should be set for acute critical care settings.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21864017     DOI: 10.1089/dia.2011.0097

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  7 in total

1.  Hematocrit interference of blood glucose meters for patient self-measurement.

Authors:  Sanja Ramljak; John Paul Lock; Christina Schipper; Petra B Musholt; Thomas Forst; Martha Lyon; Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

2.  Higher risk of hyperglycemia with greater susceptibility in females in chronic arsenic-exposed individuals in Bangladesh.

Authors:  Sudip Kumar Paul; Md Shofikul Islam; M M Hasibuzzaman; Faruk Hossain; Adiba Anjum; Zahangir Alam Saud; Md Mominul Haque; Papia Sultana; Azizul Haque; Klara Biljana Andric; Aminur Rahman; Md Rezaul Karim; Abu Eabrahim Siddique; Yeasir Karim; Mizanur Rahman; Hideki Miyataka; Lian Xin; Seiichiro Himeno; Khaled Hossain
Journal:  Sci Total Environ       Date:  2019-03-04       Impact factor: 7.963

3.  Accuracy of capillary blood glucose for 50-g glucose challenge test for gestational diabetes screening.

Authors:  Dittakarn Boriboonhirunsarn; Amporn Robkhonburi; Maria Asad-Dehghan
Journal:  Diabetol Int       Date:  2022-02-07

4.  Gestational diabetes mellitus: Screening with fasting plasma glucose.

Authors:  Mukesh M Agarwal
Journal:  World J Diabetes       Date:  2016-07-25

5.  Accuracy and precision of four main glucometers used in a Sub-Saharan African Country: a cross-sectional study.

Authors:  Simeon-Pierre Choukem; Colette Sih; Daniel Nebongo; Philomene Tientcheu; André-Pascal Kengne
Journal:  Pan Afr Med J       Date:  2019-03-14

Review 6.  Gestational Diabetes Mellitus-Innovative Approach to Prediction, Diagnosis, Management, and Prevention of Future NCD-Mother and Offspring.

Authors:  H David McIntyre; Anil Kapur; Hema Divakar; Moshe Hod
Journal:  Front Endocrinol (Lausanne)       Date:  2020-12-03       Impact factor: 5.555

7.  Polysaccharides from Enteromorpha prolifera Improve Glucose Metabolism in Diabetic Rats.

Authors:  Wenting Lin; Wenxiang Wang; Dongdong Liao; Damiao Chen; Pingping Zhu; Guoxi Cai; Aoyagi Kiyoshi
Journal:  J Diabetes Res       Date:  2015-08-10       Impact factor: 4.011

  7 in total