INTRODUCTION: The aim of this study was to investigate the postoperative prediction error measured by optical biometry and acoustic biometry in eyes after phacovitrectomy for rhegmatogenous retinal detachment (RRD) with no macular involvement.Entities:
Year: 2019 PMID: 31191998 PMCID: PMC6525934 DOI: 10.1155/2019/5964127
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Patient demographic and clinical characteristics.
| RRD group | Control group |
| |
|---|---|---|---|
| Males | 32 | 21 | |
| Females | 17 | 12 | 0.548 |
| Age (years) | 62.6 ± 7.5 (48–82) | 74.1 ± 7.1 (57–85) |
|
|
| 24.32 ± 0.90 (22.76–25.99) | 23.65 ± 0.86 (22.26–25.77) | 0.00026 |
|
| 24.05 ± 0.89 (22.58–25.75) | 23.36 ± 0.85 (22.03–25.77) | 0.000114 |
| K-average (Dpt) | 43.67 ± 1.36 (41.19–47.38) | 44.48 ± 1.37 (41.54–47.05) | 0.002276 |
| ACD (mm) | 3.42 ± 0.33 (2.57–4.09) | 3.03 ± 0.42 (2.26–4.08) |
|
Note. Data are expressed as mean ± standard deviation and range. The proportion of male patients was analyzed by the chi-squared test. The values for age, axial length, K-average, and ACD were analyzed by the Mann–Whitney U test. ACD, anterior chamber depth; RRD, rhegmatogenous retinal detachment.
Comparison of refractive prediction error between the RRD group and the control group.
| Actual postoperative refraction | Predictive refraction | Refraction error | |||
|---|---|---|---|---|---|
| Optic biometry | Acoustic biometry | Optic biometry | Acoustic biometry | ||
| RRD group | −1.50 ± 1.72 | −1.06 ± 1.40 | −0.68 ± 1.51 | −0.44 ± 0.77 | −0.81 ± 0.75 |
| Control group | −0.39 ± 1.12 | −0.67 ± 0.90 | −0.18 ± 0.86 | 0.27 ± 0.71 | −0.21 ± 0.64 |
Note. Data are expressed as mean ± standard deviation. RRD, rhegmatogenous retinal detachment.
Comparison of predicted refraction error measured by acoustic biometry and optic biometry.
| Refractive error (optic biometry) | Refractive error (acoustic biometry) |
| Difference | 95% CI | |
|---|---|---|---|---|---|
| RRD group | −0.44 ± 0.77 | −0.81 ± 0.75 |
| 0.37 | 0.22–0.51 |
| Control group | 0.27 ± 0.71 | −0.21 ± 0.64 |
| 0.48 | 0.32–0.65 |
Note. Data are expressed as mean ± standard deviation. The values for refractive error were analyzed using the paired t-test. CI, confidence interval; RRD, rhegmatogenous retinal detachment.
Comparison of predicted refraction error measured by acoustic biometry and optic biometry between the RRD and control groups.
| RRD group | Control group |
| Difference | 95% CI | |
|---|---|---|---|---|---|
| Refractive error (optic biometry) | −0.44 ± 0.77 | 0.27 ± 0.71 |
| −0.71 | −0.41 to −1.01 |
| Refractive error (acoustic biometry) | −0.81 ± 0.75 | −0.21 ± 0.64 |
| −0.60 | −0.32 to −0.88 |
Note. Data are expressed as mean ± standard deviation. The values for refractive error were analyzed using Student's t-test. CI, confidence interval; RRD, rhegmatogenous retinal detachment.