Milad Hosseinipour1,2, Rajesh Gupta3, Mark Bonnell4, Mohammad Elahinia1. 1. Dynamic and Smart Systems Laboratory, The University of Toledo, Toledo, OH, USA. 2. Department of Mechanical Engineering, Virginia Polytechnic Institute and State University, Blacksburg, VA, USA. 3. Cardiovascular Medicine Division, The University of Toledo Medical Center, Toledo, OH, USA. 4. Cardiothoracic Surgery Division, The University of Toledo Medical Center, Toledo, OH, USA.
Abstract
A detailed survey of the current trends and recent advances in rotary mechanical circulatory support systems is presented in this paper. Rather than clinical reports, the focus is on technological aspects of these rehabilitating devices as a reference for engineers and biomedical researchers. Existing trends in flow regimes, flow control, and bearing mechanisms are summarized. System specifications and applications of the most prominent continuous-flow ventricular assistive devices are provided. Based on the flow regime, pumps are categorized as axial flow, centrifugal flow, and mixed flow. Unique characteristics of each system are unveiled through an examination of the structure, bearing mechanism, impeller design, flow rate, and biocompatibility. A discussion on the current limitations is provided to invite more studies and further improvements.
A detailed survey of the current trends and recent advances in rotary mechanical circulatory support systems is presented in this paper. Rather than clinical reports, the focus is on technological aspects of these rehabilitating devices as a reference for engineers and biomedical researchers. Existing trends in flow regimes, flow control, and bearing mechanisms are summarized. System specifications and applications of the most prominent continuous-flow ventricular assistive devices are provided. Based on the flow regime, pumps are categorized as axial flow, centrifugal flow, and mixed flow. Unique characteristics of each system are unveiled through an examination of the structure, bearing mechanism, impeller design, flow rate, and biocompatibility. A discussion on the current limitations is provided to invite more studies and further improvements.
Entities:
Keywords:
Assistive technology; biomedical devices; heart failure; life support systems; mechanical circulatory support; orthotics; rehabilitation devices; ventricular assistive device
Mechanical circulatory support (MCS) is becoming a realistic alternative to heart
transplant for terminal heart failure, after exhaustion of medical and conventional surgical
treatments. Each year, there are more patients who need long-term cardiac care. Meanwhile,
the number of successful transplants remains stagnant.[1] Several modalities exist for providing support to heart-failurepatients as a[2]:Bridge to decision (BTD)Bridge to recovery (BTR)Bridge to transplant (BTT)Destination therapy (DT)Although each device has its unique characteristics, most of the available systems can be
classified into two categories: total artificial hearts (TAHs) and ventricular assistive
devices (VADs). TAHs replace a portion or all of the heart, similar to a prosthesis, with a
device that achieves systole and diastole by filling and emptying artificial chambers in
repetitive cycles. VADs, on the other hand, are designed to augment the function of a
failing heart as an orthosis without replacing the biological organ. They can provide
support to either left (LVAD), right (RVAD), or both ventricles (BiVAD).Based on the flow regime, VADs can be categorized as: positive displacement (PD-VADs) or
continuous flow (CF-VADs). In PD-VADs, the pump is connected to an external driver
(pneumatic, hydraulic, or electric) that provides alternating pressure and vacuum to fill
and empty the pump chambers. CF-VADs generally provide cardiac support by providing a
continuous output flow that increases the arterial blood flow and pressure. Angular momentum
is generated through rotation of a miniaturized impeller that is further converted to linear
flow. The speed of rotation has a direct relation with the output flow rate.
Continuous flow
While PD-VADs maintain the physiological nature of blood flow, they often have limited
lifetime and low reliability due to mechanical failure in diaphragms and valves. Today,
CF-VADs comprise over 98% of LVAD implantations in USA,[3] as they often lead to better outcomes.[4] They have shown promising results due to[5]:Lower power consumptionLower hemolysis riskLower infection riskLower sound levelLess hospitalization and equipment cost (in long-term)Moreover, these devices are more compact (25 cc vs. 150 cc) that makes them more suitable
for patients with small body-surface area (BSA) and children. The smaller size makes the
thoracic implantation achievable, which is a better site than the usual upper abdomen.[5]LVAD recipients may develop right ventricular failure during or shortly after device
implantation. This is partly contributed to rapid unloading of the left ventricle, which
may cause deviation of the interventricular septum towards the left ventricle, thus
reducing right ventricular efficiency.[6] The very first symptom of right ventricular dysfunction in these patients is often
the inability of the LVAD to fill the ventricle that leads to insufficient flow rate. The
limited filling is exacerbated when the right heart failure is followed by an elevated
transpulmonary pressure gradient or high pulmonary vascular resistance. Until recently,
TAHs were the only option for severe biventricular failure. Modern CF-VADs have shown
promising results when two devices are implanted as a BiVAD.[7] Patients with BiVAD will not be as dependent on the MCS as patients with TAH.
Device malfunctions are easier to manage with a BiVAD compared with a TAH. Moreover, BiVAD
implantation does not eliminate the possibility of myocardial recovery.
Flow regimes
The overall design of the pump including the flow regime, bearings, blood-contacting
surfaces, and flow path define the hemolytic and hemodynamic performance of the device.
Three types of continuous-flow pumps have been proven in providing durable support:
centrifugal flow pump (CFP), axial flow pump (AFP), and mixed flow pump (MFP).[8] In CFPs, the outlet is positioned tangentially to the pump housing, and rotation of
the impeller causes the centrifugal force to suck the blood from the inlet and pump it
through the outlet. In AFPs, the outlet is co-linear with the rotating section, and the
impeller blades are shaped to accelerate the blood both rotationally and axially. An MFP
combines both flow types, and impeller blades have a convex profile shape. The degree of
convexity varies based on the dominant flow model. Figure 1 provides an example of impeller design and
flow regime for each category.
Figure 1.
Different flow regimes in rotary ventricular assistive devices (reprinted with
permission[9,10]). (a) Centrifugal
flow, (b) axial flow and (c) mixed flow.
Different flow regimes in rotary ventricular assistive devices (reprinted with
permission[9,10]). (a) Centrifugal
flow, (b) axial flow and (c) mixed flow.When normalized to the pump size, CFPs are capable of generating a typical pressure
gradient, ΔP, over a wide spectrum of flow rates, Q. This is often called flat head
curve.[11] In contrast, AFPs have a steep head curve where there is an almost linear relation
between Q and ΔP.[12] MFPs tend to extend the flat region by providing high ΔP at high Q.[13]
Figure 2 compares pressure–flow
relationships of a typical CFP, AFP, and MFP at the same speed.
Figure 2.
Typical P–Q results for centrifugal, axial, and mixed CF-VADs at the same
speed.[11,14–16]
Typical P–Q results for centrifugal, axial, and mixed CF-VADs at the same
speed.[11,14-16]The flat head curve means greater sensitivity of the flow–pressure relationship that
results in greater change in Q for any given ΔP across the inlet and outlet of the pump. During a cardiac cycle
that the pressure head changes from a min Q at diastole to a
max Q in systole (e.g. 40 mmHg to 80 mmHg), the pump
exhibits larger change in flow rate (e.g. 0 lpm to 10 lpm), emulating a pulsatile flow.
This greater ΔQ increases the accuracy of the estimated flow
from pump speed and power.
Flow control
In CFPs, the motor current is used as a sensorless index of pump flow and a virtual index
of the left ventricular pressures during the cardiac cycle. This is based on the linear
current–flow relationship across the full range of operating pump flows in these pumps.
This information allows centrifugal pump controllers to monitor the pump flow and the
degree of left ventricle unloading by simply monitoring the motor current or power.[11]The correlation between flow and current in AFPs and MFPs is not nearly linear over the
full operating flow range of the pump. Therefore, motor current offers less accuracy for
flow estimation. Without a flow sensor, AFPs often cannot report accurate flow, especially
at low rates.[12] For example, the HeartMate II axial flow pump does not display flows below 3.0 lpm.
The HeartAssist 5 (ReliantHeart) uses an ultrasonic flow probe incorporated in its outlet
conduit to measure the flow rate directly.Regardless of the algorithm, accurate flow estimation depends on viscosity of patient's
blood. Currently, only one device (HeartWare HVAD) has adjustable hematocrit setting for
adjusting the blood viscosity in flow estimations. Variable viscosity and non-linear
relations between flow rate, current, speed, and power, hinder the automatic control
algorithms of CF-VADs.Although CF-VADs do not generate pulsation, the pressure difference between the pump
inlet and outlet is pulsatile since the actual heart is still beating. For some devices,
backflow may happen at lower speeds (less than 1000 rpm).[17] It has also been reported that the pressure variation can diminish at higher speeds
(more than 10,000 rpm) resulting in a non-pulsatile flow.[17] Some researchers have attempted to vary the CF-VAD RPMs throughout the cardiac
cycle to achieve some measure of pulsatility.[18] The usefulness of such control algorithms remains to be determined.Moreover, a CF-VAD is blind to the absolute pressures at the inlet and outlet ports, and
responds only to the ΔP. This can lead to inlet cannula
suction if the pump increases ΔP at low left ventricular
volume (e.g. hypovolemia). As discussed earlier, AFPs (and MFPs to some extent) have a
steep Q–P behavior that leads
to higher inlet suction at lower flow rate. Higher suction increases the risk of sucking
the ventricular wall in around the inlet cannula. This means that a flow control algorithm
should include multiple parameters, including (but not limited to):Pump speedPressure differentialBlood viscosityAbsolute inlet pressureAbsolute outlet pressureVascular and valvular resistancesSystemic, ventricular, atrial, and aortic compliancesInlet, outlet, and pump resistancesAortic inertanceInlet, outlet, and pump inertancesThe controller state variables are left ventricular pressure, left atrial pressure,
arterial pressure, aortic pressure, total flow, and pump flow.
Bearings
Various bearing mechanisms have been investigated for supporting miniaturized impellers
in CF-VADs. Summarized below and illustrated in Figure 3 are those currently in use.
Figure 3.
Bearing mechanisms in rotary ventricular assistive devices (reprinted with
permission[12,19]).
(a) Mechanical pivot, (b) hydrodynamic radial (left) and thrust (right), (c)
permanent magnet (combined with hydrodynamic bearing) and (d) electromagnetic.
Bearing mechanisms in rotary ventricular assistive devices (reprinted with
permission[12,19]).
(a) Mechanical pivot, (b) hydrodynamic radial (left) and thrust (right), (c)
permanent magnet (combined with hydrodynamic bearing) and (d) electromagnetic.
Mechanical pivot bearing
In early versions of CF-VADs, the rotor was supported mechanically by journal and
thrust bearings.[20] These bearings evolved into low-wear pivot bearings made of precision ceramic
components. Figure 3(a) is the
HeartMate II (Thoratec) inlet bearing ball and cup. This bearing set was explanted from
a pump that was in operation for 4.4 years and demonstrates low wear.[19] Although these bearings are small, they are potential sites for thrombus and
fibrin deposition, which is not present in other bearing designs that completely suspend
the rotor.[12] The concentration of hydrodynamic loads on these bearings, especially at stress
concentration points, makes them theoretically susceptible to wear and fatigue.
Hydrodynamic radial and thrust bearing
One of the biggest breakthroughs in the development of MCS systems was the introduction
of contactless bearings in 1980s. It was first achieved by filling the rotor bearing
with cross-linked blood or other biocompatible materials.[21,22] Hydrodynamic bearing has been utilized
in both radial and thrust configurations. In either configuration, a fluid film pressure
acts to separate the rotating surfaces from the stationary base. Hydrodynamic bearings
provide no life-limiting contact while in motion, however, surfaces will touch when
stationary (start and end). Moreover, they apply more shear stress on blood cells
especially at the boundary layer. Higher shear stress increases the risk of hemolysis.
An example of hydrodynamically suspended pump is HeartAsssit 5 (ReliantHeart). In Figure 3(b), the magnitude and
direction of these fluid forces are represented by the height and direction of the
arrows.
Permanent magnet bearing
Using repelling neodymium-boron-iron magnets on the rotor and housing allows for
suspending the rotor. Permanent magnet bearings allow for a larger gap between the
static motor armature and the rotor, while eliminating the need for mechanical bearings,
lubrication, sealing, and purging fluid. However, the magnetic forces are dependent on
the instantaneous position of the rotor. Therefore, the magnetic bearings are often used
in conjunction with hydrodynamic or electromagnetic bearings. An example of permanent
magnet bearing combined with hydrodynamic bearing is HeartWare HVAD (HeartWare
International), as illustrated in Figure 3(c). The rotor is aligned and separated from the conical stationary
center shaft of the pump housing using magnetic bearing. A thin film of blood generates
pressure at the tapered surface of the impeller (red line) to create hydrodynamic elevation.[12] Stator coils in the housing are supplied with electric currents having a
frequency and amplitude adjusted to the blood pressure at the pump inlet.[23-25]
Electromagnetic bearing
Electromagnetic bearings control three degrees of freedom of the impeller: axial,
radial, and tilt. Either axis can be active or passive. In active mode, the repelling
force is adjusted by changing the supplied current to the electromagnets based on the
instantaneous position or other feedback errors. In passive mode, the supplied current
to the electromagnets is constant. The electromagnetic force adjusts automatically based
on the position itself. An example of electromagnetic suspension is INCOR (Berlin Heart)
(Figure 3(d)). INCOR uses
active control for the axial position, and passive control for radial and tilt.
Electromagnetic bearings eliminate life-limiting contacts and reduce stress on blood
cells at the cost of added components, electronics, and controllers. If a momentary drop
in control signal occurs, pump failure will be inevitable. Therefore, these bearings are
often backed up with a bearing that does not require active control.
Biocompatibility
The common practice in manufacturing VADs is highly polished blood-contacting surfaces
made of biocompatible materials such as titanium. The sintered titanium surface may be
textured to reduce anticoagulation requirements and thromboembolism. To decrease the
incidence of thrombus formation, some devices coat the blood-contacting surfaces with
materials that reduce platelet activation and adhesion. Common coating materials used in
VADs are listed below.
Carmeda
BioActive Surface is a heparin-based, thrombo-resistant coating developed for medical
devices in contact with blood. The coating is applied by stepwise deposition of cationic
and anionic polymers (layer-by-layer) on the device surface, ending with covalent
end-point attachment of heparin, a well-established and widely used anticoagulant. What
is unique to Carmeda BioActive Surface is the retained functional activity of the
immobilized heparin, specifically the capacity to bind to the coagulation inhibitor
antithrombin (AT). This natural inhibitor in blood acts by forming a complex with
activated coagulation factors (enzymes), thereby neutralizing their pro-coagulant
activity. The binding of AT to heparin increases this inhibitory capacity drastically,
thus converting AT from a slow to a highly potent coagulation inhibitor. Inactive
complexes formed on the immobilized heparin are released and swept away from the surface
by the blood flow. Hence, the end-point attached heparin is not consumed but remains
active and available for further inhibition. In addition to downregulating the
coagulation, the advantages of the Carmeda BioActive Surface include platelet
compatibility and reduced activation of natural defense mechanisms.[26]
Trillium
It is a polymer coating with applied heparin to the blood-contacting surfaces. A
priming layer is bonded to the blood-contacting surface. A hydrophilic functional layer
with heparin is deposited to the prime coat and provides the key endothelial-like
behavior for the surface of the ventricular assistive device. This functional level
includes three primary elements:Non-leaching heparin molecules are covalently bonded into the surface to act like
heparin sulfate in vascular endothelium.Sulfate and sulfonate groups are incorporated into the Trillium biosurface layer
to emulate the negative charge of the vascular endothelium. These negatively
charged sulphonated polymers repel platelets (that are also negatively charged)[27] and inhibit thrombin by attaching to AT (similar to heparin).[28]As a third functional layer, polyethylene oxide polymer is deposited on the
surface as a hydrophilic molecule to create an insulating water layer between
blood and artificial surface to resist cell adhesion and protein deposition.These layers lead to a reduction in platelet activation measured by β-TG. With Trillium, a strict anticoagulation protocol should
still be followed and routinely monitored.[29]To provide a systematic review on implantable CF-VADs, eight prominent technologies are
categorized based on their flow regimes. A summary of these devices is given in Table 1. Many other MCS systems
have been introduced since 1930s, which are either now defunct or remain
investigational. A summary of these devices is provided in Table 2 as a reference. This review is intended
to discuss the devices in Table
1 from an engineering point of view, rather than clinical reports. A discussion
on the current limitations is provided to invite further developments. Where possible,
expert advice has been sought from the manufacturers to provide the most accurate and
up-to-date information.
Table 1.
Summary of commissioned rotary mechanical circulatory support systems.
Device
Manufacturer
Flow
Side
Location
Suspension and bearings
Size
Output
Approvals
HeartAssist 5
ReliantHeart
Axial
LV
Intracorporeal
Hydrodynamic (purged)
30(D) × 71(L)
5–10 lpm
FDA (BTT adult) trials, FDA (BTT pediatric), EU
(BTT, DT)
HeartMate II
Thoratec
Axial
LV
Intracorporeal
Mechanical (blood immersed)
40(D) × 60(L): 63 mL
3–10 lpm
FDA (BTT, DT), EU (BTT, DT)
INCOR
Berlin Heart
Axial
LV
Intracorporeal
Electromagnetic
30(D) × 114(L)
5 lpm
FDA (BTT) trials, EU (BTT, DT)
Impella
Abiomed
Axial
LV, RV, BV
Transvalvular
Magnetic and hydrodynamic
12 Fr to 21 Fr
2.5–5 lpm
FDA (6 h), EU
HeartWare HVAD
HeartWare
Centrifugal
LV, RV, BV
Intracorporeal
Electromagnetic and hydrodynamic
50 mL
10 lpm
FDA (BTT), FDA (DT) trials, EU (BTT, DT)
HeartMate III
Thoratec
Centrifugal
LV
Intracorporeal
Electromagnetic
69(D) × 30(H): 50 mL
13 lpm
FDA trials, EU
CentriMag
Thoratec
Centrifugal
LV, RV, BV
Extracorporeal
Magnetic
31 mL
9.9 lpm
FDA (6 days LVAD and 30 days RVAD), EU (30
days)
PediMag
Thoratec
Centrifugal
LV, RV, BV
Extracorporeal
Magnetic
14 mL
1.5 lpm
FDA (6 days), EU (30 days)
EvaHeart
EvaHeart
Centrifugal
LV
Intracorporeal
Hydrodynamic (purged)
55(D) × 64(H): 132 mL
12 lpm
FDA (BTT) trials
BPX-80
Medtronic
Centrifugal
LV
Extracorporeal
Mechanical
79(D) × 66(H): 86 mL
10 lpm
FDA (6 h), EU
BP-50
Medtronic
Centrifugal
LV
Extracorporeal
Mechanical
48 mL
1.5 l pm
FDA (6 h), EU
Synergy
HeartWare
Mixed
LV, RV, BV
Intracorporeal
Magnetic and hydrodynamic
12(D) × 49(L): 1.5 mL
0.3–6 lpm
FDA (BTT, BTR) trials, EU (BTT, BTR)
BTD: bridge to decision; BTR: bridge to recovery; BTT: bridge to transplant;
DT: destination therapy; VAD: ventricular assistive device; TAH: total
artificial heart; FDA: Food and Drug Administration; BV: biventricular; LV: left
ventricle; RV: right ventricle.
Table 2.
Summary of other rotary mechanical circulatory support systems.
Device
Manufacturer
Flow
Side
Location
Suspension and bearings
Size
Output
Approvals
Jarvik 2000
Jarvik Heart
Axial
LV
Intracorporeal
Mechanical (blood immersed)
25(D) × 55(L)
3–7 lpm
FDA (BTT, DT) trials, EU (BTT, DT)
DuraHeart
Terumo Heart
Centrifugal
LV
Intracorporeal
Magnetic and hydrodynamic
72(D) × 45(H): 196 mL
2–10 lpm
FDA (BTT, DT) trials discontinued, EU (BTT,
DT)
CorAide
Arrow
Centrifugal
LV
Intracorporeal
Magnetic and hydrodynamic
84 mL
2–8 lpm
Pre-clinical testing
DexAide
Cleveland Clinic
Centrifugal
RV
Intracorporeal
Magnetic and hydrodynamic
69 mL
2–6 lpm
Pre-clinical testing
TandemHeart
CardiacAssist
Centrifugal
LV, RV, BV
Extracorporeal
Hydrodynamic (purged)
10 mL
4 lpm
FDA trials, Canada (Class 4)
PediaFlow
PediaFlow Consortium
Mixed
LV, RV, BV
Intracorporeal
Electromagnetic
28(D) × 51(L)
0.3–1.5 lpm
In development
MiFlow
HeartWare
Mixed
LV
Intracorporeal
Electromagnetic
22 mL
2–6 lpm
Clinical trials expected
BTD: bridge to decision; BTR: bridge to recovery; BTT: bridge to transplant;
DT: destination therapy; VAD: ventricular assistive device; TAH: total
artificial heart; FDA: Food and Drug Administration; BV: biventricular; LV: left
ventricle; RV: right ventricle.
Summary of commissioned rotary mechanical circulatory support systems.BTD: bridge to decision; BTR: bridge to recovery; BTT: bridge to transplant;
DT: destination therapy; VAD: ventricular assistive device; TAH: total
artificial heart; FDA: Food and Drug Administration; BV: biventricular; LV: left
ventricle; RV: right ventricle.Summary of other rotary mechanical circulatory support systems.BTD: bridge to decision; BTR: bridge to recovery; BTT: bridge to transplant;
DT: destination therapy; VAD: ventricular assistive device; TAH: total
artificial heart; FDA: Food and Drug Administration; BV: biventricular; LV: left
ventricle; RV: right ventricle.
Axial flow devices
HeartAssist 5
HeartAssit 5 (ReliantHeart, Inc., Houston, TX, USA) is the modern version of DeBakey/NASA
VAD that has been under development since 1988. It consists of an inflow cannula, stator
in the pump housing, flow straightener, rotor, and diffuser. This electromagnetically
driven pump measures 71 mm long by 30 mm diameter and weighs 92 g. It pumps blood flow of
5 to 6 lpm at an average speed of 10,000 rpm to a maximum of 10 lpm at 12,500 rpm.[30] Due to its small size, HeartAssist 5 may be implanted above the diaphragm, so there
is no need to create a pocket for the pump. Much optimization has been done on the inlet
angle, outlet angle, axial and radial clearances of blades on flow straightener, rotor
(impeller and inducer), and diffuser to minimize the risk of hemolysis and platelet
activation. The HeartAssist 5 Pediatric VAD uses a 140° inflow cannula and a 60 mm outflow
graft for supporting pediatric patients: as young as four to six years old, BSA larger
than 0.7 m2 and smaller than 1.5 m2, weight equal or greater than
18 kg. For adults, however, HeartAssist 5 Adult VAD uses an 115° inflow cannula and 90 mm
outflow graft. Figure 4 compares
the pediatric and adult versions.
Figure 4.
HeartAssist 5 pump (Courtesy of ReliantHeart, Inc.). (a) Adult and (b) low
body-surface-area (pediatric).
HeartAssist 5 pump (Courtesy of ReliantHeart, Inc.). (a) Adult and (b) low
body-surface-area (pediatric).A plurality of magnets is placed in each impeller blade that interact with the axially
adjusted stator.[31] The straightener and impeller are designed such that they eliminate any area of
blood stagnation and prevent thrombus formation. Moreover, there is almost no gap between
the body of flow straightener and front hub of the impeller to eliminate flow
recirculation zones in contact with stationary parts, which again may be
thrombogenic.[21,32] In addition to the pump,
the LVAD system comprises a controller module, an ultrasonic flow probe, batteries, a
battery charger, and a data acquisition system for home or clinical support. A cable
powers the pump percutaneously and sends the feedback to the HeartAssist 5 Conquest
Controller. A unique feature of this device is continuous flow rate measurement using an
ultrasonic flow probe around the outflow graft (Figure 5).
Figure 5.
HeartAssist 5 ventricular assist system (Courtesy of ReliantHeart, Inc.).
HeartAssist 5 ventricular assist system (Courtesy of ReliantHeart, Inc.).All blood-contacting surfaces are made of highly polished titanium. In an attempt to
decrease the incidence of thrombus formation, all blood-contacting surfaces are coated
with Carmeda biocompatible coating. Recently, ReliantHeart designed a remote monitoring
system for HeartAssist 5, called HeartAssistRemote, which enables healthcare providers to
monitor the pump remotely. So far, HeartAssist 5 is the only remotely monitored VAD in the
world. The original DeBakey pump showed a high incidence of thromboembolic events (22%)
and pump thrombosis (11–36%) and higher mortality rate in BTT (45%).[33] A recent study showed that using device thrombogenicity emulation (DTE) can
optimize the device features. Numerical simulations reported lower stress on platelets and
hence lower platelet activation.[34] The pediatric version of HeartAssist 5 received FDA approval for BTT in November
2012, and the Conquest Controller had been given another CE Mark approval earlier in
October 2012. In Europe, HeartAssist 5 achieved CE mark approval in 2014. An
investigational device exemption (IDE) was granted to HeartAssist 5 in 2014, and clinical
trials are in progress.[3]
HeartMate II
HeartMate II LVAD (Thoratec Corp., Pleasanton, CA, USA and Texas Heart Institute,
Houston, TX, USA) is a small AFP for full circulatory support to the left heart failurepatients. It is surgically implanted in either preperitoneal or intra-abdominal cavities
with its 20 mm inflow conduit attached to the ventricular apex and a 14 mm outflow graft
connected to the ascending aorta. It provides 3 to 10 lpm continuous flow at a pump speed
of 6000 to 15,000 rpm. Sufficient flow rate and hemodynamic pressure are typically
achieved at 9000 rpm.[30] The pump housing measures 60 mm long and 40 mm diameter, approximately the size of
a D-Cell battery, and weighs 375 g.[35] The rotor, which is the only moving part of the pump, contains a magnet and is
powered by an electromagnetic motor to rotate on blood-lubricated bearings.[36] Like the other AFPs, there is a risk of generating negative intraventricular
pressure and collapsing the ventricle. Therefore, the position of the inflow cannula and
ventricular preload are important. Figure
6 shows HeartMate II with sealed grafts, and Figure 3(a) presents a section cut of the device
showing the rotor, magnets, and bearings. All internal blood-contacting surfaces including
rotor, inlet stator, outlet stator, and thin-walled duct have a smooth polished titanium
surface. The inflow conduit and outflow graft have a textured microsphere surface, which
is similar to blood-contacting surfaces on the HeartMate XVE LVAD.
Figure 6.
HeartMate II pump with sealed grafts (Courtesy of Thoratec Corp.).
HeartMate II pump with sealed grafts (Courtesy of Thoratec Corp.).In addition to HeartMate II LVAD implanted inside the body, HeartMate II Left Ventricular
Assist System (LVAS) comprises the following components out of body: system monitor,
system controller, portable power module, universal battery charger, portable rechargeable
batteries, battery clips, and a percutaneous driveline.[37] The polyester velour covered driveline exits the skin and connects the motor to the
system controller, which continuously powers and controls the whole system. Figure 7 shows an X-ray of a patient
with HeartMate II LVAS. The system may be driven by the portable batteries or the power
module independently. The controller runs the pump at fixed-speed mode, and the speed of
rotation is determined at the time of implantation based on hemodynamic needs. The
rechargeable batteries are 14 V Li-Ion or 12 V NiMH, and each pair can provide enough
energy to let the patient live tether free for up to 10 h. The 20-foot power cable in the
power module makes it possible to run the system using external AC power source.
Figure 7.
X-Ray of a patient with HeartMate II LVAS (Courtesy of Thoratec Corp.).
X-Ray of a patient with HeartMate II LVAS (Courtesy of Thoratec Corp.).The portable power module runs the pump and powers the controller. It has integrated
battery for emergency support for up to 30 min. The system monitor communicates with the
system controller and allows the user to monitor system parameters or change the operation
speed. It also provides an estimation of output blood flow based on the pump speed and
amount of power being given to the pump. The relationship between power and flow at any
speed is mostly linear, with the exception of low and high-end regions. Currently,
Thoratec (Pleasanton, CA, USA) and WiTricity (Watertown, MA, USA) engineers are developing
a fully implantable version of the device using an optimized transcutaneous energy
transmission system (TETS).[35]HeartMate II has been approved in USA and Europe for BTT (2008) and DT (2010). In 2009, a
seminal trial showed better outcomes for HeartMate II LVAD compared with HeartMate XVE in
a cohort of patients as ill as those in the original REMATCH trial. HeartMate II LVAD is
the most widely used LVAD in the world with more than 12,969 implantations between 2004
and 2012.[3] In 2015, Thoratec, Inc. announced the 20,000th implantation.
INCOR
INCOR (Berlin Heart GmbH, Berlin, Germany) is a long-term implantable axial
continuous-flow ventricular assistive device with direct drive and magnetically levitated
rotor. Complete suspension of the rotor with two electromagnetic bearings at each end
eliminates the need for mechanical bearings, lubrication, sealing, and purging fluid.
INCOR measures 30 mm outer diameter and 114 mm long, somehow the longest among axial flow devices.[30] It weighs 200 g mostly contributed to its magnetic system components. The pump is
capable of providing a maximum flow rate of 5 lpm at rotation speed between 5000 and
10,000 rpm. In such range, the pump consumes 2–4 W power against 100 mmHg pressure.
Silicon inflow cannula delivers blood from left ventricular apex, and another silicon
outflow cannula is anastomosed onto the ascending aorta. All blood-contacting surfaces are
heparin coated by Carmeda process.[38] A percutaneous cable connects the pump to control unit and battery packs out of the
body. Figure 8 shows the latest
version of INCOR.
Figure 8.
INCOR pump (Courtesy of Berlin Heart GmbH).
INCOR pump (Courtesy of Berlin Heart GmbH).All components of the INCOR system are interconnected with snap-in connectors for easier
assembly. Figure 9 was taken
during the insertion of INCOR inlet cannula into the apex. The pump housing has a
stationary flow straightener at inflow port and a stationary diffuser at outflow port,
each equipped with magnets and magnetic coils. The flow straightener and diffuser reduce
the spin of blood flow and add to the pressure gradient across the pump. The impeller has
arrangements of magnets inside it and is enclosed by the motor stator. Radial and tilting
impeller movements are passively controlled. A unique controller algorithm positions the
rotor axially within the titanium housing to create a constant laminar flow. This
controller commands the currents required for impeller rotation and stabilization. The
faster the pump rotates, the higher pressure gradient it generates across the pump,
leading to a greater tendency of the impeller displacing from its central position. The
required current to counterbalance this flow force is the principle variable of the pump
flow controller.[39]
Figure 9.
Insertion of inlet cannula of INCOR into the left ventricular apex (Courtesy of
Berlin Heart GmbH).
Insertion of inlet cannula of INCOR into the left ventricular apex (Courtesy of
Berlin Heart GmbH).The first human clinical trial of INCOR started in 2002. Out of 212 cases worldwide, 65
(31%) were bridged to transplantation, 11 (5%) were bridged to recovery, and 43 (20%) were
still on the device at the time of publishing the results.[40] Blood chemistry found no trauma or hemolysis, and there were no device-related
mortalities. INCOR has been approved to be used for BTT and DT in Europe since 2013.
Berlin Heart started clinical trials for INCOR in 2009 under an FDA IDE approval.
Impella
Impella (Abiomed, Inc., Danvers, MA, USA) is one of the world's smallest pumps and
technically an expansion of Hemopump (Medtronic, Inc., Minneapolis, MN) that works based
on the principle of Archimedes screw.[41] This miniature device, often called transvalvular pump, is small enough to be
implanted into the ventricle through the femoral artery and powered via a flexible cable
from an external (percutaneous) source of energy. Impella is developed to serve the
patients who develop cardiogenic shock, and standard medical therapy has not been able to
control or recover the complication. It is designed to restore stable hemodynamics, reduce
infarct size, and protect the myocardium from ischemic damage. An impeller is levitated
magnetically without any mechanical bearing. It is passively centered in the pump casing
via permanent magnets in combination with hydrodynamically acting driving forces. The
lateral centering of the impeller is also affected by the permanent magnets cooperating
with the external driving magnetic means. Rotation of the impeller in the outlet area
pulls blood from the inlet cannula into the root of the ascending aorta. The largest
diameter section of the device is motor housing that resides in the aortic root,
integrated with a cannula portion.[42] The small diameter cannula passes through the aortic valve with the valve leaflets
around it. It should be small enough to minimize the risk of aortic valve regurgitation.
Impella has four versions: Impella 2.5, Impella 5.0, Impella LD (left direct), and Impella
CP (cardiac power). A brief comparison of the three versions is given in Table 3.
Table 3.
Differences between Impella 2.5, Impella 5.0, Impella LD (left direct), and Impella
CP (cardiac power).
Version
Impella 2.5
Impella 5
Impella LD
Impella CP
Catheter
9 Fr
9 Fr
9 Fr
9 Fr
Motor size
12 Fr
21 Fr
21 Fr
14 Fr
Flow rate
2.5 lpm
5 lpm
5 lpm
4 lpm
Support
Five days
Seven days
Seven days
Five days
Max speed
51,000 rpm
33,000 rpm
33,000 rpm
51,000 rpm
Guide wire
0.046 × 260 cm
0.064 × 260 cm
Not Required
0.046 × 260 cm
Pressure sensor
Direct
Differential
Differential
Direct
Differences between Impella 2.5, Impella 5.0, Impella LD (left direct), and Impella
CP (cardiac power).Impella 2.5 is implanted percutaneously via a 13 Fr sheath in the femoral artery, then
ascending aorta, across the valve (aortic or tricuspid) and then into the ventricle.[43] Impella CP delivers more flow on the same platform as Impella 2.5. Impella 5 is
inserted peripherally into the femoral or axillary artery into the left ventricle. Impella
LD, however, is inserted surgically directly into the ascending aorta. Figure 10 shows the Impella 2.5 and
an illustration of the implanted device. All devices are controlled with Impella Console
(Abiomed, Inc., Danvers, MA, USA). On each version, a pressure sensor is used to monitor
catheter position and calculate the flow.[44] Purge tubing attached to the pump supplies the pump bearings with a
blood-compatible purge fluid, typically 20% dextrose in water. The increased viscosity of
this cleansing fluid helps to create a pressure barrier against the blood that the device
is exposed to, thereby minimizing thrombus formation.[45] A pressure transducer (CM-Set) measures the purge pressure in the purge lumen of
the catheter. Currently, B. Braun (Melsungen, Germany) purge pump is a separate unit in
the Impella Console System.
Figure 10.
Impella LVAD (Courtesy of Abiomed, Inc.). (a) Impella 2.5 and (b) Illustration of
the implanted device (pediatric).
Impella LVAD (Courtesy of Abiomed, Inc.). (a) Impella 2.5 and (b) Illustration of
the implanted device (pediatric).FDA granted 510(k) clearance to Impella 2.5 (2008), Impella 5 (2009), Impella LD (2009),
and Impella CP (2012) for partial circulatory support up to 6 h in cardiac procedures not
requiring cardiopulmonary bypass. In 2016, Abiomed, Inc. announced that it has received
FDA Pre-Market Approval (PMA) for the Imeplla product series. Altogether, Impella family
has been utilized to support more than 10,000 patients in USA. In Europe, Impella family
has received CE mark and is widely used.[42]
Centrifugal flow devices
HeartWare HVAD
HeartWare HVAD (HeartWare International, Inc., Framingham, MA, USA) is a miniaturized
centrifugal flow pump for full circulatory support. Figure 11 approximates the size of HeartWare HVAD.
Although mainly designed for left heart, it has also been modified for biventricular
support.[46,47] The impeller is the only
moving part of the pump and is suspended using a hybrid suspension mechanism. The hybrid
suspension incorporates passive magnets and hydrodynamic thrust bearings. A gentle
inclination is created on the upper surface of the impeller blades. As the impeller
rotates, the generated blood flow across these inclinations creates a permanent gap
between the impeller and housing such that there is no mechanical contact between the
impeller and pump housing. This hybrid suspension provides an effective wearless
operation, which significantly downsizes the pump. Figure 3(c) shows an internal view of HeartWare HVAD.
Figure 11.
Approximate size of the HeartWare HVAD (Courtesy of HeartWare International,
Inc.).
Approximate size of the HeartWare HVAD (Courtesy of HeartWare International,
Inc.).HeartWare HVAD weighs 145 g and measures 50 mm diameter and 50 cc displacement volume.
The impeller within the titanium housing rotates in the range of 2400 to 3200 rpm and is
capable of delivering up to 10 lpm flow rate. Unlike similar devices, it is designed to be
implanted above the diaphragm adjacent to the heart, thereby eliminating the need for
abdominal surgery. The inflow of the pump is taken via an integrated inflow cannula, which
is directly inserted into the ventricular apex. The pump then propels blood through an
outflow graft, equipped with strain relief, into patient's ascending aorta. Figure 12 illustrates the
implantation of HeartWare HVAD. Two independent motor stators with separate circuitry are
available in the pumping unit to easily switch between single and dual stator modes and
therefore increase the device reliability.
Figure 12.
Illustration of the HeartWare HVAD system (Courtesy of HeartWare International,
Inc.).
Illustration of the HeartWare HVAD system (Courtesy of HeartWare International,
Inc.).A single driveline exits the patient's body and connects the pump to an external
controller and battery packs. Either two rechargeable batteries or a battery plus an
adapter (AC or DC) may run the controller and pump. In the latter case, the pump may be
plugged into an AC or DC outlet. One of the batteries runs the system for 6 h. When
depleted (or adapter disconnected), the controller switches to the backup battery and the
depleted one should be recharged. In 2012, HeartWare, Inc., announced its partnership with
DUALIS MedTech GmbH (Seefeld, Germany) to develop a fully implantable version of HVAD
using transcutaneous energy transmission systems. The manufacturer is also developing a
miniaturized AFP, with an approximate displacement volume of 15 cc, called HeartWare MVAD.[48] MVAD impeller suspension is based on the same technology used in HVAD and rotates
in the range of 16,000 to 28,000 rpm within a titanium cylindrical housing, capable of
providing up to 10 lpm flow rate. Ideally, MVAD will be implanted without the need for
median sternotomy. The outflow of MVAD is a 10-mm double woven graft anastomosed to the
aorta.In 2010, the results of the company's ADVANCE trial led to approval of HeartWare HVAD as
a BTT.[49] Results showed survival to transplant or ongoing support of 92% after 180 days.
HVAD is more compact than HeartMate II, which enables its implantation in smaller
patients. This pump is designed for intra-pericardial placement that decreases the risks
of infection or other complications that may be associated with intra-abdominal or
pre-peritoneal approaches.[7] HeartWare, Inc. is currently conducting a clinical trial on HeartWare HVAD for
DT.
HeartMate III
HeartMate III (Thoratec, Pleasanton, CA) is a continuous flow centrifugal pump with a
magnetically levitated rotor. The pump is designed to be able to mimic the biological
pulsatile flow through rapid changes of the pump speed. Although not unique to HeartMate
III, this is a highly important feature considering that the long-term effects of
continuous flow devices on organ perfusion are not well understood yet. HeartMate III
works based on a self-bearing or bearingless electromagnetic design to minimize the
life-limiting friction and wear. A recent design of the pump measured 69 mm outer diameter
by 30 mm height and 200 g weight, which includes the motor, inflow cannula, flexible
recovery section, outflow graft, bend relief, and all connecting hardware (Figure 13). Rotating in the speed
range of 2000 to 5500 rpm with the prime volume of 50 cc, the pump is capable of providing
up to 13 lpm flow rate. The major parts of the pump are[50]:
Figure 13.
HeartMate III (Courtesy of Thoratec Corp.).
An upper housing that includes the inflow/outflow cannula and the top half of the
volumeA lower housing that includes the bottom half of the volume and a cavity for the
motorA motor that includes drive and levitation functions in a single magnetic
structureA rotor that includes an impeller and passive permanent magnetsHeartMate III (Courtesy of Thoratec Corp.).The spinning rotor draws blood into the pumping chamber along the rotor's axis and
propels it tangentially through an outflow graft aligned perpendicular to the inflow cannula.[51] Motor drive and magnetic levitation coils share the same stator. Thus, no separate
motor or bearing exists. The impeller rotation is caused by a moving magnetic field
generated by the drive coils.[52] The rotation and radial levitation of the rotor are controlled by a two-axis active
control system and a feedback control loop. The remaining degrees of freedom (axial and
tilting) are controlled passively by the magnetic support. Inflow cannula of the device is
inserted into the left ventricular apex, and the outflow graft is anastomosed to the
ascending aorta. All blood-contacting surfaces of the pump are made of titanium with the
exception of polytetrafluoroethylene (PTFE) washers at joints and woven polyester grafts.
Like other members of the HeartMate family, HeartMate III incorporates sintered
titanium-textured surfaces to reduce anticoagulation requirements and thromboembolism.
Both percutaneous and implantable versions of the device are under development. For the
percutaneous version, a single driveline exits skin and connects it to a belt-mounted
driver/controller. Using a pair of rechargeable batteries, the wearable power transmission
can run the system for up to 6 h.[53] The design of the HeartMate III has been modified several times during the
developmental process. In another modified version of the pump, the inflow cannula is
integrated into the pump housing and is inserted directly into the left ventricular apex.[52] Using full active control over the rotor ramp speed, it can also provide an
optional near-physiologic artificial pulse. The Momentum 3 US IDE clinical trial for
HeartMate III is ongoing with more than 1000 participants to compare the performance of
HeartMate III to that of HeartMate II. In 2015, Thoratec announced that HeartMate III has
met the primary endpoint of the CE Mark trial. This trial was a single arm, prospective,
multicenter, non-blinded and non-randomized study. The study involved up to 50 patients at
nine sites in Europe, Australia, and Canada.[54] DT and BTT trials for Heartmate III are currently ongoing.
CentriMag and PediMag
Thoratec CentriMag and PediMag (Thoratec Corp., Pleasanton, CA, USA) are a family of
extracorporeal quickly installable devices for short-term left/right or biventricular use
in patients who need cardiopulmonary support with potentially recoverable heart failure.
CentriMag and PediMag pumps use magnetic levitation to suspend the rotor by eight L-shaped
iron cores.[41] Their extracorporeal system comprises the centrifugal pump, an electromagnetic
motor, a drive console, an ultrasonic flow probe, and a tubing circuit.[55] To reduce hemolysis, mechanical gaps in the pump are more than 0.6 mm and the pump
has no valves, seals, or moving parts other than the rotor.[56] CentriMag is appropriate for patients larger than 20 kg and is capable of providing
up to 9.9 lpm flow rate under normal physiologic pressure at rotor speed up to 5000 rpm
with a priming volume of 31 mL. PediMag is optimized for pediatric patients around 20 kg
and is capable of 1.5 lpm flow rate with a priming volume of 14 mL. Figure 14 shows the Thoratec PediMag (a) and
CentriMag (b). Figure 15 shows
the Thoratec CentriMag installed on the electromagnetic motor.
Figure 14.
Thoratec PediMag and CentriMag BiVADs (Courtesy of Thoratec Corp.). (a) PediMag and
(b) CentriMag.
Figure 15.
Thoratec CentriMag installed on the electromagnetic motor (Courtesy of Thoratec
Corp.).
Thoratec PediMag and CentriMag BiVADs (Courtesy of Thoratec Corp.). (a) PediMag and
(b) CentriMag.Thoratec CentriMag installed on the electromagnetic motor (Courtesy of Thoratec
Corp.).Both CentriMag and PediMag operate on the same hardware platform, eliminating the need to
have separate system setups to support adult and pediatric patients. They can be used as
right/left or biventricular assist devices with cannulation of the left ventricular apex,
left atrium, or right atrium for inflow, and aorta or pulmonary artery for outflow.[55] In both pumps, the inlet is on the same axis as the rotor, while the outlet is
directed perpendicular to the inlet, tangent to the pump circle. The inflow is a 32 Fr
wire-reinforced cannula, whereas the outflow cannula is a 22 Fr straight one.[57,58] These pumps are currently used for
stabilizing hemodynamic needs with minimum mechanical failure and blood-related problems
such as hemolysis or pump-induced trauma. Figure 16 compares the implantation of CentriMag
with HeartMate II.
Figure 16.
Implantation of Thoratec CentriMag versus HeartMate II (Courtesy of Thoratec
Corp.)
Implantation of Thoratec CentriMag versus HeartMate II (Courtesy of Thoratec
Corp.)FDA granted 510(k) clearance to CentriMag for use up to 6 h in patients requiring
extracorporeal circulatory support during cardiac surgery. Investigational trials are
ongoing in an US pivotal trial to demonstrate safety and effectiveness for periods of
support of up to 30 days. Additionally, CentriMag is approved under an FDA Humanitarian
Device Exemption to be used as an RVAD for periods of support up to 30 days in patients
with cardiogenic shock due to acute right ventricular failure. In Europe, CentriMag has CE
Mark approval for up to 30 days of support.Both devices are cleared for clinical use for up to 6 h. CentriMag is also being
evaluated for BRT or BTT therapy. It is approved for RVAD support for up to 30 days for
patients with cardiogenic shock due to right ventricular failure. In conjunction with
CentriMag console and motor, PediMag blood pump is 510(k) cleared by FDA for support
periods of up to 6 h. Thoratec announced that they have submitted an IDE to the FDA to
begin a U.S. clinical trial examining the safety and probable benefit of the device for
use up to 30 days to support pediatric patients. Outside USA, the device is branded as
PediVAS and has CE Mark approval for support durations of up to 30 days.
BPX-80 Bio-Pump Plus and BP-50 Bio-Pump
BPX-80 Bio-Pump (Medtronic, Minneapolis, MN, USA) is a continuous extracorporeal CFP made
of molded polycarbonate for short-term supports. It has no impeller vane or roller, which
is not commonly found in other centrifugal pumps.[59] The constrained-forced vortex pumping principle is based on a series of
smooth-surfaced rotating cones that pull the blood into the vortex created by the rotation
of cones. As the blood flows toward the outlet, the vortex energy, generated by these
cones, is transferred to the blood in the form of pressure and velocity. The smooth vortex
cone design of BPX-80 Bio-Pump has shown greater air retention compared with
impeller-based centrifugal pumps. Higher air retention means less air passes through the
pump outlet and finally towards the patient. This pumping principle promotes laminar flow
and improves the blood-handling capabilities. The pump utilizes a polycarbonate journal
bearing to support the impeller, which may lead to lower durability and higher mechanical
wear. The pump also comprises a double-lip seal design over this precision bearing. The
size of the pump is as follows: cone-shaped impeller is 79 mm diameter, top housing is
50 mm high, and the bottom housing is 16 mm thick. BPX-80 Bio-Pump has a priming volume of
86 mL, weight of 200 g, and consumes 8 W power.[13] It is capable of providing up to 10 lpm flow rate at rotation speeds of up to
4500 rpm.[59,60]
Figure 17 shows the BPX-80
Bio-Pump Plus and its cutaway section.
Figure 17.
BPX-80 LVAD (Courtesy of Medtronic). (a) BPX-80 Bio-Pump Plus and (b) cross
section.
BPX-80LVAD (Courtesy of Medtronic). (a) BPX-80 Bio-Pump Plus and (b) cross
section.To decrease flow recirculation at the outlet opening and reduce the free plasma
hemoglobin, the ellipse-shaped intersection of the outlet connector and pump housing is
straightened to a nearly vertical line, based on computational fluid dynamics (CFD)
analysis of shear stress near the opening. In addition to that, the pump outlet is tapered
in order to create a more gentle blood flow. The pump is available with two biocompatible
coatings: Carmeda BioActive Surface (under license from Carmeda AB, Sweden.) and Trillium
Biosurface (under an agreement with BioInteractions, Limited, United Kingdom). The BPX-80
Bio-Pump Plus runs in conjunction with the remote-tethered Bio-Console 560 drive system
that runs the pump using integrated brushless DC motor, powers the pump by two 12 V
lead-acid gel rechargeable batteries, and controls the pump flow rate. It also detects
generated bubbles and measures the blood pressure. The Bio-Console560 weighs 17.19 kg and
measures 31.88 × 22.38 × 43.02 cm, providing a limited level of mobility. Figure 18 shows the Bio-Pump BP-80
and Bio-Console560.
Figure 18.
Bio-Pump BP-80 and Bio-Console560 (Courtesy of Medtronic).
Bio-Pump BP-80 and Bio-Console560 (Courtesy of Medtronic).BP-50 Bio-Pump (Medtronic, Minneapolis, MN, USA) is similar to BPX-80 Bio-Pump Plus,
designed for pediatric use and patients with special needs. With a 48 mL priming volume,
it is capable of providing up to 1.5 lpm flow rate with a maximum speed of 4500 rpm. It
may be controlled by the same controlling units that run BP-80 Bio-Pump Plus such as
Bio-Console 560. Similar to BP-80, BP-50 is available with Carmeda Bioactive Surface.
Figure 19 shows the BP-50
Bio-Pump and its particular console called Bio-Console 540.
Figure 19.
BP-50 LVAD (Courtesy of Medtronic). (a) BP-50 Bio-Pump and (b) Bio-Console 540.
BP-50LVAD (Courtesy of Medtronic). (a) BP-50 Bio-Pump and (b) Bio-Console 540.BPX-80 Bio-Pump Plus is a 510(k) cleared device since 1985. It is approved to be used in
conjunction with Bio-Console up to 6 h in extracorporeal cardiopulmonary bypass. BP-50 has
also got a 510(k) clearance from FDA. BPX-80, BP-50, and BP-Console are all CE marked in
the Europe.
EvaHeart
EvaHeart (Evaheart, Inc., USA, Pittsburgh, PA, USA developed by Sun Medical Technology
Corp., Nagano, Japan) is an implantable CFP designed for long-term left ventricular
support. It incorporates one water-lubricated hydrodynamic journal bearing with pure,
sterile water injected into the pump housing from an external water reservoir in the
controller package via two 2.5 mm pipes embedded in the percutaneous driveline. The water
and blood chambers are separated from each other by a thin seal called Cool-Seal. The
water lubricates the journal bearing, cools the motor coil and journal bearing down, and
flushes the inner seal faces. The hydrodynamic journal bearing supports a shaft that
connects the motor to the impeller. The pump measures 55 mm by 64 mm with a priming volume
of 132 mL and weighs 420 g. It is capable of providing up to 12 lpm flow rate at 2600 rpm
against 100 mmHg pressure. At flow rates within 5 to 9 lpm, the power consumption of the
pump is reported 9 to 10 W.[61] To provide complete bearing wash out during operation, vanes have an open-faced and
swept back shape.[2] Sixteen millimeters inflow cannula and outflow graft are connected to the
ventricular apex and ascending aorta, respectively. They are made of ePTFE or polyester,
and all blood-contacting surfaces are coated with either diamond-like carbon or a
proprietary polymer called MPC.[62] Although the pump has a completely flat pressure–flow curve, the pump output may
become pulsatile. During systole, when the pressure difference between left ventricle and
aorta drops, an instant high peak flow leads to a higher peak pressure in the aorta, which
is identical to systolic pressure. During diastole, as the left ventricle-aorta pressure
difference increases, pump flow rate decreases and thus creates an instant lowest pressure.[63] The initial clinical trial in Japan completed in 2008 after successful implantation
in 18 patients. The mean support duration was 2.76 years with a maximum of 6.29 years.
Eight patients were successfully bridged to transplantation with mean support time of
three years. The overall survival at six months was 89%.[64] The device received the final regulatory approval from the Japanese Pharmaceuticals
and Medical Devices Agency. As of December 2013, 118 patients have received the device (80
ongoing). EvaHeart is still limited to investigational purposes by Federal law in USA.
Currently, Evaheart, Inc. is conducting BTT trials under an FDA-approved IDE.[3]
Mixed flow devices
Synergy micro-pump
Synergy Micro-Pump (CircuLite, Inc., Saddle Brook, NJ, USA) is a miniaturized partial
circulatory support system developed for long-term use in NYHA class IIIb/early IV
patients. It is intended to be implanted in less ill patients who have symptomatic heart
failure, but BTR is preferred over BTT or DT for them. Synergy will augment the native
ventricle that provides some flow by itself, but not sufficiently. This allows the heart
to rest and recover, until the appropriate time for device explantation.[2] Synergy is often characterized by its hybrid or mixed flow that presents both
features of AFPs and CFPs. It combines axial, centrifugal, and orthogonal flow paths with
a single stage impeller that is driven by an integrated brushless DC motor. This titanium
micro-pump measures 12 mm diameter by 49 mm length, the approximate size of an AA battery.
The pump weighs 25 g and has a priming volume of 1.5 mL. Figure 20 approximates the size of Synergy. This
small size allows the device to be implanted surgically in the subclavicular space, into a
pacemaker-like pocket, through a right upper chest thoracotomy. This eliminates the need
for cardiopulmonary bypass or sternotomy and makes the implantation procedure less
invasive. The inflow cannula is made of silicon and reinforced by NiTi shape memory alloy
and is inserted into the left atrium. The outflow graft is made of PTFE with an inner
diameter of 8 mm and thickness of 1 mm, anastomosed to the subclavian artery. A small
percutaneous lead connects the pump to the external wearable controller that weighs around
1 kg. The rechargeable dual battery pack can run the system for 6 to 8 h. The pump's rotor
is supported by a mechanical pivot bearing and stabilized by a combination of magnetic and
hydrodynamic forces. By operating at the range of 20,000 to 28,000 rpm, the pump is
capable of providing 2 to 4.25 lpm flow rate. Although the small size and achieved an
operational speed of the device are promising, this lower flow is a concern in increasing
the risk of thrombus formation.[65] Synergy received CE Mark in 2012 and an IDE approval in 2013 for investigational
trials. The device is being supported by an NIH grant for further developments, system
validation, and clinical evaluation. The manufacturer is planning to expand the Synergy
technology to cover a broader patient population. Thus, the next-generation of the Synergy
pumps is currently under development. These modified versions are designed for
endovascular implantation, right heart support, all support, child support, and infant
support. Figure 21 compares the
implantation of these pumps. The endovascular system will be implanted with the inflow
cannula placed trans-septal into the left atrium from the subclavian vein and the outflow
graft anastomosed to the subclavian artery. Preclinical animal studies of this device are
now completed, and it is approaching clinical evaluation in Europe. A modified Synergy
pump is also being designed for right heart support or working in conjunction with the
current design as a biventricular partial circulatory support system.[66] Animal studies and cannula development of this device are in progress. All-support
platform is intended to cover 500,000 patients market with an optimized design of Synergy
for up to 6.0 lpm flow rate. It aims to accommodate the complete spectrum of a patient's
circulatory support needs from partial to full. This version has passed the
proof-of-concept step and is awaiting animal studies to support the design phase and
clinical evaluation. The child support system is also based on modified Synergy for 1.5 to
3 lpm output with a different cannulation. Input cannula will be inserted into the left
ventricular apex, while the outflow graft returns blood to the ascending aorta. The infant
support system, also called Synergy Nano-Pump, is a super small pump for 0.3 to 1.5 lpm
flow and is currently under development for completing proof-of-concept.
Figure 20.
Approximate size of Synergy Micro-Pump (Courtesy of CircuLite, Inc.)
Figure 21.
Modified versions of CircuLite MicroPump (Courtesy of CircuLite, Inc.). (a)
endovascular system, (b) right heart support, (c) all-support and (d) child
support.
Approximate size of Synergy Micro-Pump (Courtesy of CircuLite, Inc.)Modified versions of CircuLite MicroPump (Courtesy of CircuLite, Inc.). (a)
endovascular system, (b) right heart support, (c) all-support and (d) child
support.
Current limitations
Although the MCS field has improved significantly over the past three decades, many
challenges are still hindering development of these devices. Stroke remains prevalent.[67] Questions such as optimum time-point for MCS, careful patient selection criteria, and
how to maximize the biocompatibility of current systems remain unanswered or under debate.
Currently, the decision regarding when to implant an MCS is based on published scientific
evidence and INTERMACS values. Ultimately, a decision should be made between MCS
implantation at a later time point with the risk of rapidly deteriorating heart failure in
the meantime, or earlier MCS with the risk of complications associated with MCS therapy.[68]The following section provides a summary of the current limitations and challenges,
inviting further improvements and developments.
Device-related issues
Device-related problems such as coagulation disorder, gastrointestinal bleeding, pump
thrombosis, device-related infection, and cerebrovascular disorders remain the most
critical challenges for MCS systems.[68] Hemorrhage at grafts, thromboembolism at stagnant areas, and hemolysis due to high
shearing stress applied to the red blood cells are the main issues with these devices.
Heat generation is another matter; if temperature increases high enough, proteins in the
blood will denature resulting in irreversible blood damage. Since no valves are employed
in these devices, flow back into the heart could worsen the heart failure. The bearing
systems for blood pumps should be hemocompatible, reliable, and durable. Mechanical
bearings need high flow wash out at the junction between the rotating and stationary
parts, which adds to the complexity and difficulty of maintenance. For immersed bearings,
the viscous friction in the gap between the rotating impellers and the stationary housing
is very high which increases the risk of blood damage. The disadvantage of fully
magnetically levitated bearings is that they are highly dependent on the electromagnetic
feedback control. If this connection is lost, the rotor crashes in the housing.
Pulsatility
Patients supported by continuous flow pumps often undergo physiological alterations that
are not well understood.[7] Adverse clinical outcomes such as aortic insufficiency, thrombosis, exacerbation of
right heart failure, and bleeding remain a significant problem after CF-VAD implantation.[69] Although the long-term effects on survival are unclear,[70] it has been shown that the aortic insufficiency associated with CF-VADs can result
in insufficient flow, increased thrombogenicity, and reduced chance of myocardial
recovery.[71,72] Pulsation may allow more
physiologic unloading of the left ventricle by allowing unloading only during a portion of
the cardiac cycle, as opposed to the full cardiac cycle in current CF-VADs, thus reducing
pulmonary hypertension and right heart failure.[73] Although some CF-VAD recipients retain some degree of pulsatility because of the
improvements in the unloaded heart, others can show no significant pulse, the consequences
of which are not clear yet.
Accurate flow control
Accurate flow control on the current CF-VADs for achieving pulsatility remains a
challenge. Speed modulation may offer a potential solution to create pulsatility and
prevent these undesired consequences.[45] However, the performance of the flow control algorithm is highly dependent of the
accuracy and resolution of the flow measurement. Since direct flow measurement is often
impossible, an estimation algorithm is used to predict the pump output based on the pump
speed and input power. Accurate, miniaturized, and biocompatible flow feedback loops are
lacking. Heartmate III, currently under investigation in USA, has shown promising results
for improved flow accuracy and generating physiologic pulsatility.[3]Using two continuous flow pumps for biventricular support is still facing major
challenges. There is no algorithm yet for adjusting the left and right flows by
automatically and interdependently adjusting the pump speeds. Therefore, pumps still need
two separate controllers. The necessary manual adjustments complicate the care of patients
on BiVAD support.[7] The latest versions of CFPs are devices with magnetically levitated and
hydrodynamic non-contact impeller suspension systems.[74-76]
Thrombogenicity
Despite recent advances in MCS development, most notably the CF-VADs, their long-term use
remains limited by the thrombotic complications. High shear stress and recirculating flow
patterns can lead to platelet activation. Blood contact with artificial surfaces can also
stimulate blood clotting. A DT trial on HeartMate II in 2009 showed that 11% of patients
suffered a stroke after implantation due to increased shear stress on blood cells.[77] A considerable rate of thromboembolic complications was also reported for the
AbioCor TAH.[78] The non-physiologic flow patterns in CF-VADs are considered as one of the major
contributors to hemostatic response. Flow patterns along with the artificial surface
determine where thrombi will form, its size and composition, and whether it will remain
local or it will embolize.[79] Valve designs and hemodynamics also affect the platelet activation.[80] CF-VADs are much smaller than PF-VADs and have smaller gaps between components of
the pump. In older generation PF-VADs, any thrombus created in the pump could be
dislocated, possibly leading to a stroke. The smaller size of CF-VADs makes them
susceptible to thrombosis of the entire pump, where the clot stays in the device and leads
to increased hemolysis and device malfunction.[81] Despite aggressive anticoagulant therapy, MCSs can change the coagulation through
activating platelets by generating non-physiologic pulsatility, eddy formation,
turbulence, or exposure times along platelet trajectories. Development of optimization
methods such as DTE has shed more light on identifying thrombogenic regions. This
mathematical method combines CFD with particle tracking and in-silico device modeling to
determine stress of individual platelets.[79] Platelet activity state (PAS) also facilitates measurement of thrombogenicity of devices.[82] Development and experimental validation of these methods are still an ongoing
challenge. Management protocols for MCSs are usually institution dependent, with various
factors related to the clinician's experience. Anticoagulation therapy has to be
individually tailored. Otherwise, patients may not receive the correct dose of
anticoagulant. A significant variation exists between over-anticoagulation and
under-anticoagulation in a patient population where coagulation response to the MCS varies
considerably between individuals. Over-anticoagulation therapy has often been associated
with gastrointestinal bleeding and intracranial hemorrhage. Common complications due to
under-anticoagulation include pump thrombosis, hemolysis, and embolic strokes.
Hemolysis
CF-VADs have demonstrated improved survival and reduced complications compared with
PF-VADs. Nevertheless, a study in 2014 reported 18 out of 100 patients receiving HeartMate
II were diagnosed with hemolysis.[83] They concluded that hemolysis is associated with very high one-year mortality that
is more than two times greater than that observed for the non-hemolyzing patients.
Treatments involve MCS exchange or explantation, listing for transplantation, or
intensifying anticoagulation therapies. Besides thrombus formation inside the device, a
few other hemolysis risk factors are:Increased shear stress on red blood cells due to device malpositioning, kinking on
the outflow graft/cannula. These cases are often followed by an abrupt change in the
LVAD performance such as power spike, flow rate, or pulsatility indexDehydration of underfilled left ventricle with increased inlet velocityRegurgitant or stenotic valvesTransfusion-related to immune- and non-immune-mediated hemolysisDue to the inherent complexity of patients' responses to anticoagulant therapies,
clinicians may have to choose between thrombosis, bleeding, or sometimes both.[81] Therefore, achieving the optimal anticoagulation therapy remains an ongoing
challenge. Unfortunately, researchers have limited access to data regarding the prognosis
of patients with hemolysis. More single-center studies on the incident and clinical
outcomes associated with thrombosis, hemolysis, gastrointestinal bleeding, and infection
have been reported over the last 5 to 10 years.[81,83-87]
Gastrointestinal bleeding
Multiple reports have been published on the gastrointestinal (GI) bleeding in patients
who receive MCS.[88-92] In a narrative
review in 2013, 20.5% of LVAD recipients out of 1543 patients developed GI bleeding.[90] This complication has been more pronounced in CF-VADs compared with PF-VADs.
Bleeding can occur throughout the GI tract. Although the bleeding is usually manageable,
the definite bleeding site remains unidentified. Lack of enough knowledge about the
bleeding site hinders the MCS technology refinements and improvements. Despite some GI
bleedings are directly related to the device itself, LVAD replacement is seldom necessary.[93] More research is needed on the etiology of GI bleeding as well as the proper
anticoagulation management before and after GI bleeding. The role of anticoagulation
acquired von Willebrand syndrome and platelet dysfunction due to altered hemodynamics are
not clearly understood yet.
Infection
Infection occurs in up to 60% of MCS recipients.[94,95] MCS-related infections, bacteremia, and
fungemia are common in these patients. Major MCS-related infections are percutaneous
driveline, pump pocket, and cannula. The percutaneous driveline, particularly at the exit
site, is quite susceptible to infection. Visual inspection of driveline for diagnosing
infections may be hindered by purulent discharge and surrounding cellulitis. It is
reported that even with infection, systemic signs such as fever, elevated inflammatory
markers, or leukocytosis may not be present.[96] Treatment of infection may require removal of the device, since the entire device
can be susceptible to biofilm formation.[97] Cannula and pump infections are among the most serious MCS-related infections.
Infection along the cannula or inside the pump can lead to pump failure due to flow
blockage. Besides the differences in the MCS technologies themselves, heart failure risk
score, older age, diabetes, renal failure, and nutritional status are known as the
additional factors associated with the overall infection.CF-VADs have shown less MCS-related infections compared with PF-VADs due to smaller pump
size and driveline caliber. However, sepsis and non-MCS-related infections still cause the
majority of infections during the first 90 days of implantation. MCS-related infections,
especially driveline and pump pocket infections, comprise the majority of infections after
90 days of implantation.[98,99]
Although infection complications of MCS therapy have been described for many years,
current recommendations for monitoring infection after MCS implantation are based on
observational studies and expert opinions. Randomized controlled trials for studying the
MCS-related and non-MCS-related infections after implantation seem essential. An ongoing
effort towards reducing MCS-related infections is realization of TETS. TETS can recharge
the batteries wirelessly, so patients are neither tethered to a large air pumping console
nor pierced by drivelines and tubes.[100-102]
Authors: E A Rose; A C Gelijns; A J Moskowitz; D F Heitjan; L W Stevenson; W Dembitsky; J W Long; D D Ascheim; A R Tierney; R G Levitan; J T Watson; P Meier; N S Ronan; P A Shapiro; R M Lazar; L W Miller; L Gupta; O H Frazier; P Desvigne-Nickens; M C Oz; V L Poirier Journal: N Engl J Med Date: 2001-11-15 Impact factor: 91.245
Authors: K Bourque; D B Gernes; H M Loree; J S Richardson; V L Poirier; N Barletta; A Fleischli; G Foiera; T M Gempp; R Schoeb; K N Litwak; T Akimoto; M J Watach; P Litwak Journal: ASAIO J Date: 2001 Jul-Aug Impact factor: 2.872
Authors: E V Potapov; M Loebe; B A Nasseri; H Sinawski; A Koster; H Kuppe; G P Noon; M E DeBakey; R Hetzer Journal: Circulation Date: 2000-11-07 Impact factor: 29.690
Authors: Chihiro Miyagi; Barry D Kuban; Christine R Flick; Anthony R Polakowski; Takuma Miyamoto; Jamshid H Karimov; Randall C Starling; Kiyotaka Fukamachi Journal: Heart Fail Rev Date: 2021-05-01 Impact factor: 4.214
Authors: Landon H Tompkins; Steven R Prina; Barry N Gellman; Gino F Morello; Thomas Roussel; Jonathan A Kopechek; Stuart J Williams; Priscilla C Petit; Mark S Slaughter; Steven C Koenig; Kurt A Dasse Journal: Cardiovasc Eng Technol Date: 2021-09-13 Impact factor: 2.305
Authors: Patrick Borchers; Patrick Winnersbach; Sandra Kraemer; Christian Beckers; Eva Miriam Buhl; Steffen Leonhardt; Rolf Rossaint; Marian Walter; Thomas Breuer; Christian Bleilevens Journal: J Clin Med Date: 2022-06-15 Impact factor: 4.964
Authors: Landon H Tompkins; Barry N Gellman; Gino F Morello; Steven R Prina; Thomas J Roussel; Jonathan A Kopechek; Priscilla C Petit; Mark S Slaughter; Steven C Koenig; Kurt A Dasse Journal: ASAIO J Date: 2021-09-01 Impact factor: 3.826