The CentriMag: a new optimized centrifugal blood pump with levitating impeller.

Juerg Peter Mueller1, Andreas Kuenzli, Oliver Reuthebuch, Kurt Dasse, Stella Kent, Gregor Zuend, Marko Ivan Turina, Mario Louis Lachat.   

Abstract

PURPOSE: Blood pumps are routinely used for circulatory and pulmonary support. However, blood trauma and pump failure remain severe drawbacks of currently available pump models. This study evaluated the first clinical application of a new, totally bearingless centrifugal blood pump (CentriMag).
MATERIAL AND METHODS: A centrifugal pump consisting of an electromagnetic suspended impeller was used as a blood pump during beating-heart coronary artery bypass grafting in 11 patients (mean weight, 77.4 kg). Heparin in a bolus of 150 IU/kg body weight was administered, and activated clotting time was maintained at approximately 180 to 250 seconds during extracorporeal circulation. Pump-induced blood trauma was evaluated by measurement of plasma free hemoglobin (PFH), lactate dehydrogenase (LDH), hematocrit, total bilirubin, and platelet levels.
RESULTS: Mean pump flow was 3.3 +/- 0.62 L/min, and mean pressure gradient through the oxygenator was 69 +/- 4 mm Hg. No pump dysfunction occurred during a mean application time of 105 +/- 26 minutes. Inspection of the pump housings showed no internal thrombus formation despite low-dose heparinization. Only slight hemolysis was observed with a mean PFH level of 1.96 micromol/L; LDH, 460 U/L; hematocrit, 33%; total bilirubin, 25 micromol/L; and platelets, 191 x 10(3)/microL.
CONCLUSIONS: The bearingless CentriMag blood pump is a safe and reliable new device that produces only minimal hemolysis. It seems to be suited for long-term evaluation as a blood pump for extracorporeal membrane oxygenation or as ventricular assist device.

Entities:  

Mesh:

Year:  2004        PMID: 15802261     DOI: 10.1532/HSF98.20041068

Source DB:  PubMed          Journal:  Heart Surg Forum        ISSN: 1098-3511            Impact factor:   0.676


  12 in total

Review 1.  The use of computational fluid dynamics in the development of ventricular assist devices.

Authors:  Katharine H Fraser; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  Med Eng Phys       Date:  2010-11-13       Impact factor: 2.242

2.  Ground transportation of a pediatric patient on ECMO support.

Authors:  David Machin; Richard Scott; Aimee Hurst
Journal:  J Extra Corpor Technol       Date:  2007-06

3.  Bridge-to-decision therapy with a continuous-flow external ventricular assist device in refractory cardiogenic shock of various causes.

Authors:  Hiroo Takayama; Lori Soni; Bindu Kalesan; Lauren K Truby; Takeyoshi Ota; Sophia Cedola; Zain Khalpey; Nir Uriel; Paolo Colombo; Donna M Mancini; Ulrich P Jorde; Yoshifumi Naka
Journal:  Circ Heart Fail       Date:  2014-07-15       Impact factor: 8.790

Review 4.  A ventricular assist device as a bridge to recovery, decision making, or transplantation in patients with advanced cardiac failure.

Authors:  Siyamek Neragi-Miandoab
Journal:  Surg Today       Date:  2012-07-20       Impact factor: 2.549

5.  Plasma Free Hemoglobin Generation Using the EOS PMP Oxygenator and the CentriMag® Blood Pump.

Authors:  Ashley B Hodge; Matthew A Deitemyer; Victoria L Duffy; Dmitry Tumin; Dorothy A Garbin; Kathleen K Nicol; Don Hayes; Mary J Cismowski; Andrew R Yates
Journal:  J Extra Corpor Technol       Date:  2018-06

6.  The Influence of Impeller Geometries on Hemolysis in Bearingless Centrifugal Pumps.

Authors:  Pascal Puentener; Marcel Schuck; Johann W Kolar
Journal:  IEEE Open J Eng Med Biol       Date:  2020-11-16

7.  CentriMag left ventricular assist system: cannulation through a right minithoracotomy.

Authors:  Igor D Gregoric; William E Cohn; Mehmet H Akay; Saverio La Francesca; Tim Myers; O H Frazier
Journal:  Tex Heart Inst J       Date:  2008

8.  Adequate systemic perfusion maintained by a CentriMag during acute heart failure.

Authors:  Roberto R Favaloro; Alejandro Bertolotti; Mirta Diez; Liliana Favaloro; Carmen Gomez; Margarita Peradejordi; Julio Trentadue; Lorena Hellman; Yanina Arzani; Pilar Varela Otero
Journal:  Tex Heart Inst J       Date:  2008

9.  Temporary mechanical circulatory support: insights and evolving strategies.

Authors:  Anvay Mulay; Talha Meeran; Ashish Gaur; Sandeep Sinha; Pravin Kulkarni; Neeraj Kamat; Satish Javali; Amish Jasapara; Vijay Shetty
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2021-01-09

Review 10.  Current state-of-the-art of device therapy for advanced heart failure.

Authors:  Lawrence S Lee; Prem S Shekar
Journal:  Croat Med J       Date:  2014-12       Impact factor: 1.351

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.