| Literature DB >> 31125378 |
Yi Sun1, Zuojian Hu2, Zhili Huang3, Huaping Chen2, Shanzi Qin2, Zhong Jianing2, Siyuan Chen2, Xue Qin2, Yi Ye4, Chengbin Wang1.
Abstract
OBJECTIVE: The aim of this study is to evaluate the performance of different platelet counting methods (optical, impedance, fluorescence and hand counting) applied in different analysers by comparing with the international flow cytometric reference method (IRM).Entities:
Mesh:
Year: 2019 PMID: 31125378 PMCID: PMC6534315 DOI: 10.1371/journal.pone.0217298
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Descriptive analysis of different methods for platelet counting.
| Method | Minimum | Maximum | Mean | CV% |
|---|---|---|---|---|
| LH780 | 4.60 | 986.5 | 166.48 | 8.75% |
| BC-6000 Plus-O | 4.00 | 1033.5 | 178.04 | 7.28% |
| BC-6000 Plus-I | 2.00 | 987.5 | 172.92 | 3.26% |
| Sysmex XN-9000-O | 6.00 | 945.0 | 159.23 | 9.76% |
| Sysmex XN-9000-I | 4.00 | 965.0 | 175.03 | 6.33% |
| Sysmex XN-9000-F | 4.00 | 959.0 | 160.19 | 6.74% |
| Hand counting | 4.20 | 980.0 | 163.93 | 18.39% |
| IRM | 1.88 | 960.0 | 169.64 | 1.51% |
Spearman rank correlation and Bland-Altman analysis of platelets between evaluated methods and IRM.
| Levels of platelets | N | r | Average bias (mean difference) | 95% Limits of agreement | |
|---|---|---|---|---|---|
| LH780 vs IRM | 0.964 | 0.000 | +3.8 | -10.2 to +17.8 | |
| BC-6000 Plus-O vs IRM | 0.985 | 0.000 | +4.3 | -5.5 to +14.1 | |
| BC-6000 Plus-I vs IRM | 0.961 | 0.000 | +2.8 | -11.7 to +17.3 | |
| Sysmex XN-9000-O vs IRM | 0.984 | 0.000 | -1.5 | -10.8 to +7.8 | |
| Sysmex XN-9000-I vs IRM | 0.937 | 0.000 | +4.9 | -14.9 to +24.7 | |
| Sysmex XN-9000-F vs IRM | 0.988 | 0.000 | -1.5 | -9.4 to +6.4 | |
| Hand counting vs IRM | 0.922 | 0.000 | +3.1 | -25.8 to +32.1 | |
| LH780 vs IRM | 0.947 | 0.000 | -7.6 | -67.4 to +52.2 | |
| BC-6000 Plus-O vs IRM | 0.961 | 0.000 | +9.0 | -42.6 to +60.6 | |
| BC-6000 Plus-I vs IRM | 0.966 | 0.000 | +4.5 | -42.2 to +51.1 | |
| Sysmex XN-9000-O vs IRM | 0.962 | 0.000 | -16.6 | -69.8 to +36.6 | |
| Sysmex XN-9000-I vs IRM | 0.947 | 0.000 | +8.7 | -48.0 to +65.5 | |
| Sysmex XN-9000-F vs IRM | 0.959 | 0.000 | -15.5 | -67.0 to +36.0 | |
| Hand counting vs IRM | 0.841 | 0.000 | -9.7 | -101.3 to +81.9 | |
| LH780 vs IRM | 0.855 | 0.000 | -22.7 | -136.4 to +91.0 | |
| BC-6000 Plus-O vs IRM | 0.876 | 0.000 | +29.5 | -77.2 to +136.1 | |
| BC-6000 Plus-I vs IRM | 0.932 | 0.000 | +0.4 | -74.5 to +75.4 | |
| Sysmex XN-9000-O vs IRM | 0.864 | 0.000 | -33.4 | -139.6 to +72.9 | |
| Sysmex XN-9000-I vs IRM | 0.960 | 0.000 | -7.4 | -66.6 to +51.7 | |
| Sysmex XN-9000-F vs IRM | 0.831 | 0.000 | -26.9 | -153.4 to +99.5 | |
| Hand counting vs IRM | 0.851 | 0.000 | -38.0 | -173.0 to +97.0 |
Taking IRM as the gold standard, the detection characteristics of the evaluated methods.
| Cutoff | Sensitivity | Specificity | Concordance Rate | Kappa value (95% CI of K) |
|---|---|---|---|---|
| LH780 | 97.6% (0.94–0.99) | 98.8% (0.96–1.00) | 98.2% (0.96–0.99) | 0.964(0.935–0.993) |
| BC-6000 Plus-O | 94.0% (0.89–0.97) | 100% (0.98–1.00) | 97.0% (0.94–0.98) | 0.940(0.903–0.977) |
| BC-6000 Plus-I | 96.4% (0.92–0.99) | 98.8% (0.96–1.00) | 97.6% (0.95–0.99) | 0.952(0.919–0.985) |
| Sysmex XN-9000-O | 99.4% (0.97–1.00) | 98.2% (0.94–1.00) | 98.8% (0.96–1.00) | 0.976(0.953–0.999) |
| Sysmex XN-9000-I | 96.4% (0.92–0.99) | 98.2% (0.94–1.00) | 97.3% (0.95–0.99) | 0.946(0.911–0.981) |
| Sysmex XN-9000-F | 100% (0.98–1.00) | 99.4% (0.97–1.00) | 99.7% (0.98–1.00) | 0.994(0.982–1.000) |
| Hand counting | 94.6% (0.90–0.98) | 93.9% (0.89–0.97) | 94.3% (0.91–0.96) | 0.886(0.836–0.936) |
| LH780 | 96.6% (0.82–1.00) | 100% (0.99–1.00) | 99.7% (0.98–1.00) | 0.981(0.943–1.000) |
| BC-6000 Plus-O | 100% (0.88–1.00) | 99.7% (0.98–1.00) | 99.7% (0.98–1.00) | 0.981(0.945–1.000) |
| BC-6000 Plus-I | 100% (0.88–1.00) | 99.3% (0.98–1.00) | 99.4% (0.98–1.00) | 0.963(0.913–1.000) |
| Sysmex XN-9000-O | 100% (0.88–1.00) | 100% (0.99–1.00) | 100% (0.98–1.00) | 1.000(1.000–1.000) |
| Sysmex XN-9000-I | 100% (0.88–1.00) | 99.3% (0.98–1.00) | 99.4% (0.98–1.00) | 0.963(0.913–1.000) |
| Sysmex XN-9000-F | 96.6% (0.82–1.00) | 100% (0.99–1.00) | 99.7% (0.98–1.00) | 0.981(0.943–1.000) |
| Hand counting | 89.7% (0.73–0.98) | 99.3% (0.98–1.00) | 98.5% (0.96–1.00) | 0.904(0.821–0.987) |