Performance Evaluation of Automated Impedance and Optical Fluorescence Platelet Counts Compared With International Reference Method in Patients With Thalassemia.

Chaicharoen Tantanate, Ladawan Khowawisetsut, Kovit Pattanapanyasat1.   

Abstract

CONTEXT: - Spurious platelet counts from automated methods have been reported in patients with abnormal red blood cells. However, there is no specific study regarding performance of platelet counts by automated methods in patients with thalassemia.
OBJECTIVE: - To investigate the performance of automated platelet counts, including impedance (PLT-I) and optical fluorescent (PLT-O and PLT-F) methods, and compare them with the international reference method (IRM) for platelet counting in patients with thalassemia.
DESIGN: - Two hundred forty-nine thalassemia specimens from various subtypes were examined. PLT-I, PLT-O, and PLT-F from a Sysmex XN analyzer were evaluated and compared against the IRM. Demographic data, platelet counts, and red blood cell parameters are shown. Comparability between evaluated methods and IRM, as well as test characteristics, is presented. Factors involving inaccurate PLT-I were analyzed.
RESULTS: - Specimens with platelet counts ranging from 31 × 103/μL to 932 × 103/μL were included. Most patients were patients with thalassemia major. Correlation between PLT-I and IRM was lower than that of the other methods in overall patients. PLT-O and PLT-F were correlated to IRM when classifying patients according to clinically significant platelet ranges. All automated methods had acceptable sensitivities; however, specificity of PLT-I was low for diagnosis of thrombocytopenia. High RDW-CV (red blood cell distribution width-coefficient of variation) was an independent factor of inaccurate PLT-I measurement.
CONCLUSIONS: - Among the evaluated methods, PLT-I was the method least correlated to IRM, with PLT-O and PLT-F comparable to IRM in patients with thalassemia. Optical platelet counts and careful blood smear examination are recommended alternative platelet counting methods, depending on the clinical setting.

Entities:  

Mesh:

Year:  2017        PMID: 28402168     DOI: 10.5858/arpa.2016-0222-OA

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  4 in total

1.  Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting.

Authors:  Yi Sun; Zuojian Hu; Zhili Huang; Huaping Chen; Shanzi Qin; Zhong Jianing; Siyuan Chen; Xue Qin; Yi Ye; Chengbin Wang
Journal:  PLoS One       Date:  2019-05-24       Impact factor: 3.240

2.  Performance evaluation of optical platelet counting of BC-6000Plus automated hematology analyzer.

Authors:  Xuehua Hu; Yuqun Tang; Mingyi Li; Chunyun Fu; Zuoxin Deng; Wenting Tang; Huideng Zhou; Yan Chen; Yingfeng Li
Journal:  Ann Transl Med       Date:  2021-03

3.  Use of Middleware Data to Dissect and Optimize Hematology Autoverification.

Authors:  Rachel D Starks; Anna E Merrill; Scott R Davis; Dena R Voss; Pamela J Goldsmith; Bonnie S Brown; Jeff Kulhavy; Matthew D Krasowski
Journal:  J Pathol Inform       Date:  2021-04-07

4.  Alinity hq platelet count is not impacted by severe microcytosis.

Authors:  Gabriella Lakos; Zainab Mukhtar; Loredana Masi; Sabatino Valente; Fabrizio Papa
Journal:  J Clin Lab Anal       Date:  2022-03-11       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.