Hanah Kim1, Mina Hur1, Gun-Hyuk Lee1, Seung-Wan Kim1, Hee-Won Moon1, Yeo-Min Yun1.
Abstract
The performance of platelet (PLT) counting in thrombocytopenic samples is crucial for transfusion decisions. We compared PLT counting and its reproducibility between Mindray BC-6800Plus (BC-6800P, Mindray, Shenzhen, China) and Sysmex XN-9000 (XN, Sysmex, Kobe, Japan), especially focused on thrombocytopenic samples. We analyzed the correlation and agreement of PLT-I channels in both analyzers and BC-6800P PLT-O mode and XN PLT-F channel in 516 samples regarding PLT counts. Ten thrombocytopenic samples (≤2.0 × 109/L by XN PLT-F) were measured 10 times to investigate the reproducibility with the desirable precision criterion, 7.6%. The correlation of BC-6800P PLT-I and XN PLT-I was arranged moderate to very high; but the correlation of BC-6800P PLT-O and XN PLT-F was arranged high to very high. Both BC-6800P PLT-I vs. XN PLT-I and BC-6800P PLT-O vs. XN PLT-F showed very good agreement (κ = 0.93 and κ = 0.94). In 41 discordant samples between BC-6800P PLT-O and XN PLT-F at transfusion thresholds, BC-6800P PLT-O showed higher PLT counts than XN-PLT-F, except the one case. BC-6800P PLT-O exceeded the precision criterion in one of 10 samples; but XN PLT-F exceeded it in six of 10 samples. BC-6800P would be a reliable option for PLT counting in thrombocytopenic samples with good reproducibility.Entities:
Keywords: Mindray BC-6800Plus; Sysmex XN-9000; platelet; precision; thrombocytopenia
Year: 2021 PMID: 35054235 PMCID: PMC8775070 DOI: 10.3390/diagnostics12010068
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Correlation and differences of platelet counts using different channels between Mindray BC-6800Plus and Sysmex XN-9000.
| PLT Count * (×109/L) | BC-6800P PLT-I vs. XN PLT-I | BC-6800P PLT-O vs. XN PLT-F | ||||
|---|---|---|---|---|---|---|
| Equation | r (95% CI) | Mean Difference | Equation | r (95% CI) | Mean Difference | |
| ≤10 ( | y = 0.98x − 0.13 | 0.80 (0.64–0.89) | −0.2 (−6.2–5.8) | y = 1.14x + 0.92 | 0.83 (0.70–0.91) | −2.1 (−6.5–2.3) |
| 11–20 ( | y = 1.00x + 0.00 | 0.57 (0.43–0.68) | −0.2 (−8.5–8.1) | y = 1.25x − 2.50 | 0.76 (0.67–0.83) | −1.8 (−6.0–2.4) |
| 21–50 ( | y = 1.15x − 5.20 | 0.83 (0.76–0.88) | 3.6 (−68.1–75.3) | y = 1.00x + 2.00 | 0.94 (0.91–0.96) | −1.6 (−6.9–3.8) |
| 51–100 ( | y = 1.07x − 4.36 | 0.94 (0.89–0.96) | −0.9 (−11.8–10.1) | y = 1.04x − 1.07 | 0.93 (0.89–0.96) | −2.5 (−13.5–8.5) |
| >100 ( | y = 1.04x − 1.58 | 0.99 (0.98–0.99) | −7.7 (−30.4–15.0) | y = 0.96x + 5.24 | 0.99 (0.99–0.99) | 3.2 (−16.5–22.9) |
| Total ( | y = 1.03x − 0.61 | 0.99 (0.99–0.99) | −2.3 (−39.7–35.1) | y = 0.98x + 2.62 | 1.00 (1.00–1.00) | 0.0 (−13.9–14.0) |
All p-values were <0.001. * PLT counts measured by XN-F were considered as reference. Abbreviations: BC-6800P, Mindray BC-6800Plus; CI, confidence interval; PLT, platelet; XN, Sysmex XN-9000.
Agreement of platelet counts using different channels between Mindray BC-6800Plus and Sysmex XN-9000.
| XN PLT-I (×109/L) | BC-6800P PLT-I (×109/L) | XN PLT-F (×109/L)Total ( | BC-6800P PLT-O (×109/L) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| ≤10 | 11–20 | 21–50 | 51–100 | >100 | ≤10 | 11–20 | 21–50 | 51–100 | >100 | ||
| ≤10 ( | 30 | 2 | 0 | 0 | 0 | ≤10 ( | 28 | 10 | 0 | 0 | 0 |
| 11–20 ( | 5 | 64 | 19 | 0 | 0 | 11–20 ( | 0 | 82 | 30 | 0 | 0 |
| 21–50 ( | 0 | 15 | 121 | 4 | 0 | 21–50 ( | 0 | 1 | 109 | 1 | 0 |
| 51–100 ( | 0 | 0 | 3 | 54 | 3 | 51–100 ( | 0 | 0 | 0 | 57 | 3 |
| >100 ( | 0 | 0 | 1 | 0 | 195 | >100 ( | 0 | 0 | 0 | 0 | 195 |
| Cohen’s weighted kappa = 0.93 (0.91–0.95) | Cohen’s weighted kappa = 0.94 (0.93–0.96) | ||||||||||
* Cohen’s weighted kappa agreement in PLT counts ≤ 50 × 109/L (n = 261). Abbreviations: BC-6800P, Mindray BC-6800Plus; PLT, platelet; XN, Sysmex XN-9000.
Figure 1Discordant platelet (PLT) counts at transfusion threshold (10 × 109/L and 20 × 109/L) between BC-6800Plus PLT-O and Sysmex XN-9000 PLT-F (n = 41). Dark grey indicates PLT counts of BC-6800Plus PLT-O, and light grey indicates PLT counts of Sysmex XN-9000 PLT-F. The solid line indicates 10 × 109/L, and the dashed line indicates 20 × 109/L. Except case 41 (indicted by red rectangle), PLT counts of BC-6800Plus were higher than PLT counts of Sysmex XN-9000 PLT-F.
The precision of platelet counts using different channels of Mindray BC-6800Plus and Sysmex XN-9000 in 10 thrombocytopenic samples.
| PLT-I | PLT-O (or PLT-F) | |||||||
|---|---|---|---|---|---|---|---|---|
| BC-6800P | XN | BC-6800P | XN | |||||
| Mean (SD), ×109/L | CV, % | Mean (SD), ×109/L | CV, % | Mean (SD), ×109/L | CV, % | Mean (SD), ×109/L | CV, % | |
| 1 | 13.5 (1.2) | 8.7 | 20.1 (3.3) | 16.3 | 18.5 (0.5) |
| 15.9 (1.2) | 21.7 |
| 2 | 13.7 (2.0) | 14.6 | 16.2 (2.3) | 14.2 | 13.2 (0.6) |
| 12.1 (1.3) | 10.6 |
| 3 | 13.9 (1.8) | 12.9 | 17.3 (2.8) | 16.4 | 16.2 (0.9) |
| 13.9 (1.4) | 9.9 |
| 4 | 7.9 (2.6) | 32.9 | 9.7 (2.7) | 27.9 | 8.9 (0.3) |
| 7.6 (0.7) | 9.2 |
| 5 | 14.3 (1.2) | 8.1 | 15.8 (1.5) | 9.3 | 18.1 (0.7) |
| 15.4 (0.5) |
|
| 6 | 18.3 (1.3) |
| 17.8 (1.7) |
| 20.3 (0.5) |
| 16.2 (0.6) |
|
| 7 | 18.0 (1.3) |
| 18.6 (1.5) |
| 21.0 (0.8) |
| 16.9 (1.0) |
|
| 8 | 12.1 (1.7) | 13.7 | 11.8 (3.1) | 26.1 | 7.0 (0.7) | 9.5 | 3.9 (0.3) | 8.1 |
| 9 | 16.8 (2.0) | 12.2 | 20.2 (4.5) | 22.1 | 11.2 (0.6) |
| 8.8 (0.8) | 9.0 |
| 10 | 17.7 (2.1) | 11.9 | 21.5 (2.2) | 10.1 | 19.3 (0.5) |
| 18.5 (0.5) |
|
Bold types indicate less than desirable specification of precision (7.6 CV%) from the EFLM Biological Variation Database [25]. Abbreviations: BC-6800P, Mindray BC-6800Plus; CV%, percent coefficient of variation; PLT, platelet; SD, standard deviation; XN, Sysmex XN-9000.
The precision of clinical reportable platelet parameters of Mindray BC-6800Plus and Sysmex XN-9000 in 10 thrombocytopenic samples.
| IPF (%) | MPV (fL) | PCT (%) | PDW (fL) | P-LCR (%) | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BC-6800P | XN | BC-6800P | XN | BC-6800P | XN | BC-6800P | XN | BC-6800P | XN | |||||||||||
| Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | Mean (SD) | CV, % | |
| 1 | 7.3 (0.4) | 5.5 | 2.7 (0.5) | 20.4 | 10.2 (0.6) | 6.2 | 12.5 (0.6) | 4.6 | 0.026 (0.040) | 152.092 | 0.03 (0.00) | 13.23 | 15.9 (0.6) | 3.6 | 14.1 (1.4) | 9.9 | 29.4 (5.2) | 17.7 | 42.5 (2.5) | 5.9 |
| 2 | 2.6 (0.6) | 21.5 | 0.9 (0.2) | 27.7 | 9.4 (0.8) | 8.5 | 12.5 (1.6) | 12.5 | 0.013 (0.003) | 20.165 | 0.02 (0.01) | 35.36 | 16.5 (0.7) | 4.5 | 13.2 (6.3) | 48.1 | 26.5 (6.4) | 24.2 | 42.6 (7.6) | 17.8 |
| 3 | 8.4 (0.8) | 9.2 | 10.9 (1.6) | 12.5 | 10.7 (0.7) | 6.4 | NA | NA | 0.015 (0.003) | 18.298 | NA | NA | 16.0 (0.3) | 1.7 | NA | NA | 40.0 (4.1) | 10.3 | NA | NA |
| 4 | 9.3 (0.8) | 8.9 | 11.1 (1.5) | 13.1 | 10.2 (1.1) | 11.1 | NA | NA | 0.009 (0.003) | 33.432 | NA | NA | 15.8 (0.5) | 3.2 | NA | NA | 38.1 (9.9) | 25.9 | NA | NA |
| 5 | 1.5 (0.3) | 19.2 | 0.4 (0.2) | 46.1 | 8.2 (0.3) | 4.2 | 10.2 (0.6) | 6.1 | 0.012 (0.001) | 10.248 | 0.02 (0.00) | 28.41 | 15.3 (0.5) | 3.2 | 11.3 (1.5) | 13.7 | 13.3 (3.3) | 25.0 | 26.6 (4.0) | 14.9 |
| 6 | 5.1 (0.4) | 7.8 | 5.4 (0.5) | 9.3 | 11.0 (0.5) | 4.8 | 12.4 (1.0) | 7.9 | 0.020 (0.002) | 9.377 | 0.02 (0.00) | 15.79 | 16.8 (0.4) | 2.4 | 15.5 (1.7) | 10.7 | 35.4 (3.4) | 9.7 | 43.3 (4.7) | 10.8 |
| 7 | 4.2 (0.5) | 12.4 | 1.5 (0.4) | 25.3 | 10.3 (0.6) | 6.0 | 11.8 (0.4) | 3.6 | 0.035 (0.051) | 147.762 | 0.02 (0.00) | 15.06 | 16.1 (0.4) | 2.6 | 15.3 (2.8) | 18.4 | 28.4 (4.0) | 14.0 | 41.1 (3.8) | 9.2 |
| 8 | 4.2 (0.8) | 18.5 | 5.3 (1.1) | 21.3 | 9.9 (0.7) | 7.5 | NA | NA | 0.012 (0.003) | 23.011 | NA | NA | 16.4 (0.2) | 1.2 | NA | NA | 35.9 (4.2) | 11.7 | NA | NA |
| 9 | 2.1 (0.5) | 25.5 | 1.6 (0.4) | 23.8 | 10.1 (0.9) | 8.8 | NA | NA | 0.017 (0.004) | 21.785 | NA | NA | 16.0 (0.5) | 3.2 | NA | NA | 34.8 (7.8) | 22.4 | NA | NA |
| 10 | 5.4 (0.5) | 9.4 | 6.7 (0.7) | 10.4 | 10.1 (0.8) | 7.6 | 11.8 (0.8) | 6.4 | 0.022 (0.004) | 16.421 | 0.03 (0.01) | 18.75 | 16.9 (0.6) | 3.5 | 14.6 (1.6) | 10.7 | 29.8 (4.1) | 13.6 | 38.9 (3.6) | 9.4 |
Abbreviations: BC-6800P, Mindray BC-6800Plus; CV%, percent coefficient of variation; NA, not available; SD, standard deviation; XN, Sysmex XN-9000.