Literature DB >> 26629371

Accuracy of Platelet Counting by Optical and Impedance Methods in Patients with Thrombocytopaenia and Microcytosis.

Mohamed-Rachid Boulassel1, Raya Al-Farsi2, Sulaiman Al-Hashmi3, Hamad Al-Riyami3, Hammad Khan2, Salam Al-Kindi4.   

Abstract

OBJECTIVES: Obtaining accurate platelet counts in microcytic blood samples is challenging, even with the most reliable automated haematology analysers. The CELL-DYN(™) Sapphire (Abbott Laboratories, Chicago, Illinois, USA) analyser uses both optical density and electronic impedance methods for platelet counting. This study aimed to evaluate the accuracy of optical density and electrical impedance methods in determining true platelet counts in thrombocytopaenic samples with microcytosis as defined by low mean corpuscular volume (MCV) of red blood cells. Additionally, the impact of microcytosis on platelet count accuracy was evaluated.
METHODS: This study was carried out between February and December 2014 at the Haematology Laboratory of the Sultan Qaboos University Hospital in Muscat, Oman. Blood samples were collected and analysed from 189 patients with thrombocytopaenia and MCV values of <76 femtolitres. Platelet counts were tested using both optical and impedance methods. Stained peripheral blood films for each sample were then reviewed as a reference method to confirm platelet counts.
RESULTS: The platelet counts estimated by the impedance method were on average 30% higher than those estimated by the optical method (P <0.001). The estimated intraclass correlation coefficient was 0.52 (95% confidence interval: 0.41-0.62), indicating moderate reliability between the methods. The degree of agreement between methods ranged from -85.5 to 24.3 with an estimated bias of -30, suggesting that these methods generate different platelet results.
CONCLUSION: The impedance method significantly overestimated platelet counts in microcytic and thrombocytopaenic blood samples. Further attention is therefore needed to improve the accuracy of platelet counts, particularly for patients with conditions associated with microcytosis.

Entities:  

Keywords:  Anemia; Electrical Impedance; Mean Corpuscular Volume; Optical Devices; Platelet Counts; Thrombocytopenia

Year:  2015        PMID: 26629371      PMCID: PMC4664089          DOI: 10.18295/squmj.2015.15.04.004

Source DB:  PubMed          Journal:  Sultan Qaboos Univ Med J        ISSN: 2075-051X


  10 in total

1.  Immunoplatelet counting: a proposed new reference procedure.

Authors:  P Harrison; A Horton; D Grant; C Briggs; S MacHin
Journal:  Br J Haematol       Date:  2000-02       Impact factor: 6.998

2.  Platelet counting by the coulter LH 750, sysmex XE 2100, and advia 120: a comparative analysis using the RBC/platelet ratio reference method.

Authors:  Linda M Sandhaus; Ebenezer S Osei; Neeta N Agrawal; Chrisine A Dillman; Howard J Meyerson
Journal:  Am J Clin Pathol       Date:  2002-08       Impact factor: 2.493

3.  The accuracy of platelet counting in thrombocytopenic blood samples distributed by the UK National External Quality Assessment Scheme for General Haematology.

Authors:  Barbara J De la Salle; Paul N McTaggart; Carol Briggs; Paul Harrison; Caroline J Doré; Ian Longair; Samuel J Machin; Keith Hyde
Journal:  Am J Clin Pathol       Date:  2012-01       Impact factor: 2.493

4.  Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion.

Authors:  H C Segal; C Briggs; S Kunka; A Casbard; P Harrison; S J Machin; M F Murphy
Journal:  Br J Haematol       Date:  2005-02       Impact factor: 6.998

Review 5.  Continuing developments with the automated platelet count.

Authors:  C Briggs; P Harrison; S J Machin
Journal:  Int J Lab Hematol       Date:  2007-04       Impact factor: 2.877

6.  Forecasting hemoglobinopathy burden through neonatal screening in Omani neonates.

Authors:  Salam Alkindi; Shoaib Al Zadjali; Ali Al Madhani; Shahina Daar; Hamood Al Haddabi; Qamariya Al Abri; David Gravell; Tsouria Berbar; Sahaya Pravin; Anil Pathare; Rajagopal Krishnamoorthy
Journal:  Hemoglobin       Date:  2010-01       Impact factor: 0.849

7.  Evaluation of low platelet counts by optical, impedance, and CD61-immunoplatelet methods: estimation of possible inappropriate platelet transfusion.

Authors:  Joan Cid; Janilson Do Nascimento; Ana Vicent; Reyes Aguinaco; Lourdes Escoda; Aranzazu Ugarriza; Andreu Llorente
Journal:  Transfusion       Date:  2009-11-20       Impact factor: 3.157

8.  Counting platelets at transfusion threshold levels: impact on the decision to transfuse. A BEST Collaborative - UK NEQAS(H) International Exercise.

Authors:  M Lozano; A Mahon; P F van der Meer; S Stanworth; J Cid; D Devine; M K Fung; B de la Salle; N M Heddle
Journal:  Vox Sang       Date:  2013-12-12       Impact factor: 2.144

9.  Enhanced accuracy of optical platelet counts in microcytic anemia.

Authors:  Lin-Lin Pan; Ching-Mei Chen; Wan-Ting Huang; Cheuk-Kwan Sun
Journal:  Lab Med       Date:  2014

10.  A community-based study of common hereditary blood disorders in Oman.

Authors:  A A Al-Riyami; A J Suleiman; M Afifi; Z M Al-Lamki; S Daar
Journal:  East Mediterr Health J       Date:  2001-11       Impact factor: 1.628

  10 in total
  3 in total

1.  Stability of fetal hemoglobin levels in patients receiving metformin therapy.

Authors:  Mohamed-Rachid Boulassel; Abdal-Aziz Issa El-Hussain; Moataz Mohamed Hassan; Mohamed-Lamine Toumi; Sameha Merzoug; Zahra Al-Qarni; Hammad Khan; Zied Gaifer; Khalil Al-Farsi; Murtadha Al-Khabori; Mohamed Al Huneini; Salam Al-Kindi
Journal:  Haematologica       Date:  2018-04-26       Impact factor: 9.941

2.  Compare the accuracy and precision of Coulter LH780, Mindray BC-6000 Plus, and Sysmex XN-9000 with the international reference flow cytometric method in platelet counting.

Authors:  Yi Sun; Zuojian Hu; Zhili Huang; Huaping Chen; Shanzi Qin; Zhong Jianing; Siyuan Chen; Xue Qin; Yi Ye; Chengbin Wang
Journal:  PLoS One       Date:  2019-05-24       Impact factor: 3.240

3.  Alinity hq platelet count is not impacted by severe microcytosis.

Authors:  Gabriella Lakos; Zainab Mukhtar; Loredana Masi; Sabatino Valente; Fabrizio Papa
Journal:  J Clin Lab Anal       Date:  2022-03-11       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.