| Literature DB >> 31068187 |
Kai Xie1,2, Hongfei Sun1,2, Liugang Gao1,2, Tao Lin1,2, Jianfeng Sui1,2, Xinye Ni3,4.
Abstract
BACKGROUND: The radiation transmission through the multileaf collimators is undesired in modern techniques such as volumetric modulated arc therapy (VMAT). According to identical plans, in this study, we aim to investigate the dosimetric impact of jaw tracking on the VMAT plans on two adjacent targets.Entities:
Keywords: Adjacent targets; Dosimetry; Fixed-jaw; MLC transmission; VMAT
Mesh:
Year: 2019 PMID: 31068187 PMCID: PMC6505249 DOI: 10.1186/s13014-019-1284-2
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Comparison of PV and OAR doses between O-VMAT and F-VMAT
| Regions of interest | O-VMAT | F-VMAT | |
|---|---|---|---|
| PTV | |||
| CI | 0.831 ± 0.093 | 0.830 ± 0.090 | 0.833 |
| HI | 1.055 ± 0.014 | 1.050 ± 0.011 | 0.157 |
| Spinal cord | |||
| Dmax (cGy) | 3625.19 ± 376.55 | 3575.04 ± 343.55 | 0.208 |
| Spinal cord PRV | |||
| Dmax (cGy) | 4148.33 ± 280.23 | 4124.20 ± 289.31 | 0.263 |
| Intestine | |||
| V20 (%) | 29.68 ± 12.49 | 30.15 ± 11.89 | 0.624 |
| V30 (%) | 16.68 ± 11.66 | 16.07 ± 11.65 | 0.161 |
| V40 (%) | 9.44 ± 9.10 | 9.72 ± 9.41 | 0.123 |
| interOARa | |||
| Dmean (cGy) | 654.96 ± 113.38 | 490.84 ± 80.26 | 0.018 |
| V3 (%) | 70.11 ± 17.86 | 40.17 ± 12.91 | 0.018 |
| V5 (%) | 37.40 ± 12.86 | 26.71 ± 7.03 | 0.018 |
| V10 (%) | 17.47 ± 4.56 | 13.25 ± 2.26 | 0.018 |
| MUs | 766.40 ± 97.46 | 1133.21 ± 162.11 | 0.012 |
a: except for the first case because it was an outlier in the box plot
Fig. 1Relationship among Dmean of interOAR and distance between PTV1 and PTV2
Fig. 2Dose distribution of interOAR in different places from one patient. a 10 mm from PTV1 in O-VMAT; (b) 10 mm from PTV1 in F-VMAT; (c) 10 mm from PTV2 in O-VMAT; (d) 10 mm from PTV2 in F-VMAT; (e) at the center of interOAR in O-VMAT, and (f) at the center of interOAR in F-VMAT