Dosimetric comparison between jaw tracking and static jaw techniques in intensity-modulated radiotherapy.

Zhongsu Feng1, Hao Wu2, Yibao Zhang3, Yunjun Zhang4, Jinsheng Cheng5, Xu Su6.   

Abstract

PURPOSE: To compare the dosimetric differences between jaw tracking technique (JTT) and static jaw technique (SJT) in dynamic intensity-modulated radiotherapy (d-IMRT) and assess the potential advantages of jaw tracking technique.
METHODS: Two techniques, jaw tracking and static jaw, were used respectively to develop the d-IMRT plans for 28 cancer patients with various lesion sites: head and neck, lungs, esophageal, abdominal, prostate, rectal and cervical. The dose volume histograms (DVH) and selected dosimetric indexes for the whole body and for organs at risk (OARs) were compared. A two dimensional ionization chamber Array Seven29 (PTW, Freiburg, Germany) and OCTAVIUS Octagonal phantom (PTW, Freiburg, Germany) were used to verify all the plans.
RESULTS: For all patients, the treatment plans using both techniques met the clinical requirements. The V5, V10, V20, V30, V40 (volumes receiving 5, 10, 20, 30 and 40 Gy at least, respectively), mean dose (Dmean) for the whole body and V5, V10, V20, Dmean for lungs in the JTT d-IMRT plans were significantly less than the corresponding values of the SJT d-IMRT plans (p < 0.001). The JTT d-IMRT plans deposited lower maximum dose (Dmax) to the lens, eyes, brainstem, spinal cord, and right optic nerve, the doses reductions for these OARs ranged from 2.2% to 28.6%. The JTT d-IMRT plans deposited significantly lower Dmean to various OARs (all p values < 0.05), the mean doses reductions for these OARs ranged from 1.1% to 31.0%, and the value reductions depend on the volume and the location of the OARs. The γ evaluation method showed an excellent agreement between calculation and measurement for all techniques with criteria of 3%/3 mm.
CONCLUSIONS: Both jaw tracking and static jaw d-IMRT plans can achieve comparable target dose coverage. JTT displays superior OARs sparing than SJT plans. These results are of clinical importance, especially for the patients with large and complex targets but close to some highly radio-sensitive organs to spare, and for patients with local recurrent or secondary primary malignant lesion within a previously irradiated area.

Entities:  

Mesh:

Year:  2015        PMID: 25623899      PMCID: PMC4326511          DOI: 10.1186/s13014-015-0329-4

Source DB:  PubMed          Journal:  Radiat Oncol        ISSN: 1748-717X            Impact factor:   3.481


Introduction

Nowadays, dynamic intensity-modulated radiotherapy (d-IMRT) is widely used in the treatment of cancer patients. Compared with three dimensional conformal radiotherapy, d-IMRT allows to increase the dose conformity of the target while decreasing normal tissue doses [1]. In dynamic IMRT plans, the multi-leaf collimators (MLCs) translate continuously at variable speeds during the irradiation while the upper and lower jaws stay static. According to the study by LoSasso et al. [2], the MLC transmission increases with increasing jaw field size and beam energy. Cadman et al. [3] found that the transmission through the jaw and the MLC together is smaller than 0.1%. Jaw tracking technique provided by linear accelerators keeps jaws during dose delivery as close as possible to the MLC aperture, and further minimizes leakage and transmission through the MLC leaves. Joy et al. [4] evaluated the dosimetric effects of jaw tracking in step-and-shoot IMRT, but failed to indicate which patients would benefit most from jaw tracking. The dosimetric benefits of jaw tracking for prostate and head and neck (H&N) patients using d-IMRT and volumetric-modulated arc therapy (VMAT) were also evaluated by others [5,6]. It showed that, the organs far from the target showed larger sparing in jaw-tracking static arc than the organs adjacent to the target. TrueBeam (Varian Medical Systems, Palo Alto, CA) is a new generation of linear accelerator providing jaw tracking technique (JTT), which traces the MLC aperture with jaws to minimize the leakage and transmission of the MLC leaves, hence further reduces the OAR doses adjacent to the target, and potentially improves the dose fall-off towards the surrounding critical structures. The main purpose of this study is to assess the potential advantages of the JTT provided by TrueBeam accelerators in reducing the doses to the organs at risk (OARs) while preserving adequate target dose coverage, and to evaluate which patients and which OARs would benefit most from JTT plans.

Materials and methods

Ethical consideration and consent

The use of the radiotherapy database for retrospective research has been approved by the committee of the Ethical Review Board of the Chinese Center for Disease Control and Prevention (No. LLSC2014-005). This research was waivered informed consent.

Patient characteristics

Twenty eight cancer patients (median age 58 years, range 31–77 years) treated with radiotherapy in our department were retrospectively included in this study. The lesion distributions of the selected cases included head and neck, lungs, esophageal, abdominal, prostate, rectal and cervical, as shown in Table 1. One local recurrent nasopharyngeal carcinoma (NPC) patient was included in the H&N group.
Table 1

Lesion distributions and treatment characteristics of the selected cases

Lesion sites n Total dose (Gy) PTV (cm 3 )
GTV CTV
Mean dose (range) Mean dose (range) Mean SD
Head and neck666.4(60–70)55.8(45–60)693.38168.8
Lungs560.7(56–66)54.9(50.6-60)27050.07
Esophageal465(64,66)50536.2839.4
Abdominal353.5(50–60)46(45–48)454.1181.24
Prostate27265(60,70)174.8535.75
Rectal650.641.81163.5755.14
Cervical2-47.5(45,50)1204.05139.45

Abbreviations: SD standard deviation.

Lesion distributions and treatment characteristics of the selected cases Abbreviations: SD standard deviation.

Contouring of targets and OARs

All patients were immobilized with custom-made thermoplastic mask and underwent computed tomography (CT) scan with slice thickness of 3 mm (for head and neck patients) or 5 mm (for thoracic, abdominal and pelvic patients ) using the Siemens Somatom Sensation Open 40 slice CT scanner (Siemens, Forchheim, Germany). The targets and OARs were contoured slice by slice on the treatment planning CT images. Relevant organs were delineated for different anatomic regions. The body was segmented automatically by the Eclipse (Versions 11.0) Treatment Planning System (TPS) (Varian Medical Systems, Palo Alto, CA). Isotropic margins of 3 mm (for head and neck patients) or 5 mm (for thoracic, abdomen and pelvic patients) were created around the clinical target volume (CTV) and gross tumor volume (GTV) to generate the planning target volume (PTV) and planning gross tumor volume (PGTV), respectively.

Treatment planning

The static jaw technique (SJT) d-IMRT plans were created using the Eclipse TPS using sliding window dynamic delivery and fixed beam angles. The selected energies were 6 MV (for head and neck patients) and 10 MV (for thoracic, abdominal and pelvic patients) respectively from a TrueBeam linear accelerator. The collimator angles were optimized according to the target shape. After IMRT optimization using the dose volume optimizer algorithm (version 11.0.31), the volumetric doses were calculated using the anisotropic analytical algorithm (version 11.0.31) with a dose calculation grid of 2 mm. The reference volume for the treatment planning was PTV. The normal tissue dose volume constraints used for the treatment planning are listed in Table 2. For all patients, the dose distributions met the planning requirements, i.e. at least 95% of the PTV received the prescribed dose, while the doses to the surrounding tissues were minimized. To create JTT d-IMRT plans, the SJT d-IMRT plans were duplicated and jaw tracking function was selected when calculating the leaf motions and volumetric doses. All the machine parameters and optimization parameters were identical to those of the SJT d-IMRT plans. The JTT plans reduce the dose leakage and transmission through the MLCs, which required renormalization to achieve the same PTV coverage as in the SJT d-IMRT plans.
Table 2

Dose constraints used in treatment planning

Organ Dose volume parameters Organ Dose volume parameters
LensDmax < 7GyHeartDmean < 26Gy
EyeDmax < 50GyV30 < 50%
Dmean < 35GyLiverDmean < 30Gy
Optic nerveDmax < 54Gy1/3 volume of liver avoid irradiation
Brain stemDmax < 54GyKidneyDmean < 18Gy
ParotidV30 < 50%V12 < 55%
Dmean < 26GyV20 < 32%
Spinal cordDmax < 40GyIntestineV40 < 30%
EsophagusV35 < 50%BladderV30 < 50%
Dmean < 34GyFemoral HeadV20 < 50%
LungV5 < 70%V30 < 15%
V10 < 50%RectumV30 < 50%
V20 < 30%V60 < 35%
Dmean < 15GyV70 < 20%
Dose constraints used in treatment planning

Plan evaluation

The DVHs and the isodose curves of the JTT plans were compared with that of the SJT plans. The dose changes to the whole body were evaluated using V5, V10, V20, V30, V40, Dmax and Dmean as listed in Table 3. The parameters of V5, V10, V20, Dmax, and Dmean were used to evaluate the dose changes to the lungs, as shown in Table 4. The dosimetric disparities of OARs were evaluated using the maximum dose (Dmax) and mean dose (Dmean) as listed in Table 5 and Table 6.
Table 3

Comparison of the whole body doses between the two techniques ( )

Body SJT JTT p
V5 (%)37.2 ± 19.136.1 ± 18.6<0.001
V10 (%)29.0 ± 16.528.2 ± 16.0<0.001
V20 (%)19.9 ± 12.919.3 ± 12.4<0.001
V30 (%)12.0 ± 9.611.6 ± 9.2<0.001
V40 (%)7.6 ± 7.07.5 ± 6.9<0.001
Dmax(Gy)60.3 ± 8.560.5 ± 8.80.071
Dmean(Gy)9.8 ± 6.09.6 ± 5.8<0.001
Table 4

Comparison of lung doses between the two techniques ( )

Lung SJT JTT p
V5 (%)48.3 ± 6.745.7 ± 5.9<0.001
V10 (%)34.9 ± 6.233.6 ± 5.8<0.001
V20 (%)20.5 ± 5.519.9 ± 5.1<0.001
Dmax(Gy)59.9 ± 6.460.00 ± 6.50.155
Dmean(Gy)11.4 ± 2.611.0 ± 2.5<0.001
Table 5

Comparison of OARs D between the two techniques ( )

OARs D max (Gy) SJT JTT p
Right lens5.6 ± 2.24.00 ± 1.70.031
Left lens5.7 ± 1.64.2 ± 1.30.031
Right eye26.3 ± 17.324.5 ± 17.80.031
Left eye26.6 ± 13.924.9 ± 14.90.031
Right optic nerve44.3 ± 8.943.4 ± 8.70.031
Left optic nerve41.8 ± 10.840.8 ± 12.00.156
Brain stem39.1 ± 12.037.2 ± 12.90.016
Right parotid62.8 ± 10.963.2 ± 11.40.063
Left parotid63.2 ± 7.363.6 ± 7.40.156
Spinal cord31.0 ± 8.030.3 ± 7.9<0.001
Esophagus62.9 ± 5.562.8 ± 5.50.500
Heart57.9 ± 5.957.9 ± 6.00.527
Liver49.2 ± 6.149.2 ± 6.10.094
Right kidney36.3 ± 15.936.3 ± 16.00.406
Left kidney37.2 ± 9.337.0 ± 9.10.406
Intestine50.6 ± 3.750.7 ± 3.90.313
Bladder56.1 ± 9.056.2 ± 8.90.156
Femoral head39.7 ± 5.239.8 ± 5.30.156
Rectum60.6 ± 8.960.2 ± 9.40.156
Table 6

Comparison of OARs D between the two techniques ( )

OARs D mean (Gy) SJT JTT p
Right lens4.9 ± 2.03.38 ± 1.440.031
Left lens5.0 ± 1.43.59 ± 1.070.031
Right eye7.1 ± 3.85.56 ± 3.550.031
Left Eye7.7 ± 2.65.99 ± 2.830.031
Right optic nerve25.5 ± 12.423.87 ± 13.330.031
Left optic nerve26.0 ± 11.124.37 ± 12.260.031
Brain stem23.7 ± 13. 522.03 ± 14.080.016
Right parotid29.3 ± 4.927.41 ± 4.020.031
Left parotid28.7 ± 4.127.23 ± 3.460.031
Spinal cord19.0 ± 8.018.1 ± 7.6<0.001
Esophagus29.4 ± 8.428.9 ± 8.10.016
Heart17.9 ± 11.917.5 ± 11.80.004
Liver12.2 ± 9.911.9 ± 9.80.031
Right kidney8.5 ± 4.18.2 ± 4.00.031
Left kidney6.6 ± 4.86.3 ± 4.50.031
Intestine27.4 ± 3.427.1 ± 3.40.016
Bladder33.5 ± 8.733.1 ± 8.50.004
Femoral head14.1 ± 1.613.7 ± 1.50.004
Rectum35.8 ± 4.535.2 ± 4.70.031
Comparison of the whole body doses between the two techniques ( ) Comparison of lung doses between the two techniques ( ) Comparison of OARs D between the two techniques ( ) Comparison of OARs D between the two techniques ( )

Plan verification

The Eclipse TPS was used to create plans for verification by converting the SJT and JTT plans to OCTAVIUS Octagonal phantoms (PTW, Freiburg, Germany). Then the volumetric doses were re-calculated, and the dose distributions in effective measurement points of the two dimensional ionization chamber Array Seven29 (PTW, Freiburg, Germany) were exported for all plans. By using the γ evaluation method of VeriSoft software (version 5.1) (PTW, Freiburg, Germany), the measured dose planes were compared with the computed dose distribution using criteria of 3% dose difference and 3 mm distance to agreement.

Statistical analysis

The data were presented as the averages of all patients followed by the standard deviation (). The PASW Statistics (version 18.0) software (SPSS; Chicago, IL, USA) was applied for statistical analysis. To compare the results of the two techniques, Wilcoxon Signed-Ranks test was performed with p values <0.05 considered as significant.

Results

As shown in the DVH comparison (Figure 1), JTT d-IMRT plans reduce doses to the normal tissues, especially in the low dose regions.
Figure 1

DVH comparison between SJT and JTT plans for different body sites: (a) H&N, (b) Thoracic, (c) Abdominal, (d) Pelvic case. CTV: Clinical target volume; PTV: Planning target volume; GTV: Gross tumor volume; PGTV: Planning gross tumor volume.

DVH comparison between SJT and JTT plans for different body sites: (a) H&N, (b) Thoracic, (c) Abdominal, (d) Pelvic case. CTV: Clinical target volume; PTV: Planning target volume; GTV: Gross tumor volume; PGTV: Planning gross tumor volume. Table 3 lists the comparison of the whole body doses between the two techniques. The JTT d-IMRT plans displayed significantly lower percentages of V5, V10, V20, V30, V40 and significantly lower mean doses of the whole body (p < 0.001). There was no significant difference of maximum doses delivered to the whole body (p = 0.071). Table 4 lists the statistical data of the lungs for all thoracic cases. The V5, V10, V20, Dmean of lungs in JTT d-IMRT plans are significantly lower than the corresponding values of the SJT d-IMRT plans (p < 0.001), the mean reduction was 2.6%, 1.3%, 0.6% and 0.4%, respectively. There was no significant difference of maximum doses delivered to the lungs (p = 0.155). This study demonstrates that the mean values of the maximum doses to the lens, eyes, brainstem, spinal cord, and right optic nerve are significantly reduced in the JTT plans (all p values <0.05), as shown in Table 5. The doses reduction for these OARs ranged from 2.2% to 28.6%, and the maximum dose to lens is reduced most obviously. Table 6 lists the mean doses to various OARs for all patients. The JTT d-IMRT plans give significantly lower mean doses to the studied OARs (all p values <0.05). The mean doses reduction for these OARs ranged from 1.1% to 31.0%, and the reduction values are depend on the volume and the location of the OARs. Treatment plan verification of the two techniques is generally performed with the γ evaluation method. The γ was compared locally, and the γ pass rate for SJT and JTT is 93.9 ± 1.98 and 94.1 ± 1.95, respectively. The comparison of γ between the two techniques showed no statistical difference (p = 0.132). Dose calculation and delivery are both accurate for SJT and JTT plans at criteria of 3%/3 mm using γ analysis.

Discussion

Jaw tracking methods have also been used in other treatment modalities aiming to achieve better target coverage and critical structure sparing. Joy et al. [4] evaluated the dosimetric effects of jaw tracking in step-and-shoot IMRT, showing an overall reduction of normal tissue doses. Most patients had reductions of V5, V10, and V20 by less than 2% in the normal tissues. Schmidhalter et al. [5] evaluated the leaf transmission reduction using moving jaws in dynamic MLC IMRT, and demonstrated that the undesired doses to the body volume minus the planning target volume decreased by up to 1.8% and 1.5% for prostate and H&N patients. Simultaneously, the MU increased by up to 3.1% and 2.8%, respectively. Kim et al. [6] assessed the potential advantages of jaw tracking technique by using control point sequences of VMAT planning, showing that, for H&N cases, the mean dose reductions for all the OARs ranged from 4.3% to 11.9%, and for prostate patients, the organs distant from the target were spared better in jaw-tracking static arc (JTSA) plans. The dose reductions were more significant in the dose regions of D80, D90 and D95 for all the patients in JTSA plans. However, previous studies [4-6] included no abdominal cases or other pelvic lesion except prostate: the number of OARs was less than that of this study. Moreover, they didn’t indicate which cases and which organs would benefit most from JTT. This study selected cases of various body sites. For large and complex targets, such as NPC, the large target volumes are surrounded by many critical neural tissues and sensitive structures such as lens, optic nerves, brain stem, parotid glands and spinal cord. This study demonstrates that the mean values of the maximum doses to the lens, eyes, brainstem, spinal cord, and right optic nerve are significantly reduced in the JTT plans (all p values <0.05) as shown in Table 5. Perhaps because these OARs do not overlap with the target in the selected cases, yet the left optic nerve and other OARs are partially or wholly covered by PTVs. The reduction of the maximum dose to OARs may reduce the risk of radiation injury [7,8], such as radiation induced myelitis and cataract. It is also more meaningful for the radiotherapy patients with local recurrent or second primary malignant lesion within or adjacent to a previously irradiated area. The JTT technique also significantly reduces the low dose regions. In the thoracic cases, the JTT technique significantly reduces the V5, V10, V20 to the lungs, as shown in Table 4, compared to the SJT plan (p < 0.001). The results may mostly benefit Hodgkin’s lymphoma patients whose lungs are largely within the treatment fields during radiotherapy. The JTT technique may reduce the risk of pulmonary toxicity since radiation pneumonitis rate is correlated with low dose volumes [9-12]. The risk of radiation induced secondary malignancies is another concern, which is strongly associated with low dose exposure of normal tissues during IMRT [13]. In addition, this study suggests that the JTT plans give significantly lower mean doses to various OARs (all p values <0.05, refer to Table 6 for details). The JTT associated OAR dose reductions are more obvious for the mean doses to the structures of small volume (such as lens) or structures far from the targets (such as left optic nerve, esophagus, and kidneys in this study). The mean doses of OARs have been widely used to predict the probability of radiation toxicity [9-12], such as pulmonary toxicity. Therefore, JTT plans may decrease the probability of developing late side effects or secondary neoplasm. Some problems exist in the current jaw tracking technique though. Firstly, jaw tracking is not considered in the optimization of JTT d-IMRT or in the step-and-shoot jaw tracking IMRT plans [4], as limited by the optimization algorithm of TPS. However, the Eclipse TPS (version 11.0) provides jaw tracking option in the VAMT plan optimization, which is applicable on Varian Truebeam accelerators. Secondly, the output factor of the accelerator varies dramatically in jaw tracking technique. Usually, the output factors are measured on the central axis of the beam. However, there are many off-axis fields in jaw tracking plans, which would affect the output factor and should be considered in treatment planning. Joy et al. [4] calculated the off-axis output factors and demonstrated decreases of 1% to 3% when shifting away from central axis. Although all the SJT and JTT plans in this study can pass the dosimetric verification, this issue was not solved so far and further research is needed. Thirdly, the performance of flattening filter-free (FFF) module combined with the jaw tracking technique of TrueBeam accelerator is still uncertain. There were several reports [14-19] about the clinical usage of FFF beams, indicating more potential of reducing the doses to OARs with comparable target dose coverage. The potential benefit of applying jaw tracking technique in FFF beams will be further studied in the future work.

Conclusions

This study demonstrates that all the SJT and JTT plans could meet the clinical objectives and are clinically applicable. The dose measurements agreed well with TPS calculated doses. JTT displays superior OARs sparing than SJT plans. These results are of clinical importance, especially for the patients with large and complex targets but close to some highly radio-sensitive organs to spare (such as lens or gonad), and for patients with local recurrent or secondary primary malignant lesion within a previously irradiated area. More studies on off-axis output factors and combined application of FFF module should be conducted in the future.
  19 in total

1.  Volumetric modulation arc radiotherapy with flattening filter-free beams compared with static gantry IMRT and 3D conformal radiotherapy for advanced esophageal cancer: a feasibility study.

Authors:  Giorgia Nicolini; Sarbani Ghosh-Laskar; Shyam Kishore Shrivastava; Sushovan Banerjee; Suresh Chaudhary; Jai Prakash Agarwal; Anusheel Munshi; Alessandro Clivio; Antonella Fogliata; Pietro Mancosu; Eugenio Vanetti; Luca Cozzi
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-02       Impact factor: 7.038

2.  Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams.

Authors:  S Subramaniam; S Thirumalaiswamy; C Srinivas; G A Gandhi; M Kathirvel; K K Kumar; S Mallik; M Babaiah; Y Pawar; A Clivio; A Fogliata; P Mancosu; G Nicolini; E Vanetti; L Cozzi
Journal:  Strahlenther Onkol       Date:  2012-03-10       Impact factor: 3.621

Review 3.  Current status and future perspective of flattening filter free photon beams.

Authors:  Dietmar Georg; Tommy Knöös; Brendan McClean
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

4.  Analysis of clinical and dosimetric factors associated with treatment-related pneumonitis (TRP) in patients with non-small-cell lung cancer (NSCLC) treated with concurrent chemotherapy and three-dimensional conformal radiotherapy (3D-CRT).

Authors:  Shulian Wang; Zhongxing Liao; Xiong Wei; Helen H Liu; Susan L Tucker; Chao-Su Hu; Rodhe Mohan; James D Cox; Ritsuko Komaki
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-09-25       Impact factor: 7.038

5.  Dose-volume modeling of the risk of postoperative pulmonary complications among esophageal cancer patients treated with concurrent chemoradiotherapy followed by surgery.

Authors:  Susan L Tucker; H Helen Liu; Shulian Wang; Xiong Wei; Zhongxing Liao; Ritsuko Komaki; James D Cox; Radhe Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-09-11       Impact factor: 7.038

6.  Dosimetric comparison of three different involved nodal irradiation techniques for stage II Hodgkin's lymphoma patients: conventional radiotherapy, intensity-modulated radiotherapy, and three-dimensional proton radiotherapy.

Authors:  Bhishamjit S Chera; Christina Rodriguez; Christopher G Morris; Debbie Louis; Daniel Yeung; Zuofeng Li; Nancy P Mendenhall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-04-20       Impact factor: 7.038

7.  A nomogram to predict radiation pneumonitis, derived from a combined analysis of RTOG 9311 and institutional data.

Authors:  Jeffrey D Bradley; Andrew Hope; Issam El Naqa; Aditya Apte; Patricia E Lindsay; Walter Bosch; John Matthews; William Sause; Mary V Graham; Joseph O Deasy
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-08-06       Impact factor: 7.038

8.  Dosimetric effects of jaw tracking in step-and-shoot intensity-modulated radiation therapy.

Authors:  Sarah Joy; George Starkschall; Stephen Kry; Mohammed Salehpour; R Allen White; Steven H Lin; Peter Balter
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

9.  Assessment of potential jaw-tracking advantage using control point sequences of VMAT planning.

Authors:  Jung-in Kim; Jong Min Park; So-Yeon Park; Chang Heon Choi; Hong-Gyun Wu; Sung-Joon Ye
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

10.  Advanced nasopharyngeal carcinoma radiotherapy with volumetric modulated arcs and the potential role of flattening filter-free beams.

Authors:  Mingzan Zhuang; Tuodan Zhang; Zhijian Chen; Zhixiong Lin; Derui Li; Xun Peng; Qingchun Qiu; Renhua Wu
Journal:  Radiat Oncol       Date:  2013-05-14       Impact factor: 3.481

View more
  11 in total

1.  Dosimetric comparison and biological evaluation of fixed-jaw intensity-modulated radiation therapy for T-shaped esophageal cancer.

Authors:  Hua Chen; Ying Huang; Hao Wang; Yan Shao; Ning J Yue; Hengle Gu; Yanhua Duan; Aihui Feng; Zhiyong Xu
Journal:  Radiat Oncol       Date:  2021-08-19       Impact factor: 3.481

2.  Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy.

Authors:  Sangutid Thongsawad; Chirasak Khamfongkhruea; Chirapha Tannanonta
Journal:  J Med Phys       Date:  2018 Jan-Mar

3.  Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study.

Authors:  Karthick Raj Mani; Sagar Upadhayay; K J Maria Das
Journal:  Radiat Oncol J       Date:  2017-03-31

Review 4.  Influence of the jaw tracking technique on the dose calculation accuracy of small field VMAT plans.

Authors:  Ans C C Swinnen; Michel C Öllers; Erik Roijen; Sebastiaan M Nijsten; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2017-01-03       Impact factor: 2.102

5.  A comparative study of identical VMAT about two adjacent targets with and without fixed-jaw technique.

Authors:  Kai Xie; Hongfei Sun; Liugang Gao; Tao Lin; Jianfeng Sui; Xinye Ni
Journal:  Radiat Oncol       Date:  2019-05-08       Impact factor: 3.481

6.  Fixed-jaw technique to improve IMRT plan quality for the treatment of cervical and upper thoracic esophageal cancer.

Authors:  Wei Song; Hong Lu; Jie Liu; Di Zhao; Jun Ma; Biyun Zhang; Dahai Yu; Xinchen Sun; Jinkai Li
Journal:  J Appl Clin Med Phys       Date:  2019-08-28       Impact factor: 2.102

7.  Selection Strategy of Jaw Tracking in VMAT Planning for Lung SBRT.

Authors:  Wuji Sun; Yinghua Shi; Yu Li; Chao Ge; Xu Yang; Wenming Xia; Kunzhi Chen; Libo Wang; Lihua Dong; Huidong Wang
Journal:  Front Oncol       Date:  2022-02-08       Impact factor: 6.244

8.  A comparative study of identical VMAT plans with and without jaw tracking technique.

Authors:  Hao Wu; Fan Jiang; Haizhen Yue; Qiaoqiao Hu; Jian Zhang; Zhuolun Liu; Jian Gong; Sha Li; Jianhao Geng; Yibao Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

9.  Dosimetric superiority of IMRT with jaw tracking technique for whole esophagus and T-shaped field radiotherapy in advanced esophageal cancer.

Authors:  Liwan Shi; Youqun Lai; Shanyu Chen; Lirong Fu; Qin Lin
Journal:  PLoS One       Date:  2018-09-05       Impact factor: 3.240

10.  The use of collimator angle optimization and jaw tracking for VMAT-based single-isocenter multiple-target stereotactic radiosurgery for up to six targets in the Varian Eclipse treatment planning system.

Authors:  Lauren M M Pudsey; Dean Cutajar; Alex Wallace; Anastasia Saba; Laurel Schmidt; Andrej Bece; Catherine Clark; Yoshiya Yamada; Giordano Biasi; Anatoly Rosenfeld; Joel Poder
Journal:  J Appl Clin Med Phys       Date:  2021-07-19       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.