Biomechanical Analysis of Cortical Versus Pedicle Screw Fixation Stability in TLIF, PLIF, and XLIF Applications.

Edward K Nomoto1, Guy R Fogel2, Alexandre Rasouli1, Justin V Bundy3, Alexander W Turner4.   

Abstract

STUDY
DESIGN: Cadaveric biomechanical study.
OBJECTIVES: Medial-to-lateral trajectory cortical screws are of clinical interest due to the ability to place them through a less disruptive, medialized exposure compared with conventional pedicle screws. In this study, cortical and pedicle screw trajectory stability was investigated in single-level transforaminal lumbar interbody fusion (TLIF), posterior lumbar interbody fusion (PLIF), and extreme lateral interbody fusion (XLIF) constructs.
METHODS: Eight lumbar spinal units were used for each interbody/screw trajectory combination. The following constructs were tested: TLIF + unilateral facetectomy (UF) + bilateral pedicle screws (BPS), TLIF + UF + bilateral cortical screws (BCS), PLIF + medial facetectomy (MF) + BPS, PLIF + bilateral facetectomy (BF) + BPS, PLIF + MF + BCS, PLIF + BF + BCS, XLIF + BPS, XLIF + BCS, and XLIF + bilateral laminotomy + BCS. Range of motion (ROM) in flexion-extension, lateral bending, and axial rotation was assessed using pure moments.
RESULTS: All instrumented constructs were significantly more rigid than intact (P < .05) in all test directions except TLIF + UF + BCS, PLIF + MF + BCS, and PLIF + BF + BCS in axial rotation. In general, XLIF and PLIF + MF constructs were more rigid (lowest ROM) than TLIF + UF and PLIF + BF constructs. In the presence of substantial iatrogenic destabilization (TLIF + UF and PLIF + BF), cortical screw constructs tended to be less rigid (higher ROM) than the same pedicle screw constructs in lateral bending and axial rotation; however, no statistically significant differences were found when comparing pedicle and cortical fixation for the same interbody procedures.
CONCLUSIONS: Both cortical and pedicle trajectory screw fixation provided stability to the 1-level interbody constructs. Constructs with the least iatrogenic destabilization were most rigid. The more destabilized constructs showed less lateral bending and axial rotation rigidity with cortical screws compared with pedicle screws. Further investigation is warranted to understand the clinical implications of differences between constructs.

Entities:  

Keywords:  cortical screw; lumbar interbody fusion; screw fixation; screw trajectory

Year:  2018        PMID: 30984495      PMCID: PMC6448199          DOI: 10.1177/2192568218779991

Source DB:  PubMed          Journal:  Global Spine J        ISSN: 2192-5682


Introduction

Segmental pedicle screw-rod fixation, pioneered by Roy-Camille[1] and modified by Weinstein,[2] is considered the gold standard for internal stabilization of the thoracolumbar spine. Complications with pedicle screws include superior facet violation and morbidity associated with soft tissue dissection. Superior facet violation may accelerate degenerative changes.[2-6] Dissection of the paraspinal muscles is needed as far lateral as the transverse processes. The multifidus muscle is considered especially vulnerable due to its monosegmental innervation.[7-9] Segmental nerve injury may lead to muscle atrophy, segmental instability, disc degeneration, and herniation.[10,11] Medial-to-lateral trajectory cortical screws have been proposed as an alternative to conventional pedicle screws as the medial screw entry point allows for less disruptive dissection, only as far lateral as the pars, in order to introduce the fixation and reduced likelihood of facet violation. Cortical screws have been described in clinical applications for posterior approaches such as transforaminal lumbar interbody fusion (TLIF) and posterior lumbar interbody fusion (PLIF).[12-18] In addition, cortical screws may be appropriate to provide internal stabilization to a transpsoas lateral lumbar interbody fusion (extreme lateral interbody fusion [XLIF]/lateral lumbar interbody fusion) where posterior fixation via a mini-open rather than a percutaneous approach is preferred, or when direct decompression is needed. The increased proportion of cortical bone along the screw trajectory and use of cortical thread forms may also offer improved screw fixation strength. There is an increasing number of biomechanical studies examining the stability provided by cortical screws for thoracolumbar spinal fixation. Screw toggle and pullout,[19-24] screw insertion torque,[21,25,26] single-level construct stability,[27,28] and multievel construct stability[29] using cortical screws have been investigated. While the stability of TLIF and XLIF constructs with pedicle and cortical screws has been investigated, the stability of PLIF constructs has not yet been reported. Furthermore, the biomechanical effect of extent of PLIF decompression and the consequence of XLIF direct decompression are also unknown. Finally, TLIF, PLIF, and XLIF constructs have not all been compared within the same analysis. In this lumbar cadaveric study, the multidirectional stability of bilateral medial-to-lateral trajectory cortical screw constructs and conventional trajectory pedicle screw constructs were compared when providing supplemental fixation to single-level TLIF, PLIF, or XLIF constructs.

Materials and Methods

Test Conditions

Specimens were initially tested intact. TLIF, PLIF, and XLIF interbody constructs were then tested with bilateral pedicle and cortical screw-rod fixation under the conditions listed in Table 1.
Table 1.

Test Conditions.

InterbodyPosterior DecompressionFixationCondition Name
TLIFUnilateral facetectomyBilateral pedicle screwsTLIF + UF + BPS
Unilateral facetectomyBilateral cortical screwsTLIF + UF + BCS
PLIFMedial facetectomyBilateral pedicle screwsPLIF + MF + BPS
Bilateral facetectomyBilateral pedicle screwsPLIF + BF + BPS
Medial facetectomyBilateral cortical screwsPLIF + MF + BCS
Bilateral facetectomyBilateral cortical screwsPLIF + BF + BCS
XLIFNoneBilateral pedicle screwsXLIF + BPS
NoneBilateral cortical screwsXLIF + BCS
Bilateral laminotomyBilateral cortical screwsXLIF + BL + BCS

Abbreviations: TLIF, transforaminal lumbar interbody fusion; UF, unilateral facetectomy; BPS, bilateral pedicle screw; BCS, bilateral cortical screw; PLIF, posterior lumbar interbody fusion; MF, medial facetectomy; BF, bilateral facetectomy; XLIF, extreme lateral interbody fusion; BL, bilateral laminotomy.

Test Conditions. Abbreviations: TLIF, transforaminal lumbar interbody fusion; UF, unilateral facetectomy; BPS, bilateral pedicle screw; BCS, bilateral cortical screw; PLIF, posterior lumbar interbody fusion; MF, medial facetectomy; BF, bilateral facetectomy; XLIF, extreme lateral interbody fusion; BL, bilateral laminotomy.

Specimen Preparation

Fresh frozen cadaver spines from T12 to the sacrum were obtained. Muscle tissue was removed, while ligament and disc tissue were retained. Spines were subjected to dual-energy X-ray absorptiometry (DEXA) scanning (Discovery C, Hologic, Inc, Bedford, MA) to rule out osteoporotic specimens and for specimen grouping, and A-P and lateral fluoroscopy to exclude deformity and significant degeneration. Two motion segments from each specimen (L1-2 and L3-4 in half of the specimens and L2-3 and L4-5 in the remaining specimens) were tested resulting in 8 constructs for each interbody and screw trajectory combination (TLIF with pedicle screws, TLIF with cortical screws, PLIF with pedicle screws, PLIF with cortical screws, XLIF with pedicle screws, and XLIF with cortical screws). Numerous studies have tested constructs with different levels from the same or different specimens.[30-34] Bone mineral density (BMD) was matched between groups. Cephalad and caudal specimen ends were potted in polyurethane resin (Smooth-Cast, Smooth-On, Inc, Easton, PA).

Surgical Procedures

Screw Fixation

Bilateral pedicle screws (BPS; Precept Pedicle Screw System, NuVasive, Inc, San Diego, CA) were placed along a conventional lateral-to-medial trajectory.[2] Screw diameters were Ø5.5 to 7.5 mm with lengths of 40 to 55 mm. Bilateral cortical screws (BCS; Precept Pedicle Screw System) were placed using a medial-to-lateral technique.[22] Screws diameters were all Ø5.0 mm with lengths of 30 to 35 mm.

Interbody and Posterior Decompression

TLIF specimens received a complete unilateral facetectomy (UF) and single interbody cage with 10 × 25 mm, 10 × 30 mm, or 10 × 35 mm footprint sizes (CoRoent LO, NuVasive, Inc) delivered along an oblique trajectory. PLIF specimens initially received medial facetectomies (MFs) that included bilateral resection of the medial aspect of the inferior articular processes. Bilateral interbody cages with 9 × 23 mm or 9 × 28 mm footprint sizes (CoRoent LMP, NuVasive, Inc) were delivered as far lateral as possible. Subsequently, complete bilateral facetectomies (BFs) were performed, whereby the articular processes were removed. XLIF specimens received laterally applied interbody cages with 18 × 50 mm or 18 × 55 mm footprint sizes (CoRoent XL, NuVasive, Inc) spanning the apophyseal ring. Cortical screw constructs initially had no posterior decompression; then, a bilateral laminotomy (BL) was performed, while pedicle screw constructs were only tested with intact posterior elements.

Biomechanical Testing

For each test condition, specimens were mounted on a custom 6 degrees of freedom spine test system described previously.[35] The loading protocol consisted of 3 cycles of pure, unconstrained moments to ±7.5 N m in each motion direction (flexion-extension, lateral bending, and axial rotation) without axial preload.[36,37] Motion was recorded across each tested level using infrared LED arrays attached to the vertebrae immediately cephalad and caudal to the interbody implant. The LEDs were tracked using an optoelectronic motion capture system (Optotrak Certus, Northern Digital, Inc, Waterloo, Ontario, Canada).

Data Analysis

Data from the third loading cycle in each test direction was analyzed. Range of motion (ROM) was normalized to intact. BMD and intact ROM were compared across all groups using 1-way analysis of variance. Normalized ROM of instrumented conditions were compared using 1-way analysis of variance and Tukey-Kramer post hoc comparisons, with P ≤ .05 considered significant.

Results

Average BMD was 0.897 g/cm2 (range = 0.767-1.089 g/cm2) for the PLIF specimens, 0.897 g/cm2 (range = 0.750-1.081 g/cm2) for the TLIF specimens, and 0.868 g/cm2 (range = 0.695-1.053 g/cm2) for the XLIF specimens. This was not found to be statistically significantly different between groups (P = .871). Average intact ROM for all groups in flexion-extension was 7.1 ± 2.6°, in lateral bending 8.1 ± 3.4°, and in axial rotation 2.9 ± 1.6°. There were no significant differences in intact ROM for any of the test groups (flexion-extension: P = .903, lateral bending: P = .706, axial rotation: P = .970). Comparing the instrumented constructs with intact (Figure 1), there was a significant (P ≤ .05) reduction in ROM for all TLIF, PLIF, and XLIF reconstructions in flexion-extension and lateral bending, with all producing less than 38.8% intact ROM. In axial rotation, all constructs significantly reduced ROM with regard to intact except TLIF + UF + BCS (88.6 ± 35.6% intact ROM), PLIF + MF + BCS (64.5 ± 24.0% intact ROM), and PLIF + BF + BCS (95.2 ± 32.3% intact ROM). Despite relatively modest average percent reductions in axial rotation ROM with regard to intact for these conditions, initial intact axial rotation ROM was much smaller than the other loading directions, and the instrumented ROM (in degrees) was lower than in lateral bending for the same constructs.
Figure 1.

Mean normalized range of motion (% intact ROM) for each test condition in: (A) flexion-extension, (B) lateral bending, and (C) axial rotation. Error bars indicate ±1 standard error of the mean (SEM). Test condition abbreviations are provided in Table 1.

Mean normalized range of motion (% intact ROM) for each test condition in: (A) flexion-extension, (B) lateral bending, and (C) axial rotation. Error bars indicate ±1 standard error of the mean (SEM). Test condition abbreviations are provided in Table 1. Evaluating the instrumented conditions, the most rigid construct (lowest ROM) in flexion-extension was XLIF + BL + BCS (8.8 ± 2.8%), and in lateral bending and axial rotation was XLIF + BPS (14.2 ± 5.2% and 38.7 ± 13.8%, respectively). Conversely, the least rigid (greatest ROM) construct in flexion-extension and axial rotation was PLIF + BF + BCS (20.2 ± 10.4% and 95.2 ± 32.3%, respectively), and in lateral bending it was TLIF + UF + BCS (38.8 ± 17.8%). Statistical analysis revealed that in flexion-extension XLIF + BCS, XLIF + BL + BCS, and PLIF + MF + BPS were significantly more rigid (P ≤ .025) than PLIF + BF + BCS. In lateral bending, XLIF + BPS and PLIF + MF + BPS were both significantly more rigid than TLIF + UF + BCS (P ≤ .023) as well as PLIF + BF + BCS (P ≤ .045). Likewise in axial rotation, XLIF + BPS and PLIF + MF + BPS were both significantly more rigid than TLIF + UF + BCS (P ≤ .015) as well as PLIF + BF + BCS (P ≤ 0.002). Additionally, XLIF + BCS was significantly more rigid than PLIF + BF + BCS (P = .050). Comparing the interbody constructs with pedicle screw fixation, no statistically significant differences were detected (P ≥ .605); however, generally XLIF provided the lowest ROM, followed by PLIF + MF and then TLIF + UF/PLIF + BL. Comparing interbody constructs with cortical screw fixation, XLIF with or without BL was more rigid than PLIF + BF in flexion-extension (P ≤ 0.007), and XLIF without BL was more rigid than PLIF + BF in axial rotation (P = .050), with no differences in lateral bending (P ≥ .614). Again, generally XLIF provided the lowest ROM, followed by PLIF + MF and then TLIF + UF/PLIF + BL. Last, comparing pedicle and cortical fixation for the same interbody procedures, no statistically significant differences were found (P ≥ .130).

Discussion

Medial-to-lateral trajectory cortical screws potentially offer an alternative thoracolumbar fixation option for interbody fusion techniques. In the current study, the XLIF and PLIF constructs retaining bilateral facet joints provided the most rigid stability. TLIF with UF and PLIF with BF were similar and both allowed more motion. The finding that XLIF is more stable than TLIF has been reported previously, attributable to the lack of iatrogenic destabilization and large buttressing interbody spacer with XLIF.[28,38] XLIF has not to our knowledge previously been compared with PLIF and was found here to have similar ROM provided only MFs were performed. PLIF has previously been shown to be more stable than TLIF, consistent with this study, again if most facet joint integrity is retained.[39,40] After bilateral facet resection, this study showed equivalent ROM for TLIF and PLIF in all directions. The relative stability of the various interbody constructs was found to be consistent when comparing those instrumented with pedicle screw as well as those with cortical screw, indicating that the underlying stability comes from retention of anatomical structures associated with the interbody technique. The importance of the facet joints in stabilizing the spine was demonstrated in this study when comparing the 2 PLIF constructs, where the BF test condition allowed numerically greater ROM compared with the MF in all directions tested. The largest differences were seen in axial rotation; however, no statistically significant differences were identified for any test directions. Facet joints are known to limit axial rotation in particular, as demonstrated by biomechanical studies reporting increased axial rotation following complete facetectomy.[41-43] The BLs performed with the XLIF constructs instrumented with BCS were found to have negligible impact on construct ROM (less than 5% intact ROM difference) and can thus be performed without introducing instability. A similar result can be expected for XLIF constructs with BPS. In this study, no statistically significant differences were detected between pedicle and cortical screw stabilization of the same interbody technique. There was a trend toward less ROM in flexion-extension with cortical screws supplementing XLIF and TLIF constructs, and greater ROM with cortical screws in lateral bending and axial rotation for all interbody constructs. Perez-Orribo et al[28] compared pedicle and cortical screw trajectories in TLIF and XLIF interbody applications. They also reported no significant differences between screw types when augmenting XLIF constructs, although they found similar trends to the current study with reduced motion with the cortical screws in flexion-extension and increased motion in lateral bending and axial rotation. Perez-Orribo et al[28] did detect a significantly reduced stiff zone with the TLIF construct and cortical screws compared with pedicle screw fixation in lateral bending. The clinical implication of ROM and stiffness differences between screw trajectories requires further evaluation. Long-term results for interbody fusions supplemented with cortical screw fixation are lacking. Short-term follow-up TLIF and PLIF studies indicate comparable clinical and radiographic outcomes at approximately 1 year to comparable interbody techniques with pedicle screws.[15,17] Additionally, these studies reported shorter incision length, quicker operative duration, and less blood loss compared with pedicle screws. Some case studies have reported screw loosening and loss of reduction; however, the majority of these did not include interbody support, which is likely a contributing factor.[14,44,45] Previous cadaveric cyclic toggle loading and pullout studies have reported equivalent or improved fixation strength for cortical trajectory screws over pedicle screws,[20,22-24] although a recent study reported lower fixation strength with cortical screws.[19] No screw loosening was observed in the current biomechanical study. This study has limitations in common with other cadaveric biomechanical investigations. Sample size in each group was limited to 8, based on availability, although multiple spinal levels were used in each specimen to increase the sample sizes. Additional samples may have allowed for detection of more statistical differences between test conditions; however, since statistical significance may not be associated with clinical significance, P values should be interpreted with caution. The loading used is simplified and excludes the effect of surrounding musculature and body weight; however, it is a repeatable and accepted method that is independent of specimen size. Another limitation of the study is that the MF PLIF condition retaining the majority of the facet joints would not be clinically feasible at the higher lumbar levels in most cases as the space would be insufficient to insert the 2 PLIF cages around the thecal sac. Despite this, we were able to evaluate the biomechanical stability of this condition, which should be applicable to the lower lumbar levels where the use of cortical screws is more prevalent. Strengths of the study included testing of 3 different interbody techniques under identical loading conditions.

Conclusion

Supplemental fixation of single-level TLIF, PLIF, or XLIF constructs with either pedicle or cortical screws provided significant increases in immediate postoperative stability in flexion-extension and lateral bending compared with the intact condition. In axial rotation, most conditions were also more rigid than intact. For the same interbody and iatrogenic destabilization procedure, cortical trajectory screws typically allowed greater lateral bending and axial rotation ROM than pedicle trajectory screws, although not statistically different. For the same screw trajectory, iatrogenic destabilization was the primary ROM contributor with XLIF constructs being most rigid, followed by PLIF with facet joints mostly retained. TLIF with UF and PLIF with BF were least stable. Further investigation is warranted to understand the clinical implications of differences between constructs.
  44 in total

1.  Influence of graded facetectomy and laminectomy on spinal biomechanics.

Authors:  T Zander; A Rohlmann; C Klöckner; G Bergmann
Journal:  Eur Spine J       Date:  2003-04-26       Impact factor: 3.134

2.  Radiologic evaluation of adjacent superior segment facet joint violation following transpedicular instrumentation of the lumbar spine.

Authors:  Rajesh R Shah; Saeed Mohammed; Asif Saifuddin; Benjamin A Taylor
Journal:  Spine (Phila Pa 1976)       Date:  2003-02-01       Impact factor: 3.468

Review 3.  Anatomic and technical considerations of pedicle screw fixation.

Authors:  J N Weinstein; B L Rydevik; W Rauschning
Journal:  Clin Orthop Relat Res       Date:  1992-11       Impact factor: 4.176

4.  Biomechanical comparison of posterior lumbar interbody fusion and transforaminal lumbar interbody fusion performed at 1 and 2 levels.

Authors:  Christopher P Ames; Frank L Acosta; John Chi; Jaicharan Iyengar; William Muiru; Emre Acaroglu; Christian M Puttlitz
Journal:  Spine (Phila Pa 1976)       Date:  2005-10-01       Impact factor: 3.468

5.  Revision of transforaminal lumbar interbody fusion using anterior lumbar interbody fusion: a biomechanical study in nonosteoporotic bone.

Authors:  Avraam Ploumis; Chunhui Wu; Amir Mehbod; Gustav Fischer; Antonio Faundez; Wentien Wu; Ensor Transfeldt
Journal:  J Neurosurg Spine       Date:  2010-01

6.  Technical factors related to the incidence of adjacent superior segment facet joint violation after transpedicular instrumentation in the lumbar spine.

Authors:  Zhiming Chen; Jie Zhao; Hao Xu; Aigang Liu; Jiandong Yuan; Cong Wang
Journal:  Eur Spine J       Date:  2008-09-16       Impact factor: 3.134

7.  Computed tomography evaluation of superior-segment facet-joint violation after pedicle instrumentation of the lumbar spine with a midline surgical approach.

Authors:  Ali Moshirfar; Louis G Jenis; Leo R Spector; Patrick J Burke; Elena Losina; Jeffrey N Katz; Frank F Rand; Scott G Tromanhauser; Robert J Banco
Journal:  Spine (Phila Pa 1976)       Date:  2006-10-15       Impact factor: 3.468

8.  In vitro, biomechanical comparison of an anterior lumbar interbody fusion with an anteriorly placed, low-profile lumbar plate and posteriorly placed pedicle screws or translaminar screws.

Authors:  Brian P Beaubien; Alihan Derincek; William D Lew; Kirkham B Wood
Journal:  Spine (Phila Pa 1976)       Date:  2005-08-15       Impact factor: 3.468

9.  Cortical bone trajectory for lumbar pedicle screws.

Authors:  B G Santoni; R A Hynes; K C McGilvray; G Rodriguez-Canessa; A S Lyons; M A W Henson; W J Womack; C M Puttlitz
Journal:  Spine J       Date:  2008-09-14       Impact factor: 4.166

10.  Effects of facetectomy and crosslink augmentation on motion segment flexibility in posterior lumbar interbody fusion.

Authors:  Norman B Chutkan; Haitao Zhou; John P Akins; Karl H Wenger
Journal:  Spine (Phila Pa 1976)       Date:  2008-10-15       Impact factor: 3.468

View more
  3 in total

1.  Lateral lumbar interbody fusion using a cellular allogeneic bone matrix in the treatment of symptomatic degenerative lumbar disc disease and lumbar spinal instability.

Authors:  William C Tally; H Thomas Temple; J Kenneth Burkus
Journal:  J Spine Surg       Date:  2021-09

Review 2.  Dynamic Stabilization Adjacent to Fusion versus Posterior Lumbar Interbody Fusion for the Treatment of Lumbar Degenerative Disease: A Meta-Analysis.

Authors:  Xiangyao Sun; Zhaoxiong Chen; Siyuan Sun; Wei Wang; Tongtong Zhang; Chao Kong; Shibao Lu
Journal:  Biomed Res Int       Date:  2020-05-20       Impact factor: 3.411

3.  Cadaveric biomechanical analysis of multilevel lateral lumbar interbody fusion with and without supplemental instrumentation.

Authors:  Oujie Lai; Yunlin Chen; Qixin Chen; Yong Hu; Weihu Ma
Journal:  BMC Musculoskelet Disord       Date:  2021-03-15       Impact factor: 2.362

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.