Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) versus conventional electro surgery in post-bariatric body-contouring procedures: A retrospective randomised clinical study.

Torsten Schlosshauer1, Marcus Kiehlmann1, Pablo Ramirez1, Marc-Oliver Riener2, Gabriel Djedovic1, Ulrich M Rieger1.   

Abstract

Postoperative wound-healing problems are relatively high in post-bariatric body-contouring procedures, partly because of electrosurgery and the associated thermal tissue damage. This study is a retrospective randomised evaluation of the effect of a low-thermal plasma dissection device (PEAK PlasmaBlade, Medtronic, Minneapolis, Minnesota) in comparison with conventional electrosurgery. A total of 24 patients undergoing upper arm or medial thigh lifting were randomised to PEAK PlasmaBlade on one side and to monopolar electrosurgery on the other side of the same patient. Wounds of 10 patients were examined histologically for acute thermal injury depth. Significantly lower total volume of drain output (61,1 ± 70,2 mL versus 95,1 ± 176,0 mL; P = .04) was found on the PEAK PlasmaBlade side compared with the electrosurgery side. Furthermore, the PEAK PlasmaBlade side showed fewer seromas (no case of seroma versus three seromas in the electrosurgery group) and less thermal damage (40% versus 70%; P = .26). Acute thermal injury depth from the PEAK PlasmaBlade was less than from monopolar electrosurgery (425 ± 171 μm versus 686 ± 1037 μm; P = .631). PEAK PlasmaBlade appears to be superior to traditional monopolar electrosurgery for post-bariatric body-contouring procedures because it demonstrated less tissue damage, lower total volume of drain output, and fewer postoperative seromas resulting in faster wound healing.
© 2019 Medicalhelplines.com Inc and John Wiley & Sons Ltd.

Entities:  

Keywords:  PEAK PlasmaBlade; body-contouring procedure; drain output; electrosurgery; seroma

Mesh:

Year:  2019        PMID: 30938101      PMCID: PMC7949418          DOI: 10.1111/iwj.13124

Source DB:  PubMed          Journal:  Int Wound J        ISSN: 1742-4801            Impact factor:   3.315


  26 in total

1.  Pulsed electron avalanche knife (PEAK) for intraocular surgery.

Authors:  D V Palanker; J M Miller; M F Marmor; S R Sanislo; P Huie; M S Blumenkranz
Journal:  Invest Ophthalmol Vis Sci       Date:  2001-10       Impact factor: 4.799

2.  Effects of the pulsed electron avalanche knife on retinal tissue.

Authors:  Daniel V Palanker; Michael F Marmor; Andre Branco; Philip Huie; Jason M Miller; Steven R Sanislo; Alexander Vankov; Mark S Blumenkranz
Journal:  Arch Ophthalmol       Date:  2002-05

3.  Comparison of wound-healing characteristics with feedback circuit electrosurgical generators in a porcine model.

Authors:  Harrison S Pollinger; Gamal Mostafa; Kristi L Harold; Catherine E Austin; Kent W Kercher; Brent D Matthews
Journal:  Am Surg       Date:  2003-12       Impact factor: 0.688

Review 4.  Concepts, considerations, and concerns on the cutting edge of radiofrequency ablation.

Authors:  Daniel B Brown
Journal:  J Vasc Interv Radiol       Date:  2005-05       Impact factor: 3.464

5.  Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) versus conventional electro surgery in post-bariatric body-contouring procedures: A retrospective randomised clinical study.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Pablo Ramirez; Marc-Oliver Riener; Gabriel Djedovic; Ulrich M Rieger
Journal:  Int Wound J       Date:  2019-04-01       Impact factor: 3.315

6.  Comparative healing of rat fascia following incision with three surgical instruments.

Authors:  Edward I Chang; Grace A Carlson; Joshua G Vose; Eric J Huang; George P Yang
Journal:  J Surg Res       Date:  2011-01-22       Impact factor: 2.192

7.  Comparative healing of human cutaneous surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a standard scalpel.

Authors:  Manuel E Ruidiaz; Davorka Messmer; Dominique Y Atmodjo; Joshua G Vose; Eric J Huang; Andrew C Kummel; Howard L Rosenberg; Geoffrey C Gurtner
Journal:  Plast Reconstr Surg       Date:  2011-07       Impact factor: 4.730

8.  Complications in body contouring procedures: an analysis of 1797 patients from the 2005 to 2010 American College of Surgeons National Surgical Quality Improvement Program databases.

Authors:  John P Fischer; Ari M Wes; Joseph M Serletti; Stephen J Kovach
Journal:  Plast Reconstr Surg       Date:  2013-12       Impact factor: 4.730

9.  Complications following body contouring surgery after massive weight loss: a meta-analysis.

Authors:  Emir Hasanbegovic; Jens Ahm Sørensen
Journal:  J Plast Reconstr Aesthet Surg       Date:  2013-10-28       Impact factor: 2.740

10.  Wound healing of skin incisions produced by ultrasonically vibrating knife, scalpel, electrosurgery, and carbon dioxide laser.

Authors:  R Hambley; P A Hebda; E Abell; B A Cohen; B V Jegasothy
Journal:  J Dermatol Surg Oncol       Date:  1988-11
View more
  4 in total

1.  Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) versus conventional electro surgery in post-bariatric body-contouring procedures: A retrospective randomised clinical study.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Pablo Ramirez; Marc-Oliver Riener; Gabriel Djedovic; Ulrich M Rieger
Journal:  Int Wound J       Date:  2019-04-01       Impact factor: 3.315

2.  Effect of low-thermal dissection device versus conventional electrocautery in mastectomy for female-to-male transgender patients.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Marc-Oliver Riener; Jens Rothenberger; Robert Sader; Ulrich M Rieger
Journal:  Int Wound J       Date:  2020-05-05       Impact factor: 3.315

3.  Comparative analysis on the effect of low-thermal plasma dissection device (PEAK PlasmaBlade) vs conventional electrosurgery in post-bariatric abdominoplasty: A retrospective randomised clinical study.

Authors:  Torsten Schlosshauer; Marcus Kiehlmann; Marc-Oliver Riener; Robert Sader; Ulrich M Rieger
Journal:  Int Wound J       Date:  2019-09-17       Impact factor: 3.315

4.  Post-Bariatric Plastic Surgery: Abdominoplasty, the State of the Art in Body Contouring.

Authors:  Payam Sadeghi; Daniela Duarte-Bateman; Wanyan Ma; Ryan Khalaf; R'ay Fodor; Gorizio Pieretti; Feliciano Ciccarelli; Hamed Harandi; Roberto Cuomo
Journal:  J Clin Med       Date:  2022-07-25       Impact factor: 4.964

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.