| Literature DB >> 30899771 |
Takeshi Kubo1, Ayami Ohno Kishimoto1, Kaori Togashi1.
Abstract
PURPOSE: To determine the reproducibility and productivity of reduced dose chest computed tomography (CT) using a nodule detection task.Entities:
Keywords: Automatic exposure control; Computed tomography; Pulmonary nodules; Radiation dose reduction
Year: 2019 PMID: 30899771 PMCID: PMC6412012 DOI: 10.1016/j.ejro.2019.02.001
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
The standard deviations of the CT numbers in the regions of interest.
| Image types | ||||
|---|---|---|---|---|
| Original | Simulation A | Simulation B | ||
| Image level | Pulmonary apex | 37±13.1 | 101.2±48.9 | 160.7±82.2 |
| Aortic arch | 36.6±14.6 | 92.7±36.7 | 147.7±64.3 | |
| Interventricular septum | 45.9±41 | 83.4±36.1 | 119±38.2 | |
| Right diaphragm | 69.5±85.8 | 103.8±79.3 | 138.4±80.1 | |
Statistically significant difference between original and simulation A images and simulation B images (p < 0.00001).
Fig. 1Area of ground-glass opacity in the right lower lobe measuring 7 mm was detected by a reader on the original image (left) but missed on the simulation A image (middle) and simulation B image (right). Another 5-mm solid nodule (not shown) is found in the right middle lobe and identified on both original and simulation images.
Fig. 2Area of ground-glass opacity measuring 21 mm in the left upper lobe was detected by a reader on the original image (left) and simulation A image (middle) and missed on simulation B image (right). The increased noise and streak artifacts decrease disease conspicuity.
Fig. 3A solid 5-mm nodule in the left upper lobe is depicted in the original image (left) with a well-defined border. The margin of the nodule become blurred by the overlapping streak artifacts in simulation A image (middle) and almost effaced in simulation B image (right).
Image quality scores recorded by three readers for original images and simulation A and B images.
| Original | Simulation | ||
|---|---|---|---|
| A | B | ||
| 1, Very good | 162 | 1 | 0 |
| 2, Good | 96 | 24 | 0 |
| 3, Fair | 6 | 198 | 46 |
| 4, Poor | 0 | 40 | 205 |
| 5, Very poor | 0 | 1 | 13 |
The number of detected nodules.
| Reader 1 | Reader 2 | Reader 3 | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| <5 mm | 5–9 mm | 10–29 mm | ≥30 mm | <5 mm | 5–9 mm | 10–29 mm | ≥30 mm | <5 mm | 5–9 mm | 10–29 mm | ≥30 mm | ||
| Original | GGN | 57 | 34 | 11 | 0 | 3 | 36 | 16 | 1 | 23 | 19 | 6 | 0 |
| Solid | 44 | 49 | 24 | 4 | 165 | 79 | 39 | 5 | 265 | 73 | 23 | 4 | |
| Total | 101 | 83 | 35 | 4 | 168 | 115 | 55 | 6 | 288 | 92 | 29 | 4 | |
| Simulation A | GGN | 39 | 36 | 10 | 0 | 7 | 39 | 20 | 0 | 29 | 28 | 6 | 0 |
| Solid | 31 | 37 | 26 | 4 | 164 | 91 | 39 | 5 | 211 | 56 | 24 | 5 | |
| Total | 70 | 73 | 36 | 4 | 171 | 130 | 59 | 5 | 240 | 84 | 30 | 5 | |
| Simulation B | GGN | 35 | 38 | 8 | 0 | 9 | 26 | 17 | 0 | 35 | 29 | 4 | 2 |
| Solid | 32 | 36 | 28 | 5 | 152 | 84 | 38 | 5 | 187* | 53 | 27 | 4 | |
| Total | 67* | 74 | 36 | 5 | 161 | 110 | 55 | 5 | 222 | 82 | 31 | 6 | |
Simulation A images, simulated images for standard deviation of 21; simulation B images, simulated images for standard deviation of 29; GGN, ground-glass nodules;*p-value <0.05 (Wilcoxon signed-rank test).
Concordance and discordance between the first reading and the second reading.
| Simulation A images | |||||||
|---|---|---|---|---|---|---|---|
| Reader 1 | Simulated | Reader 2 | Simulated | ||||
| Concordant | Discordant | Concordant | Discordant | ||||
| (a) All nodules < 5 mm | |||||||
| Original | Concordant | 451 | 30 | Original | Concordant | 383 | 38 |
| Discordant | 18 | 29 | Discordant | 38 | 69 | ||
| p = 0.083 | p = 1 | ||||||
| (b) Pure ground-glass nodules > 5 mm | |||||||
| Original | Concordant | 428 | 40 | Original | Concordant | 472 | 21 |
| Discordant | 30 | 30 | Discordant | 22 | 13 | ||
| p = 0.232 | p = 0.879 | ||||||
| (c) Solid or mixed nodules >5 and ≤ 10 mm | |||||||
| Original | Concordant | 483 | 16 | Original | Concordant | 443 | 27 |
| Discordant | 13 | 16 | Discordant | 22 | 36 | ||
| p = 0.577 | p = 0.475 | ||||||
| (d) Solid or mixed nodules >10 mm | |||||||
| Original | Concordant | 481 | 21 | Original | Concordant | 435 | 29 |
| Discordant | 12 | 14 | Discordant | 29 | 35 | ||
| p = 0.117 | p = 1 | ||||||
Simulation A images, simulated images for standard deviation of 21; simulation B images, simulated images for standard deviation of 29.
Reading time spent on one case by three readers.
| Original | Simulated A | p-value | Simulated B | p-value | |
|---|---|---|---|---|---|
| Reader 1 | 409.9±245.7 | 409.6±317.2 | 0.995 | 371.1±199.2 | 0.2 |
| Reader 2 | 320.3±96.1 | 323.8±99.6 | 0.77 | 344.4±100.2 | 0.057 |
| Reader 3 | 125.4±63.1 | 115.4±50 | 0.059 | 122.2±59.5 | 0.597 |
Means and standard deviations of the reading time are expressed in seconds. P-values for the differences between original images and simulated images are shown for simulation A images A and simulation B images.