Literature DB >> 15284432

Reduced radiation dose helical chest CT: effect on reader evaluation of structures and lung findings.

John R Mayo1, Kun-Il Kim, Sharyn L S MacDonald, Takeshi Johkoh, Peter Kavanagh, Harvey O Coxson, Sverre Vedal.   

Abstract

PURPOSE: To assess, by using computer simulation, the effect of the use of reduced computed tomographic (CT) tube current on reader evaluation of structures and lung findings on images obtained at clinically indicated chest CT examinations.
MATERIALS AND METHODS: The noise level in the raw scan data of 150 clinically indicated conventional tube current (200-320-mA) chest CT examinations was modified to simulate tube current reduction to 100 and to 40 mA. A total of 450 image sets were thus available. Four radiologists blinded to the tube current used assessed the image sets in random order for 14 structures and lung findings and ranked subjective image quality by using a five-point scale (1 = nondiagnostic, 2 = inferior, 3 = adequate, 4 = good, 5 = excellent). After a 3-week interval, the 150 conventional tube current image sets were rescored so that intraobserver agreement could be assessed. The McNemar statistic was used to determine whether there were more scoring disagreements between interpretations of the conventional and those of the reduced tube current scans or between the two interpretations of the conventional tube current scans.
RESULTS: When overall agreement for 14 structures and lung findings was pooled over four observers, significantly more disagreements (P <.05) were seen when scores were compared between conventional and reduced tube current scans than when scores were compared between repeated interpretations of the conventional tube current scans. There was a significant decrease (P <.05) in the subjective image quality of reduced tube current scans compared with the subjective image quality of conventional tube current scans.
CONCLUSION: These data indicate that reduced tube current does affect reader evaluation of structures and lung findings and reduces a reader's subjective assessment of image quality. Copyright RSNA, 2004

Mesh:

Year:  2004        PMID: 15284432     DOI: 10.1148/radiol.2323030899

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  10 in total

1.  A method to analyse observer disagreement in visual grading studies: example of assessed image quality in paediatric cerebral multidetector CT images.

Authors:  K Ledenius; E Svensson; F Stålhammar; L-M Wiklund; A Thilander-Klang
Journal:  Br J Radiol       Date:  2010-03-24       Impact factor: 3.039

2.  A computer-simulated liver phantom (virtual liver phantom) for multidetector computed tomography evaluation.

Authors:  Yoshinori Funama; Kazuo Awai; Osamu Miyazaki; Yoshiharu Nakayama; Da Liu; Taiga Goto; Yasuyuki Yamashita; Shinichi Hori
Journal:  Eur Radiol       Date:  2005-10-20       Impact factor: 5.315

Review 3.  CT protocols in interstitial lung diseases--a survey among members of the European Society of Thoracic Imaging and a review of the literature.

Authors:  Helmut Prosch; Cornelia M Schaefer-Prokop; Edith Eisenhuber; Daniela Kienzl; Christian J Herold
Journal:  Eur Radiol       Date:  2012-12-13       Impact factor: 5.315

4.  Observer-independent nodule-detectability index for low-dose lung cancer screening CT: a pilot study.

Authors:  Masaki Ohkubo; Shinichi Wada; Satoshi Kanai; Kazuhiro Ishikawa; Janaka C Marasinghe; Toru Matsumoto
Journal:  Radiol Phys Technol       Date:  2013-06-09

5.  Ultra-low-dose MDCT of the chest: influence on automated lung nodule detection.

Authors:  Ji Young Lee; Myung Jin Chung; Chin A Yi; Kyung Soo Lee
Journal:  Korean J Radiol       Date:  2008 Mar-Apr       Impact factor: 3.500

6.  Efficiency and reproducibility in pulmonary nodule detection in simulated dose reduction lung CT images.

Authors:  Takeshi Kubo; Ayami Ohno Kishimoto; Kaori Togashi
Journal:  Eur J Radiol Open       Date:  2019-03-11

7.  Performance of a new quantitative computed tomography index for interstitial lung disease assessment in systemic sclerosis.

Authors:  Marialuisa Bocchino; Dario Bruzzese; Michele D'Alto; Paola Argiento; Alessia Borgia; Annalisa Capaccio; Emanuele Romeo; Barbara Russo; Alessandro Sanduzzi; Tullio Valente; Nicola Sverzellati; Gaetano Rea; Serena Vettori
Journal:  Sci Rep       Date:  2019-07-01       Impact factor: 4.379

8.  Is weight-based adjustment of automatic exposure control necessary for the reduction of chest CT radiation dose?

Authors:  Priyanka Prakash; Mannudeep K Kalra; Matthew D Gilman; Jo-Anne O Shepard; Subba R Digumarthy
Journal:  Korean J Radiol       Date:  2009-12-28       Impact factor: 3.500

9.  Radiation dose reduction of chest CT with iterative reconstruction in image space - Part I: studies on image quality using dual source CT.

Authors:  Hye Jeon Hwang; Joon Beom Seo; Jin Seong Lee; Jae-Woo Song; Song Soo Kim; Hyun Joo Lee; Chae Hun Lim
Journal:  Korean J Radiol       Date:  2012-10-12       Impact factor: 3.500

10.  Low dose chest CT protocol (50 mAs) as a routine protocol for comprehensive assessment of intrathoracic abnormality.

Authors:  Takeshi Kubo; Yoshiharu Ohno; Mizuki Nishino; Pei-Jan Lin; Shiva Gautam; Hans-Ulrich Kauczor; Hiroto Hatabu
Journal:  Eur J Radiol Open       Date:  2016-04-27
  10 in total