Estimates of Patient Radiation Doses in Digital Radiography Using DICOM Information at a Large Teaching Hospital in Oman.

Ibrahim I Suliman1,2,3.   

Abstract

In this study, we sought to estimate the patient radiation doses in the digital radiography X-ray examinations conducted in a large hospital. The patient exposure factors and kerma-area product (PKA) were retrospectively recorded via the Digital Imaging and Communications in Medicine (DICOM) header for 547 patients. The entrance surface air kerma (ESAK) was estimated from the measurements of the X-ray tube output and recorded exposure factors, as well as from the console that displayed PKA as an alternative method. Effective doses were estimated from ESAK and PKA values using the appropriate conversion coefficient. In the chest PA, chest LAT, cervical spine AP, cervical spine LAT, abdomen AP, pelvis AP, lumbar spine AP, and lumbar spine LAT, the median ESAK (mGy) was found to be 0.13, 0.27, 0.35, 0.52, 0.70, 1.06, 2.33, and 4.18 mGy, respectively. Median PKA values were 0.10, 0.26, 0.14, 0.17, 0.77, 0.68, 0.81, and 1.11 Gy cm2, respectively. The estimated effective dose from ESAK and PKA values yielded comparable results. The comparison revealed that the ESAK and PKA values fell far below the reported in the literature. The results showed that the information of the DICOM deader is valuable for dosimetry and optimization.

Entities:  

Keywords:  Diagnostic reference levels; Digital radiography; Entrance surface air kerma; Kerma-area product; Patient doses

Mesh:

Year:  2020        PMID: 30891650      PMCID: PMC7064681          DOI: 10.1007/s10278-019-00199-y

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  20 in total

1.  Managing patient dose in digital radiology. A report of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  2004

2.  Development of Diagnostic Reference Levels Using a Real-Time Radiation Dose Monitoring System at a Cardiovascular Center in Korea.

Authors:  Jungsu Kim; Deoknam Seo; Inseok Choi; Sora Nam; Yongsu Yoon; Hyunji Kim; Jae Her; Seonggyu Han; Soonmu Kwon; Hunsik Park; Dongheon Yang; Jungmin Kim
Journal:  J Digit Imaging       Date:  2015-12       Impact factor: 4.056

3.  Current state of practice regarding digital radiography exposure indicators and deviation indices: Report of AAPM Imaging Physics Committee Task Group 232.

Authors:  Jaydev K Dave; A Kyle Jones; Ryan Fisher; Katie Hulme; Lynn Rill; David Zamora; Andrew Woodward; Samuel Brady; Robert D MacDougall; Lee Goldman; Susan Lang; Donald Peck; Bruce Apgar; S Jeff Shepard; Robert Uzenoff; Charles Willis
Journal:  Med Phys       Date:  2018-10-17       Impact factor: 4.071

4.  ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging.

Authors:  E Vañó; D L Miller; C J Martin; M M Rehani; K Kang; M Rosenstein; P Ortiz-López; S Mattsson; R Padovani; A Rogers
Journal:  Ann ICRP       Date:  2017-10

5.  Impact of digital imaging on radiation doses to the patient during X-ray examination of the urinary tract.

Authors:  B Sjöholm; H Geijer; J Persliden
Journal:  Acta Radiol       Date:  2005-10       Impact factor: 1.990

6.  Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients.

Authors:  A Herrmann; H Bonél; A Stäbler; C Kulinna; C Glaser; N Holzknecht; B Geiger; M Schätzl; F Reiser
Journal:  Eur Radiol       Date:  2001-11-13       Impact factor: 5.315

7.  Estimation of adult patient doses for common diagnostic X-ray examinations in Wad-madani, Sudan: derivation of local diagnostic reference levels.

Authors:  I I Suliman; T S Mohammedzein
Journal:  Australas Phys Eng Sci Med       Date:  2014-03-02       Impact factor: 1.430

8.  Patient dose management solution directly integrated in the RIS: "Gray Detector" software.

Authors:  A Nitrosi; A Corazza; M Bertolini; R Sghedoni; P Pattacini; M Iori
Journal:  J Digit Imaging       Date:  2014-12       Impact factor: 4.056

Review 9.  Radiation dose descriptors: BERT, COD, DAP, and other strange creatures.

Authors:  Edward L Nickoloff; Zheng Feng Lu; Ajoy K Dutta; James C So
Journal:  Radiographics       Date:  2008 Sep-Oct       Impact factor: 5.333

10.  DOSIMETRY OF ADULT AND PEDIATRIC PATIENTS FOR COMMON DIGITAL RADIOGRAPHY EXAMINATIONS.

Authors:  Bhupendra Singh Rana; Sanjeev Kumar; Inderjeet Singh Sandhu; Narinder Paul Singh
Journal:  Radiat Prot Dosimetry       Date:  2018-06-01       Impact factor: 0.972

View more
  1 in total

1.  Local diagnostic reference levels for routine chest X-ray examinations at a public sector hospital in central South Africa.

Authors:  Maurice Junda; Henra Muller; Hesta Friedrich-Nel
Journal:  Health SA       Date:  2021-08-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.