Literature DB >> 11870439

Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients.

A Herrmann1, H Bonél, A Stäbler, C Kulinna, C Glaser, N Holzknecht, B Geiger, M Schätzl, F Reiser.   

Abstract

The purpose of this study was to evaluate the performance of standard-dose and low-dose cesium iodide (CsI)-doted amorphous silicon (a-Si) flat-panel detector technology (FDT) as compared with storage-phosphor technology (SPT) in the depiction of relevant anatomical structures in chest radiography. In 75 patients referred for thoracic CT, digital chest radiographs were randomly obtained with either SPT at a standard dose (speed class S400, n=25), standard-dose FDT (S400, n=25) or FDT at a low dose (S800, n=25). Five radiologists evaluated the visibility of eight pulmonary and mediastinal anatomical structures using a five-point rating scale. To determine statistically significant differences between the three groups, the Mann-Whitney U-test was employed. No statistically significant differences were found in the depiction of eight criteria between SPT and standard-dose or low-dose FDT chest radiographs. The performance of FDT S400 was equal to SPT for most criteria and better for retrocardiac structures and soft tissue. FDT S800 was inferior to both SPT and FDT S400. Standard-dose FDT is equivalent to SPT in the depiction of relevant anatomical structures of the chest. Our results also indicate that a dose reduction of 50% with FDT may result in small but not significant decrease of image quality.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11870439     DOI: 10.1007/s00330-001-1166-4

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  9 in total

1.  Clinical results of CsI-detector-based dual-exposure dual energy in chest radiography.

Authors:  Jens Ricke; Frank Fischbach; Torsten Freund; Ulf Teichgräber; Enrique Lopez Hänninen; Rainer Röttgen; Ulrike Engert; Hermann Eichstädt; Roland Felix
Journal:  Eur Radiol       Date:  2003-06-19       Impact factor: 5.315

Review 2.  [Modern diagnosis of lung nodules].

Authors:  N D Abolmaali; T J Vogl
Journal:  Radiologe       Date:  2004-05       Impact factor: 0.635

Review 3.  Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literature.

Authors:  Markus Völk; Okka W Hamer; Stefan Feuerbach; Michael Strotzer
Journal:  Eur Radiol       Date:  2004-02-17       Impact factor: 5.315

Review 4.  Digital detectors for mammography: the technical challenges.

Authors:  A Noel; F Thibault
Journal:  Eur Radiol       Date:  2004-10-08       Impact factor: 5.315

5.  Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice.

Authors:  Wouter J H Veldkamp; Lucia J M Kroft; Mireille V Boot; Bart J A Mertens; Jacob Geleijns
Journal:  Eur Radiol       Date:  2005-08-31       Impact factor: 5.315

6.  Direct detector radiography versus dual reading computed radiography: feasibility of dose reduction in chest radiography.

Authors:  Michael Gruber; Martin Uffmann; Michael Weber; Mathias Prokop; Csilla Balassy; Cornelia Schaefer-Prokop
Journal:  Eur Radiol       Date:  2006-01-11       Impact factor: 5.315

7.  [The future of bedside chest radiography: Comparative study of mobile flat-panels and needle-image plate storage phosphor systems].

Authors:  K Bremicker; D Gosch; T Kahn; G Borte
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-04-24       Impact factor: 0.840

8.  Estimates of Patient Radiation Doses in Digital Radiography Using DICOM Information at a Large Teaching Hospital in Oman.

Authors:  Ibrahim I Suliman
Journal:  J Digit Imaging       Date:  2020-02       Impact factor: 4.056

9.  Copper filtration in pediatric digital X-ray imaging: its impact on image quality and dose.

Authors:  Philippe Brosi; Anja Stuessi; Francis R Verdun; Peter Vock; Rainer Wolf
Journal:  Radiol Phys Technol       Date:  2011-03-23
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.