Saied Nokar1, Amirali Reza Rasouli-Ghahroudi2, Elham Shidvash3, Faezeh Atri4. 1. Department of Prosthodontics, Dental Research Center, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran. 2. Department of Periodontics, Dental Implants Research Center, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran. 3. Department of Prosthodontics, Faculty of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran. 4. Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
Abstract
BACKGROUND: Primary stability is one of the prerequisites of immediate loading. The aim of this study was to compare the primary stability of four different implant systems in low-density bone models. MATERIALS AND METHODS: In this in vitro experimental study, 20 fixtures from four implant body designs were selected: Zimmer Tapered Screw-Vent (ZTSV), NobleReplace Tapered (NRT), Replace Select Tapered (RST), and Dentium SuperLine (DSL). Fixtures were inserted in low-density bone models according to manufacturer drilling protocol by one surgeon. Measurement of insertion torque value (ITV), resonance frequency analysis (implant stability quotient [ISQ]), and reverse torque value (RTV) was recorded for each fixture. The data were analyzed with one-way ANOVA and post hoc tests (Tukey honestly significant difference) (P < 0.05). RESULTS: ZTSV had significantly lower amount of insertion torque in comparison to other systems (P = 0.045). RTV was significantly lower in ZTSV in comparison to DSL and NRT (P = 0.004). ISQ value in NRT (ISQ = 67) was significantly higher than other systems (P = 0.000). The lowest amount of ISQ was in RST system (ISQ = 53) with significant difference (P = 0.000). CONCLUSION: This study demonstrated that primary stability of different implant systems was not comparable and implant design was effective on ITV, RTV, and ISQ.
BACKGROUND: Primary stability is one of the prerequisites of immediate loading. The aim of this study was to compare the primary stability of four different implant systems in low-density bone models. MATERIALS AND METHODS: In this in vitro experimental study, 20 fixtures from four implant body designs were selected: Zimmer Tapered Screw-Vent (ZTSV), NobleReplace Tapered (NRT), Replace Select Tapered (RST), and Dentium SuperLine (DSL). Fixtures were inserted in low-density bone models according to manufacturer drilling protocol by one surgeon. Measurement of insertion torque value (ITV), resonance frequency analysis (implant stability quotient [ISQ]), and reverse torque value (RTV) was recorded for each fixture. The data were analyzed with one-way ANOVA and post hoc tests (Tukey honestly significant difference) (P < 0.05). RESULTS: ZTSV had significantly lower amount of insertion torque in comparison to other systems (P = 0.045). RTV was significantly lower in ZTSV in comparison to DSL and NRT (P = 0.004). ISQ value in NRT (ISQ = 67) was significantly higher than other systems (P = 0.000). The lowest amount of ISQ was in RST system (ISQ = 53) with significant difference (P = 0.000). CONCLUSION: This study demonstrated that primary stability of different implant systems was not comparable and implant design was effective on ITV, RTV, and ISQ.
Immediate loading in dental implants has been become more popular due to patients’ demand of having teeth in the period of osseointegration. Immediate loading is defined as restoration of implant with fixed functional interim prosthesis at the time of implant insertion up to 48 hours later.[1]There are some essential prerequisites for the possibility of immediate loading: (1) selection of an implant system with great primary stability, (2) high bone implant surface contact,[2] and (3) reduction of micro-motion to <100 μm.[3]Implant primary stability is a mechanical fact, defined as the absence of clinical mobility at the time of insertion, and has an important role on immediate loading success because instability leads to fibrous encapsulation and implant failure.[4]Primary stability is related to implant design, method of implant insertion, quality and quantity of surrounding bone.[4] Some surgical methods are suggested to increase primary stability such as bicortical implant placement,[5] under-sized bed preparation,[6] and bone compression technique.[7] Some implant designs result in more stability such as conical fixtures[8] and wide diameter implants.[9] High-density bone provides more stability and in case of low-density bone, exact selection of implant design and surgical protocol is more important.[10]Methods of evaluation of implant stability are classified to invasive and noninvasive. Invasive techniques include histological analysis (evaluation of bone-implant contact in specimen), removal torque analysis, tensional and push out, pull out test. These methods are limited to nonclinical experimental because of ethical concerns. Noninvasive methods are surgeon's perception, radiographic analysis, cutting torque resistance, insertion or seating torque measurement, vibration test, percussion test, pulsed oscillation waveform, periotest, resonance frequency analysis (RFA), and magnetic technology.[11]To achieve the proper primary stability, bone quality has an important role, which is not changeable. Surgical methods of increasing stability are successful in some extent although they are not feasible in all cases and they require surgeon experience and dexterity. It seems that the most promising and logical way for getting higher stability in weak bone is to choose appropriate implant design.[10] The purpose of this study was to compare primary stability of four different popular implant designs, which are recommended by their companies for immediate loading. Primary stability in each design was determined through measuring insertion torque value (ITV), RFA and reverse torque value (RTV). It is hypothesized that primary stability of all of the implants would be equal.
MATERIALS AND METHODS
In this in vitro experimental, 20 dental implant with four different designs were selected. The sample size was determined at least five sample in each group according to Lachmann et al. study[12] by Minitab software, with α = 0.05 and β = 0.2 and standard deviation of 4.4. Selected dental implants were the factory recommended type for immediate loading: Zimmer Tapered Screw-Vent (ZTSV), Noble Replace Tapered (NRT), Replace Select Tapered (RST), and Dentium Super Line (DSL). Implant features are shown in Table 1.
Table 1
Features of implants used in the study groups
Features of implants used in the study groupsPolyurethane blocks (General Plastics manufacture Co, WA, USA) with density of 0.3 cm3/g was chose for bone model, which is categorized as low-density type according to ASTM F1839 (American Society for Testing and Materials).[13] This density is similar to type D4 of Lekholm classification and simulate thin layer of cortical bone surrounding a core of low-density trabecular bone.[14] Five polyurethane blocks were prepared in size of 40 mm in length, 30 mm in width, and 30 mm in height and divided into 4 equal rectangular. The center of each rectangular was the insertion place of each implant, and therefore, four type of implant were inserted in each blocks [Figure 1].
Figure 1
Prepared polyurethane block to place four types of implant.
Prepared polyurethane block to place four types of implant.All the fixtures were inserted according to the manufacturer's protocol by one expert clinician.The ITV was measured with connecting torque meter probe to superior part of fixtures to seat in prepared socket. Then, Osstell Mentor device (Osstell, Integration Diagnostic AB, GoteborgSvagen, Sweden) was used to measure the implant stability in implant stability quotient (ISQ). The appropriate smart peg of the instrument was selected according to company catalog in relation to size and type of implant and attached to fixture, to determine the primary stability of each implant according to the RFA values.Then, the least RTV needed for unscrewing the fixture was measured with torque meter (TQ 8800, Lutron Electronic Enterprise CO, Taipei, Taiwan) [Figure 2].
Figure 2
Measuring insertion torque value (a), resonance frequency analysis value (b), reverse torque value (c).
Measuring insertion torque value (a), resonance frequency analysis value (b), reverse torque value (c).Data analysis was done with SPSS (SPSS for Windows Inc. Version 22. Chicago, Illinois, IBM Corporation, USA). Normality of data was proved according to Kolmogorov–Smirnov analysis. One-way ANOVA was used to compare data in each group. Multiple comparison was done with post hoc test (Tukey honestly significant difference [HSD] type). P < 0.05 was considered as statistically significant.
RESULTS
The mean, maximum, minimum, and standard deviation of variables of each group is summarized in Table 2.
Table 2
Amount of resonance frequency analysis (implant stability quotient), insertion torque (N/Cm), and reverse torque (N/Cm) value in study groups
Amount of resonance frequency analysis (implant stability quotient), insertion torque (N/Cm), and reverse torque (N/Cm) value in study groupsEvaluation of the ISQ, ITV, and RTV in different groups by one-way ANOVA test revealed that the study groups were different in terms of each variable (P < 0.0001).According to post hoc analysis (Tukey HSD), the lowest amount of ISQ was in RRST with significant difference (P = 0.000). ISQ of ZTSV and DSL was equal, and the highest amount of ISQ was in NRT with significant difference with others (P = 0.000).ZTSV had the lowest ITV with significant difference with other systems (P = 0.045). The ITV between other systems had no significant difference.According to post hoc analysis, ZTSV had the lowest RTV, and the difference was significant (P = 0.004) with DSL and NRT implant. The RTV between other systems had no significant difference.The multiple comparison of systems is shown in Figure 3.
Figure 3
Multiple comparison of implant systems stability.
Multiple comparison of implant systems stability.
DISCUSSION
Based on the result, the study hypotheses are rejected, and primary stability of different implant systems was not equal.Primary stability is one of the prerequisites of immediate loading. It is proved that high primary stability decreases micromotion, bone resorption, and fibrous formation.[10] The main purpose of measuring the primary stability is to evaluate the micromotion.[15] Micromotion in D1 and D2 is less and primary stability could be easily achievable but in D3 and D4 bone, reaching to primary stability is difficult.[16] In this study, primary stability of four implant designs in D4 bone model is evaluated because in this bone quality, implant design seems to be more critical[10] and the purpose is to evaluate that in compromised situation, which implant design could be more efficient.Implant macroscopic features have great role in primary stability. Fixtures with more length provide more stability. It is proved that primary stability is in risk in fixture lengths <10 mm and length more than 15 mm does not benefit more stability.[17] Furthermore, more thread depth and width increase functional surface and stability.[18] Taper implants provide more stability than cylindrical form and are recommended to use in immediate loading[19] and when the use of short implants is necessary.[20] Although implant surface characteristic has no direct effect on primary stability, it improves bone healing and reduces micro movement because of coarse and bioactive surface, which could be effective on secondary stability.[1921] In the present study, four popular implant designs, which are the companies recommended types for immediate loading, were selected. They contained similar features such as taper body form, length of 12 or 13 mm and regular diameters.In this study, ISQ amount was in following sequence: NRT > ZTSV = DSL > RRST. ISQ was significantly higher in NRT system in comparison to other systems (mean = 67). Probably, it is related to double groovy (double thread) design, square form thread, with 0.42 mm depth and 0.7 mm width of threads, which increase functional surface and bone contact.[18] This result is in line with Ostman[22] Meredith[23] and Friberg et al.[24] studies, which showed the effect of macro design on ISQ value.ISQ value in ZTSV (mean = 61.8) and DSL (mean = 64) had no significant difference. In ZTSV system, thread form was V shape and in DSL system, fixture had double thread design and reverse buttress form. The reason of the same ISQ versus these design difference could be related to more number of threads in each unit of surface in ZTSV implant, which increase functional surface and bone contact. In addition, DSL implant contains self-cutting blade in half-epical part, which increases ease of insertion but decrease ISQ value. This result is compatible with Kim et al.[25] that showed ISQ is higher in nonself-cutting blade implants.ISQ value in RST system was significantly lower in comparison to other systems. It could be results of lacking of some specific macro design for increasing functional surface such as double thread or large number threads per unit.Although some studies showed, that ISQ is a good evaluation of primary stability for immediate loading,[2627] some studies mentioned that single time ISQ evaluation is not a proper determinant of stability and success of implant.[2829] ISQ shows amount of lateral stiffness and could not demonstrate the actual implant micromotion. Micromotion should be limited to <50–100 micron unless fibrous formation and bone loss would happen around implant.[1529] However, it is proved that there is a significant correlation between ISQ and micromotion.[29] It is reported that ISQ is more reliable in high-density bone and not trustworthy in poor quality bone (D3, D4).[30]It is claimed that ISQ should be analyzed for each system over the time and ISQ is not reliable indicator for comparison of several systems during loading in polyurethane block or specific area of jaw.[3132]In this study, to assess primary stability, ITV was recorded as well. ITV is correlated with implant micromotion. Trisi et al.[33] showed maximum ITV in low- and high-density bone could be 35 N/Cm and 100 N/Cm, respectively. Each 10N/Cm increase in ITV decreases the micromotion about 4 micron. ITV can be affected by noncalibrated surgical handpiece, hand pressure of surgeon, and unstable blocks during drilling, however, it is insignificant.[34]In this study, ITV was in following sequence: DSL > NRT = RRST > ZTSV. ZTSV had significantly lower ITV (mean = 9) in comparison to others. ZTSV has V-shaped thread, it is proved that square and reverse buttress thread shapes provide more stability than buttress and V-shaped thread.[18] Furthermore, ZTSV fixture is self-tapping which eases the insertion and reduces ITV, and the same finding of self-tapping effect is supported by Ostman[22] Javed and Romanos,[35] and Piattelli et al.[36]DSL showed the highest ITV (mean = 19) and it could be the result of double-thread design. The effect of this macro design is compatible with Chiapasco et al.,[37] Fazel et al.,[38] and Cochran et al.[39]RTV is an indicator of bone implant interfacial interface and was measured with torque controller device and the least torque needed to unscrew and remove the fixture was recorded.[40] RTV in ZTSV implant was lowest (mean = 7.72) with significant difference with NRT and DSL. This lower value in ZTSV system could be in association with V-shaped thread and consequently lower functional surface and lower force for detorquing.Lachmann et al.[12] showed that removal torque value is lower in implants with self-cutting blade in half-apical part. The least removal torque value for long-term stability is reported to be 30 NCm.[22]In this study, RTV was lower than ITV in all systems the reason could be related to deformation and compression of the surrounding bone model by lateral forces of implant placement.[33]In this study, although all systems had high ISQ, ITV and RTV were low, and they were not proper for immediate loading in D4 bone. In comparison, DSL and NRT with higher ISQ and ITV are better choice in low-density bone to prevent failures in the early healing period.
CONCLUSION
This study demonstrated that primary stability of different implant systems was not comparable and implant design was effective on ITV, RTV, and ISQ. According to study's limitation, the lowest amount of ITV and RTV was in ZTSV with no significant difference between others. The lowest and highest amount of ISQ value was in RST and NRT, respectively.
Financial support and sponsorship
This study was supported by Deputy of Research Affairs, Faculty of Dentistry, Tehran University of Medical Sciences.
Conflicts of interest
The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or nonfinancial in this article.
Authors: Luca Comuzzi; Margherita Tumedei; Ana Emilia Pontes; Adriano Piattelli; Giovanna Iezzi Journal: Int J Environ Res Public Health Date: 2020-04-11 Impact factor: 3.390