| Literature DB >> 30705961 |
Harminder Singh1, Donald R Duerksen1, Gale Schultz2, Carol Reidy3, Pat DeGagne4, Nancy Olson4, Zoann Nugent1, Michelle J Alfa4,5.
Abstract
Background and study aims Prevention of infection transmission from contaminated endoscopes would benefit from a rapid test that could detect low levels of viable bacteria after high level disinfection. The aim of this study was to evaluate the rapid NOW! (RN) test's ability to detect endoscope contamination. Materials and methods The RN test kit and the accompanying fluorometer were evaluated. The manufacturer states that a fluorometer signal > 300 units is indicative of viable Gram-negative bacteria. Suspension testing of varying concentrations of Escherichia coli, Pseudomonas aeruginosa and Enterococcus faecalis were used to determine the RN test limit of detection. Simulated-use testing was done using a duodenoscope inoculated with 10 % blood containing approximately 35 CFU E. coli per channel. Samples were extracted from the duodenoscope instrument channel and tested using the manufacturer's instructions. Results The RN test could consistently detect 10 CFU of E. coli and P. aeruginosa (fluorescent signal of 9,000 to 11,000 units) but not E. faecalis. Sensitivity and specificity for Gram-negative bacteria were 93 % and 90 %, respectively, using all of the suspensions in the study. Extraction of E. coli from an inoculated duodenoscope instrument channel repeatedly provided a positive signal (i. e. > 2,000 units). Conclusions The RN test can reliably detect low levels of Gram-negative bacteria in suspension as well as from samples extracted from endoscope channels. These preliminary findings are encouraging but further assessment of extraction efficacy, impact of organic residuals and clinical workflow are still needed.Entities:
Year: 2019 PMID: 30705961 PMCID: PMC6353648 DOI: 10.1055/a-0808-4342
Source DB: PubMed Journal: Endosc Int Open ISSN: 2196-9736
Simulated-use evaluation of the RN test for detection of E. coli in duodenoscope instrument channel.
| Replicate number | CFU Inoculum per channel |
Maximum Total CFU Tested
| NOW! test Incubation time (Hrs) | Numeric fluorometer value (NFV) |
|
| ||||
| 1 | 320.00 | 32.00 | 21.75 | 2376.00 |
| 2 | 360.00 | 36.00 | 21.67 | 2810.00 |
| 3 | 390.00 | 39.00 | 22.25 | 1832.00 |
| Average: | 356.70 | 35.67 | 21.89 | 2339.33 |
| SD: | 35.12 | 3.51 | 0.31 | 490.03 |
|
| ||||
| 1 | 15600.00 | 1560.00 | 20.00 | 2178.00 |
| 2 | 30000.00 | 3000.00 | 20.25 | 4534.00 |
| 3 | 22000.00 | 2200.00 | 21.92 | 4405.00 |
| Average: | 22500.00 | 2253.33 | 21.39 | 3705.67 |
| SD: | 7214.80 | 721.48 | 1.04 | 1324.57 |
Bacteria were suspended in (10 % sheep blood) and the instrument channel of a duodenoscope was inoculated and then allowed to dry for 2 hours before extraction as per MIFU (see Materials and Methods for details). The negative control from the test kit showed an NFV of 0 and 10 % blood without any bacteria had an NFV of 3.
The inoculum per channel was extracted in 5 mL extraction fluid so the maximum CFU tested is calculated as (CFU inoculated/5 mL) × 0.5 mL (volume of extracted channel sample that is tested).
Performance of the RN test for various bacterial concentrations suspended in sRO water.
| Numeric fluorometer value (NFV) for various CFU in suspension | |||||
|
Test 1
|
Test 2
|
Test 3
| |||
| Total CFU tested | NFV | Total CFU tested | NFV | Total CFU tested | NFV |
|
| |||||
| 13.50 | 178.00 | 20.00 | 331.00 | 15.50 | 43.00 |
| 1.35 | 84.00 | 2.00 | 145.00 | 1.55 | 255.00 |
| 0.14 | 240.00 | 0.20 | 320.00 | 0.16 | 270.00 |
| 0.01 | 193.00 | 0.02 | 121.00 | 0.02 | 246.00 |
|
| |||||
| 150.00 | 14373.00 | 225.00 | 15966.00 | 105.00 | 15348.00 |
| 15.00 | 10825.00 | 22.50 | 10052.00 | 10.50 | 11891.00 |
| 1.50 | 8389.00 | 2.25 | 9753.00 | 1.05 | 2109.00 |
| 0.15 | 117.00 | 0.23 | 1.00 | 0.11 | 124.00 |
|
| |||||
| 16.00 | 9551.00 | 23.50 | 24196.00 | 22.00 | 10896.00 |
| 1.60 | 13333.00 | 2.35 | 51.00 | 2.20 | 12036.00 |
| 0.16 | 13.00 | 0.24 | 38.00 | 0.22 | 27.00 |
| 0.02 | 70.00 | 0.02 | 0.00 | 0.02 | 5.00 |
Tests 1, 2 and 3 were independent bacterial suspensions prepared on three different days and each of the three organisms was evaluated separately (i. e. each microbe tested separately). Any Total CFU tested for a Gram-negative bacterium that is < 1 CFU would be considered to be a negative sample. Negative control test using sRO water showed an NFV of 92, 40 and 0 on three separate test days.