Literature DB >> 22317858

Development and validation of rapid use scope test strips to determine the efficacy of manual cleaning for flexible endoscope channels.

Michelle J Alfa1, Nancy Olson, Pat Degagné, Patricia J Simner.   

Abstract

BACKGROUND: Cleaning of flexible endoscopes is most commonly performed using manual methods that are often performed inadequately. The aim of this study was to validate the sample collection protocol and the Rapid Use Scope Test (RUST) and then assess its usefulness in clinical use.
METHODS: The benchmarks for adequate cleaning were protein <6.4 μg/cm(2), hemoglobin <2.2 μg/cm(2), and carbohydrate <1.2 μg/cm(2). Sample collection consisted of flushing 10 mL of sterile reverse osmosis water through the suction-biopsy port to the distal end. Validation of the RUST audit tool included simulated-use and in-use testing in 43 endoscopy clinics across Canada.
RESULTS: Simulated-use testing validated that improperly cleaned endoscopes that exceeded the cleaning benchmarks would be flagged by the RUST test. The clinical-use study indicated that 96.6% of 1,489 scope channels tested were RUST negative; however, 19% and 12% of elevator guide-wire channels and endoscopic retrograde colangiopancreatography channels, respectively, exceeded the benchmarks. The survey indicated that reprocessing personnel valued a rapid audit tool for assessing compliance with manual cleaning.
CONCLUSION: The validated RUST test provides health care users with a rapid audit tool for manual cleaning that can be integrated into the quality program in endoscopy.
Copyright © 2012 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22317858     DOI: 10.1016/j.ajic.2011.10.006

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  5 in total

Review 1.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

Review 2.  Infectious diseases linked to cross-contamination of flexible endoscopes.

Authors:  Nikki Kenters; Elisabeth G W Huijskens; Corianne Meier; Andreas Voss
Journal:  Endosc Int Open       Date:  2015-08

Review 3.  Recent Update on Microbiological Monitoring of Gastrointestinal Endoscopes after High-Level Disinfection.

Authors:  Suk Pyo Shin; Won Hee Kim
Journal:  Clin Endosc       Date:  2015-09-30

4.  Recent update of gastrointestinal endoscope reprocessing.

Authors:  Kyong Hee Hong; Yun Jeong Lim
Journal:  Clin Endosc       Date:  2013-05-31

5.  Evaluation of an overnight non-culture test for detection of viable Gram-negative bacteria in endoscope channels.

Authors:  Harminder Singh; Donald R Duerksen; Gale Schultz; Carol Reidy; Pat DeGagne; Nancy Olson; Zoann Nugent; Michelle J Alfa
Journal:  Endosc Int Open       Date:  2019-01-30
  5 in total